Teledyne HPM-2002-OBE instruction manual Device Status, Default RS232/485 Specifications

Page 20

Restore Factory Defaults

/#<CR>

Restore Calibration Register Default Values

Set Zero

/0 <CR>

Store Present Pressure as Instrument ZERO

 

 

(if Piezo <32 Torr & Pirani <50 mTorr)

Force Zero

/!<CR> Override Limit Checks and Store Present

 

 

Pressure as Instrument ZERO

3.3.7.Device Status

When requested to transmit its status the HPM-2002-OBE responds with a five digit number which is explained in the following:

digit#:

1

2

3

4

5

Serial Receiver Overflow

 

 

 

 

 

 

 

 

 

 

=4

 

 

 

 

 

 

 

 

Main Board EEPROM Error

=2

 

 

 

 

 

 

 

 

Probe EEPROM (any) Error

=1

 

 

 

 

 

 

 

 

Probe EEPROM Not Responding

 

 

 

 

 

 

 

 

 

 

 

 

=8

 

 

 

 

 

 

Probe EEPROM Read Error

 

 

=1

 

 

 

 

 

 

Probe EEPROM Checksum Error

 

 

 

 

 

 

 

 

 

 

 

 

 

 

=4

 

 

 

 

Probe EEPROM Verification Error

 

 

 

 

=2

 

 

 

 

Probe EEPROM Identification Error

 

 

 

 

=1

 

 

 

 

Communications Syntax Error

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

=4

 

 

Piezo Sensor Bad (voltage out of range)

 

 

 

 

 

 

=2

 

 

Pirani Sensor Bad (voltage out of range)

 

 

 

 

 

 

=1

 

 

High Setpoint Alarm (pressure exceeds setpoint)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

=4

Low Setpoint Alarm (pressure less than setpoint)

 

 

 

 

 

 

=2

Gas# Changed (not the same as when unit last calibrated)

 

 

 

 

=1

3.3.8.Default RS232/485 Specifications

Baud Rate

9600

Character Length

Eight data bits

Parity

None

Stop Bits

1

RS 485 Universal Address

00

HPM-2002-OBEVacuum Gauge

Page 20 of 39

Image 20
Contents Teledyne Hastings Manual Print History Table of Contents Troubleshooting Features Model 2002 SensorsHPM-2002-OBE Analog Output Module HPM-2002-OBE Analog Output 4-20mA ModuleHPM-2002-OBE DeviceNet Module SpecificationsHPM-2002-OBE RS232/485 Output Module Transducer Installation OBE Module InstallationReceiving Inspection Quick StartInitial Operation Pin Volt 20 mA RS232/485High and Low Set Point Modes Analog OutputOverall Functional Description Run Mode Cal ModeAnalog Output GAS ModeGas Selection Table HPM-2002-OBEVacuum Gauge Analog Output 4-20mA Cal Mode Analog Output 4-20mA HPM-2002-OBEVacuum Gauge Command Syntax RS232/485 With DisplayInterrogation Commands Paramater Modification CommandsCalibration Adjustment Commands Reset / Restore CommandsDefault RS232/485 Specifications Device StatusModular Connector Pinout R2-232 RJ-11 connector on left side Modular Connector Pinout RS-485Run Mode DeviceNet TM DeviceNet descriptionZero Coefficient Adjustment Midrange Coefficient AdjustmentAtmosphere Coefficient Adjustment Calibration of the HPM-2002-OBETheory of Operation Piezoresistive Sensor Vo = SPV+V1HPM-2002-OBEVacuum Gauge Pirani Sensor Et = Es + Er + EgEs = K ∆T Am/L Er = σεTh4-Ta4AsHPM-2002-OBEVacuum Gauge Dual Sensor Operation Eg = Kg ∆T As/∆xEg = arLt273/Th1/2Th-TaAgP HPM-2002-OBEVacuum Gauge Advanced Setup Guide RS485 AddressTurnaround Delay Restore EepromBaud Rate Data Bits/Parity Bits/Stop BitsRestore Calibration Restore Factory DefaultsSoftware Reset Frequently Asked Questions Why does my display read, Err1?Why does my display read, U.U? Why does my display read, U .?Why does my display read, . U? What are Err3, Err5, Err6 and Err7?Warranty Repair Policy Non-Warranty Repair PolicyDiagrams and Drawings HPM-2002-OBEVacuum Gauge HPM-2002-OBEVacuum Gauge