3M eg4 user manual Dosimeter setting, Dosimeter screen

Page 30

24

Customizing set-up parameters in DMS

Dosimeter setting

Dosimeter setting

The Edge general settings are used to setup the instrument with basic parameters. The eg4 model features two virtual dosimeters in one with two customized, user-defined settings. Please refer to the Edge dosimeter settings table below for an explanation of setup details.

1.Click on the Dosimeter tab from the Edge’s Instrument Configuration panel (see 1). To change the dosimeter settings, follow one (or all) of the selection options below.

Click on the checkbox to enable the dosimeter (LED) heartbeat, Dose indicator LED, and/or the Meter 1/Meter 2 Threshold settings.

Click in the circular button to select the response time (fast or slow), peak weighting (C or Z), and/or RMS Weighting (A or C).

To set the numeric fields either type in a value or use the arrows to enter appropriate values for the Meter1/Meter 2 settings (threshold, upper limit, exchange rate, criterion level, and RMS weighting).

2.When Meter 1 and Meter 2 settings are “enabled”, this will log (for eg4 model only) two different setups simultaneously. Typically, users will set the meter 1 and meter 2 with different regulatory or local standards (such as EU directive and OSHA).

3.To set the dosimeter settings, save and send to the Edge (for more details see “Saving and

sending configurations”).

Dosimeter settings

explained in the Edge dosimeter settings (below).

1

Figure 3-6: Dosimeter screen

053-644, Rev H

Page 24

eg4

Image 30
Contents Edge 3M The Edge Personal Noise DosimeterTable of Contents Setting-up the Edge via DMS Time history……………………………………………………………………………………………………Technical Support Stopping a studyViewing measurements on the display Replacing microphone and windscreenPhysical attributes of EdgeDock1 and EdgeDock5 INDEX…………………………………………………………………………………List of Figures List of TablesDosimetry Edge and dosimetryNoise dosimeter Illustration of how the dosimeter operates and processesRange of sound What do the standards say?Applications of the Edge Running your studies Setting-up your dosimeter parameters for your noise studiesNoise assessment steps View your results via DMSParameter setup screen for eg4 Edge ModelChecking the equipment Identifying your equipmentLED indicator Turning onLED indicators explained Turning offDiagram Display and keypad3MTMDetection Mangement Software DMS Docking station functionalityLoading Docking station components EdgeDock1 or EdgeDock5EdgeDock1 and EdgeDock5 components Powering EdgeDock1 or EdgeDock5  Powering EdgeDock1 or EdgeDock5 How to charge Charging your dosimeterMounting the docking station to a wall Storing the dosimeterUnderstanding what data is stored Noise studiesSummary data Time historyViews on the Edge LoggingDefinitions of data log values and views Views on ExplanationDosimeter views and definitions Dose WeightingViews on Explanation Edge Response LevelDescriptors Variants Lmax variants Dosimeter measurement descriptors/variantsDocking Communicating & downloading dataCustomizing set-up parameters in DMS New Hardware wizard for Edge DockDosimeter setting Dosimeter screenEdge dosimeter settings Dosimeter fields explained for the EdgeDisplay setting Edge display screenSecurity setting Edge security settingAuto Run setting Edge auto run screenFirmware update 10 Edge firmware update screenSaving and sending/setting configurations 11 Saving and sending Edge setupsOverview of running a study Quick Help List typical dosimetry sessionWindscreen and maintenance  How to attach the dosimeterPositioning the Edge on your shirt Checking battery power  Checking battery powerCalibrating  Calibrating the dosimeterResetting the dosimeter  ResetSetting contrast on dosimeter  ContrastOn/off/enter button  How to run a noise study Running studiesStarting a run Indicator Icon Explanation Indicators/Icons on the display How to stop a noise study Indicators i.e., run, stop, battery, and OLViewing measurements on the display Locking or securing your dosimeter during run Viewing measurements Locking or Securing  How to lock or secure How to unlock UnlockingDownloading data  DownloadingDMS start screen Viewing data and data finder  Selecting and viewing dataReport View Example of Edge Report in DMSPanel Layout Viewcharts/graphs Panel layout view viewing the data in charts and graphsLED indicator disable/enable Technical SupportReplacing microphone and windscreen Attaching/Detaching Attaching/Detaching the windscreen and microphone Hard Hat Attachment Kit Hard Hat Attachment kitMechanical characteristics StandardsElectrical characteristics Environmental effectsUser Interface EdgeDock1 and EdgeDock5 Physical attributes of EdgeDock1 and EdgeDock5Measurements TypesMeasurement parameters Security CommunicationsSpecial functions Auto-RunEdge Block Diagram Edge Block Diagram Appendix B Service Department and Technical Support 1 800 3M ServiceWarranty Contacting 3MGlossary of Terms Background Noise, effectsCombining Noise sources Crest FactorCriterion Level CL Dose Criterion TimeDecibel dB Exchange Rate ER also known as the Doubling RateLavg Average Level Free field setting microphoneHearing Conservation HC Hertz Hz Lep,dMaximum Level Lmax Leq Equivalent LevelLogging Measurement RangeRandom incidence microphone PEL Permissible Exposure LevelProjected Time Response TimeTWA Time Weighted Average ThresholdWeighting A, C, Z, etc Upper Limit ULWindscreen Measurement equations Index Running studies About 3M Personal Safety About Us