Guidance and manufacturer’s declaration – electromagnetic immunity

The 1528L is intended for use in the electromagnetic environment specified below. The customer or the user of the 1528L should assure that it is used in such an environment.

Immunity test

IEC60601

Compliance level

Electromagnetic environment –

test level

guidance

 

 

 

 

 

 

Electrostatic discharge

±6 kV Contact

±6 kV Contact

Floors should be wood, concrete or

(ESD)

±8 kV air

±8 kV air

ceramic tile. If floors are covered with

IEC61000-4-2

 

 

synthetic material the relative humidity

 

 

 

should be at least 30 %

 

 

 

 

Electrical fast

±2 kV for power supply

±2 kV for power supply

Main power quality should be that of a

transient/burst

lines

lines

typical commercial or hospital

IEC61000-4-4

±1 kV for input/output

±1 kV for input/output

environment.

 

lines

lines

 

 

 

 

 

Surge

±1 kV differential mode

±1 kV differential mode

Main power quality should be that of a

IEC6100-4-5

±2 kV common mode

±2 kV common mode

typical commercial or hospital

 

 

 

environment.

 

 

 

 

Voltage dips, short

<5 % UT

<5 % UT

Main power quality should be that of a

interruptions and voltage

(>95 % dip in UT)

(>95 % dip in UT)

typical commercial or hospital

variations on the power

for 0,5 cycle

for 0,5 cycle

environment. If the user of the 1528L

supply input lines

40 % UT

40 % UT

requires continued operation during

IEC61000-4-11

power mains interruptions, it is

 

(60 % dip in UT)

(60 % dip in UT)

recommended that the 1528L be

 

for 5 cycles

for 5 cycles

powered from an uninterruptible power

 

70 % UT

70 % UT

supply or a battery.

 

 

 

(30 % dip in UT)

(30 % dip in UT)

 

 

for 25 cycles

for 25 cycles

 

 

<5 % UT

<5 % UT

 

 

(>95 % dip in UT)

(>95 % dip in UT)

 

 

for 5 sec

for 5 sec

 

 

 

 

 

Power frequency

3 A/m

3 A/m, 30 A/m

Power frequency magnetic fields

(50/60 Hz)

 

 

should be at levels characteristic of a

Magnetic field

 

 

typical location in a typical commercial

IEC61000-4-8

 

 

or hospital environment.

 

 

 

 

NOTE UT is the ac. mains voltage prior to application of the test level.

User Manual: 1528L

SW200095 Rev A, Page 7 of 39

Image 7
Contents Elo Touch Solutions 1528L Touchmonitors Trademark Acknowledgments DisclaimerSW200095 Rev A, Page 3 Critical functions are Environmental conditions for transport and storage ClassificationEMC Guidance Symbol DescriptionSW200095 Rev A, Page 7 Portable and mobile RF communications equipment Separation distance according to frequency of transmitter Table of Contents Product Description LCD Monitor Performance FeaturesExternal Medical Grade Power Adapter IntroductionUnpacking Your Touchmonitor Installation and SetupFront View Rear View Product OverviewSide View Kensington Lock Remove the Cable Cover USB Interface ConnectionSW200095 Rev A, Page 16 Replace the Cable Cover Installing the Touch Driver SoftwareOptimizing the LCD Monitor Installing the USB Touch Driver Operation About Touchmonitor AdjustmentsControls and Adjustment Auto Adjust Recall DefaultsPhase OSD TimeoutMonitor Angle desktop only Power LED Monitor & Power SavingGeneral Power Saving Mode Power Saving Indicator Mode ConsumptionSolutions to Common Problems TroubleshootingProblem Suggestions Input Video 15.0 LCD Native ResolutionVGA Svga XGA Touchmonitor Safety Waste Electrical and Electronic EquipmentWEEE Directive Care and Handling of Your TouchmonitorMonitor Modes Technical SpecificationsResolution Type Scan KHz Scan Hz Polarity Display ModesDimensions H x W Weight Net Certifications Value 1528L Touch Display SpecificationsHumidity non- condensing North America Power Adapter Cord SelectionEquivalent Types Cert Country Agency Mark Cord selection for other than North AmericaHAR Flexible Cord Mechanical Light Transmission per Astm D1003 Visual ResolutionOptical AccuTouch resistive Technology Touchscreen SpecificationsDurability CarollTouch infrared Technology Touchscreen SpecificationsElectrical EnvironmentalReplacement Parts Part Number Contact Elo Touch SolutionsContact Elo TouchSystems Regulatory Information Electrical Safety InformationII. Emissions and Immunity Information SW200095 Rev A, Page 37 Warranty Check out Elo’s Website Getting in Touch with Elo