Recommended separation distances between

Portable and mobile RF communications equipment and the 1528L

The 1528L is intended for use in which radiated RF disturbances are controlled. The customer or the user of the 1528L can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the 1528L as recommended below, according to the maximum output power of the communications equipment.

 

Separation distance according to frequency of transmitter

Rated maximum output power of

 

M

 

 

 

 

transmitter

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2,5 GHz

W

__

__

__

 

d = 1,2 √P

d = 1,2 √P

d = 2,3 √P

 

 

 

 

0,01

0,12

0,12

0,23

 

 

 

 

0.1

0,38

0,38

0.73

 

 

 

 

1

1,2

1,2

2,3

 

 

 

 

10

3,8

3,8

7,3

 

 

 

 

100

12

12

23

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

This device complies with all applicable electromagnetic emission and immunity standards for medical device equipment. This device is designed to not cause harmful interference, and to accept any interference received, including interference that may cause undesired operation. The performance of this device is limited to the emission and immunity standards that have been applied. Other devices which are not designed to withstand emission levels as specified in the medical device standards may be susceptible to interference from this device. Subjecting the device to conditions beyond the rated performance capabilities may result in emissions in excess of the standard. If it is determined that this device produces electromagnetic or other interference it must be disconnected from power until the cause of the problem has been determined and resolved. If it is determined that this device is functioning improperly due to electromagnetic and other interference it must be disconnected from power until the cause of the problem has been determined and resolved.

User Manual: 1528L

SW200095 Rev A, Page 9 of 39

Image 9
Contents Elo Touch Solutions 1528L Touchmonitors Trademark Acknowledgments DisclaimerSW200095 Rev A, Page 3 Critical functions are Environmental conditions for transport and storage Classification EMC Guidance Symbol DescriptionSW200095 Rev A, Page 7 Portable and mobile RF communications equipment Separation distance according to frequency of transmitter Table of Contents External Medical Grade Power Adapter LCD Monitor Performance FeaturesIntroduction Product DescriptionUnpacking Your Touchmonitor Installation and SetupFront View Rear View Product OverviewSide View Kensington Lock Remove the Cable Cover USB Interface ConnectionSW200095 Rev A, Page 16 Installing the Touch Driver Software Replace the Cable CoverOptimizing the LCD Monitor Installing the USB Touch Driver Operation About Touchmonitor AdjustmentsControls and Adjustment Phase Recall DefaultsOSD Timeout Auto AdjustGeneral Power Saving Mode Power LED Monitor & Power SavingPower Saving Indicator Mode Consumption Monitor Angle desktop onlyTroubleshooting Solutions to Common ProblemsProblem Suggestions Native Resolution Input Video 15.0 LCDVGA Svga XGA Touchmonitor Safety Waste Electrical and Electronic EquipmentWEEE Directive Care and Handling of Your TouchmonitorResolution Type Scan KHz Scan Hz Polarity Technical SpecificationsDisplay Modes Monitor Modes1528L Touch Display Specifications Dimensions H x W Weight Net Certifications ValueHumidity non- condensing Power Adapter Cord Selection North AmericaEquivalent Types Cert Country Agency Mark Cord selection for other than North AmericaHAR Flexible Cord Mechanical Light Transmission per Astm D1003 Visual ResolutionOptical AccuTouch resistive Technology Touchscreen SpecificationsElectrical CarollTouch infrared Technology Touchscreen SpecificationsEnvironmental DurabilityContact Elo Touch Solutions Replacement Parts Part NumberContact Elo TouchSystems Electrical Safety Information Regulatory InformationII. Emissions and Immunity Information SW200095 Rev A, Page 37 Warranty Check out Elo’s Website Getting in Touch with Elo