Restore Factory Defaults | /#<CR> | Restore Calibration Register Default Values |
Set Zero | /0 <CR> | Store Present Pressure as Instrument ZERO |
|
| (if Piezo <32 Torr & Pirani <50 mTorr) |
Force Zero | /!<CR> Override Limit Checks and Store Present | |
|
| Pressure as Instrument ZERO |
3.3.7.Device Status
When requested to transmit its status the
digit#: | 1 | 2 | 3 | 4 | 5 | |||||
Serial Receiver Overflow |
|
|
|
|
|
|
|
|
|
|
=4 |
|
|
|
|
|
|
|
| ||
Main Board EEPROM Error | =2 |
|
|
|
|
|
|
|
| |
Probe EEPROM (any) Error | =1 |
|
|
|
|
|
|
|
| |
Probe EEPROM Not Responding |
|
|
|
|
|
|
|
|
|
|
|
| =8 |
|
|
|
|
|
| ||
Probe EEPROM Read Error |
|
| =1 |
|
|
|
|
|
| |
Probe EEPROM Checksum Error |
|
|
|
|
|
|
|
|
|
|
|
|
|
| =4 |
|
|
|
| ||
Probe EEPROM Verification Error |
|
|
|
| =2 |
|
|
|
| |
Probe EEPROM Identification Error |
|
|
|
| =1 |
|
|
|
| |
Communications Syntax Error |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| =4 |
|
| ||
Piezo Sensor Bad (voltage out of range) |
|
|
|
|
|
| =2 |
|
| |
Pirani Sensor Bad (voltage out of range) |
|
|
|
|
|
| =1 |
|
| |
High Setpoint Alarm (pressure exceeds setpoint) |
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| =4 | ||||
Low Setpoint Alarm (pressure less than setpoint) |
|
|
|
|
|
| =2 | |||
Gas# Changed (not the same as when unit last calibrated) |
|
|
|
| =1 |
3.3.8.Default RS232/485 Specifications
Baud Rate | 9600 |
Character Length | Eight data bits |
Parity | None |
Stop Bits | 1 |
RS 485 Universal Address | 00 |