The setpoints may also be entered as a decimal number, e.g. [H=760.99<CR>] will be same as entering [H=7.6099E+2<CR>] .
When inputting setpoint data, it should be entered in the same Units of Pressure as the presently selected Units of Measurement (i.e. Torr, mbar or Pascal). The data is only checked to be a valid number with a one digit exponent before being accepted. There are no limit checks on the data; the user is free to choose any value appropriate to his use of the instrument.
The Turnaround Delay and
If the
Caution: Since all units will respond to the ‘UNIVERSAL ADDRESS’, make sure that only the unit to be modified is connected to the
The ‘UNIVERSAL ADDRESS’ is used for setup only; (i.e. to set all units to the same gas #) data is never transmitted when the ‘UNIVERSAL ADDRESS’ is used.
The value entered for the Turnaround Delay is used to modify an internal timer which normally runs at ~8 millisecond, e.g. [*{aa}T=10<CR>] will set the delay to ~80 milliseconds.
If the command syntax is not met or if the number is out or range, the
3.3.5.Calibration Adjustment Commands
Command Description | Format | Valid Range: | |
Set Full Scale Calibration | CF={m.d}E{e}<CR> | 5.12e+2 | to 1.023e+3 Torr |
|
| 6.83e+2 | to 1.365e+3 mbar |
|
| 6.83e+4 | to 1.365e+5 Pascal |
Set Midpoint Calibration | CM={m.d}E{e}<CR> | 4.00e+0 | to 7.999e+0 Torr |
|
| 5.34e+0 | to 1.066e+1 mbar |
|
| 5.34e+2 | to 1.066e+3 Pascal |
Set Low Scale Calibration | CL={m.d}E{e}<CR> | 0 to | |
|
| 0 to | |
|
| 0 to 1.666e+1 Pascal |
Notes:
The calibration adjustment data may also be entered as a decimal number, e.g. [CF=760.99<CR>] will be same as entering [CF=7.6099E+2<CR>].
When inputting calibration adjustment data, it must be within the valid range of the presently selected Unit of Measurement (i.e. Torr, mbar or Pascal). The data is checked to be valid before being accepted.
If the command syntax is not met or if the number is out or range, the
The ‘UNIVERSAL ADDRESS’ may be used to calibrate all connected units simultaneously, e.g. [*00CF=760<CR>].
3.3.6.Reset / Restore Commands
Command Description | Format | Notes: |
Escape | <Esc> | Reset Command Buffer (ignore prior Input) |
Software Reset | /R<CR> | Reinitialize Software |
|
| Page 19 of 39 |