3M eg4 user manual Setting contrast on dosimeter,  Contrast, On/off/enter button

Page 42

36

Overview of running a study

Setting contrast on dosimeter

Setting contrast on dosimeter

Contrast

1.

Ensure the meter is Off.

 

 

2.

Turn the meter on by pressing the on/off/enter

button.

 

3.

When the splash screen appears, press and hold the run/stop

and on/off enter

 

buttons. (See figure 4-5 for splash screen).

 

 

4.The Setup menu appears on the Edge. Repeatedly press the arrow key until the Contrast symbol is selected.

NOTE: Setup menu is used to reset the dosimeter, change the contrast, and return to the main view.

Contrast icon

Checkmark icon Is the selection to exit the setup menu

Arrows and enter key Indicates the valid keys are selected

Figure 4-7: Changing the contrast from the keypad

5.Next press the on/off enter button.

6.A number will appear on your display (i.e., 50). Press the up or down arrow to increase or decrease the contrast of the Edge.

7.Press on/off enter button to accept the new contrast value.

8.To return to the measurement view screen, press the down arrow and press the

on/off/enter button.

053-644, Rev H

Page 36

eg4

Image 42
Contents Edge 3M The Edge Personal Noise DosimeterTable of Contents Setting-up the Edge via DMS Time history……………………………………………………………………………………………………Technical Support Stopping a studyViewing measurements on the display Replacing microphone and windscreenPhysical attributes of EdgeDock1 and EdgeDock5 INDEX…………………………………………………………………………………List of Figures List of TablesDosimetry Edge and dosimetryNoise dosimeter Illustration of how the dosimeter operates and processesRange of sound What do the standards say?Applications of the Edge Running your studies Setting-up your dosimeter parameters for your noise studiesNoise assessment steps View your results via DMSParameter setup screen for eg4 Edge ModelChecking the equipment Identifying your equipmentLED indicator Turning onLED indicators explained Turning offDiagram Display and keypad3MTMDetection Mangement Software DMS Docking station functionalityLoading Docking station components EdgeDock1 or EdgeDock5EdgeDock1 and EdgeDock5 components Powering EdgeDock1 or EdgeDock5  Powering EdgeDock1 or EdgeDock5 How to charge Charging your dosimeterMounting the docking station to a wall Storing the dosimeterUnderstanding what data is stored Noise studiesSummary data Time historyViews on the Edge LoggingDefinitions of data log values and views Views on ExplanationDosimeter views and definitions Dose WeightingViews on Explanation Edge Response LevelDescriptors Variants Lmax variants Dosimeter measurement descriptors/variantsDocking Communicating & downloading dataCustomizing set-up parameters in DMS New Hardware wizard for Edge DockDosimeter setting Dosimeter screenEdge dosimeter settings Dosimeter fields explained for the EdgeDisplay setting Edge display screenSecurity setting Edge security settingAuto Run setting Edge auto run screenFirmware update 10 Edge firmware update screenSaving and sending/setting configurations 11 Saving and sending Edge setupsOverview of running a study Quick Help List typical dosimetry sessionWindscreen and maintenance  How to attach the dosimeterPositioning the Edge on your shirt Checking battery power  Checking battery powerCalibrating  Calibrating the dosimeterResetting the dosimeter  ResetSetting contrast on dosimeter  ContrastOn/off/enter button  How to run a noise study Running studiesStarting a run Indicator Icon Explanation Indicators/Icons on the display How to stop a noise study Indicators i.e., run, stop, battery, and OLViewing measurements on the display Locking or securing your dosimeter during run Viewing measurements Locking or Securing  How to lock or secure How to unlock UnlockingDownloading data  DownloadingDMS start screen Viewing data and data finder  Selecting and viewing dataReport View Example of Edge Report in DMSPanel Layout Viewcharts/graphs Panel layout view viewing the data in charts and graphsLED indicator disable/enable Technical SupportReplacing microphone and windscreen Attaching/Detaching Attaching/Detaching the windscreen and microphone Hard Hat Attachment Kit Hard Hat Attachment kitMechanical characteristics StandardsElectrical characteristics Environmental effectsUser Interface EdgeDock1 and EdgeDock5 Physical attributes of EdgeDock1 and EdgeDock5Measurements TypesMeasurement parameters Security CommunicationsSpecial functions Auto-RunEdge Block Diagram Edge Block Diagram Appendix B Service Department and Technical Support 1 800 3M ServiceWarranty Contacting 3MGlossary of Terms Background Noise, effectsCombining Noise sources Crest FactorCriterion Level CL Dose Criterion TimeDecibel dB Exchange Rate ER also known as the Doubling RateLavg Average Level Free field setting microphoneHearing Conservation HC Hertz Hz Lep,dMaximum Level Lmax Leq Equivalent LevelLogging Measurement RangeRandom incidence microphone PEL Permissible Exposure LevelProjected Time Response TimeTWA Time Weighted Average ThresholdWeighting A, C, Z, etc Upper Limit ULWindscreen Measurement equations Index Running studies About 3M Personal Safety About Us