Falcon F110 manual Email Settings

Page 22

3F110 Web Interface

3.3.2Email

The Email tab stores settings for the F110 to send Email Alert Notifications when alert thresholds are exceeded. In order to enable email alerting from the F110's web server interface, your mail server must have SMTP Relay enabled for either your network or the IP address of the F110. Contact your Email Administrator if you need assistance.

Enter the mail server address for the mail server you are using in the Mail Server Address field. The address can be entered as either the IP address or the DNS name of your mail server. If you wish to use a DNS name for the Mail Server Address field, you must assign a Static IP address, Gateway IP address, DNS Server IP address, and Subnet Mask on the Network tab.

In the Email Recipients fields, you can enter a single/regular email address, distribution list email address or mobile phone, pager or PDA email address (using Email-To-SMS) for email alert when alert thresholds are exceeded. Only one address may be entered.

The Return Address (From) field MUST contain an email address for an active account on the mail server specified in the Mail Server Address field. For example, if 'mail.rletech.com' is entered in the Mail Server Address field, the Return Address (From) field must contain an email address with the 'mail.rletech.com' domain. Contact your email administrator if you require additional assistance.

The F110 can also do authenticated SMTP (ESMTP). Enter your user name and password in the fields and select the check box for Enable Authentication. You MUST save your settings before sending a test email. Refer to xrefpg. X for Trace output, to view mail transfer information between the F110 and mail server for troubleshooting.

Note You must save your settings before sending a test email.

Figure 3.4 Email Settings

www.rletech.com

22

970.484.6510

Image 22
Contents Monitoring Revision History Product Registration Rletech.com 970.484.6510 Contents Rletech.com Figures Rletech.com 970.484.6510 Tables Rletech.com System Overview Product DescriptionFront Panel Indicators and Controls Front PanelTerminal Block Designations F110 Terminal Block DesignationsSystem Overview 970.484.6510 Getting Started InstallationCommunication Set IP SettingsSearch IP Address Assignment for Static Only NetworksUpdate F110 WiringPower Supply 2 RJ45 Ethernet Connection Physical ConnectionF110 Web Interface Overview F110 WEB InterfaceStatus Screen Settings Screen Authentication Security Prompt no Username or PasswordNetwork Settings NetworkEmail Settings Snmp Settings SnmpExternal Temperature/Humidity Channel 2, Channel 3 and ChannelSwitched Sensor Alarm Configuration SecuritySecurity Settings AdvancedReset Defaults Restore Defaults PromptHelp Screen About ScreenTelnet Configuration Modbus Configuration VIA TelnetF110 Modbus Capabilities Enable MODBUS? N? Modbus Registers Register Name Description Units RangeHexadecimal Decimal Value Description Switch Sensor 1 Alarm On 1 = Closed, 0 = Open Modbus Registers 970.484.6510 Technical Specifications Inputs Temperature/HumidityOperating Environment Temperature