25
Customizing 
Dosimeter setting
Edge dosimeter settings
| 
 | 
 | 
 | 
 | 
| Dosimeter settings for the Edge | 
 | Name Field | 
 | 
| 
 | 
 | *Note: refer to Glossary for more detailed information. | 
 | 
| Heartbeat enable | 
 | When enabled (or checked), the LED will flash green while running | 
 | 
| 
 | 
 | and flash amber when stopped. | 
 | 
| Response | 
 | Either Slow Time Response or Fast Time Response is applied to | 
 | 
| 
 | 
 | the measurements. (Typically Slow time response is used in the | 
 | 
| 
 | 
 | US market while Fast is used in the EU communities.) | 
 | 
| 
 | When enabled (or checked), C minus A will be calculated. | 
 | |
| Projected Time | 
 | A selectable time parameter via the up/down arrows or type in a | 
 | 
| 
 | 
 | value. The projected time value is used to make projected dose | 
 | 
| 
 | 
 | calculation for the dose measurement. | 
 | 
| Peak Weighting | 
 | Select either A, C or Z for the peak frequency weighting. | 
 | 
| Dose indicator LED | 
 | When enabled (or checked), the LED will flash red when the dose | 
 | 
| 
 | 
 | value exceeds the designated entered value. | 
 | 
| Setup name | 
 | (Optional field) A setup name may be entered into the Meter 1 | 
 | 
| 
 | 
 | and/or Meter 2 fields. When set, it appears on the top of Edge’s | 
 | 
| 
 | 
 | display (example setup names: OHSHA, MSHA, EU directive). | 
 | 
| Threshold enable | 
 | When the checkbox is selected and a value is entered for meter 1 | 
 | 
| 
 | 
 | and/or meter 2, the threshold is used to calculate Lavg, TWA, and | 
 | 
| 
 | 
 | Dose measurements. | 
 | 
| Upper Limit | 
 | Select a UL dB level by the up/down arrows or by typing in a value. | 
 | 
| 
 | 
 | (Note: The upper limit (UL) is a measurement setting in which the | 
 | 
| 
 | 
 | user selects a specific decibel level. The UL value records the | 
 | 
| 
 | 
 | amount of time that the Sound level is at or greater than the | 
 | 
| 
 | 
 | entered level.) | 
 | 
| RMS Weighting | 
 | Select either A or C RMS Weighting. (Note: The RMS (root mean | 
 | 
| 
 | 
 | square) are the results from the Edge with the response time and | 
 | 
| 
 | 
 | weighting settings taken into account.) | 
 | 
| Criterion Level | 
 | Select a Criterion level via the up/down arrow or by typing in a | 
 | 
| 
 | 
 | value. (Note: Criterion level is the average SPL that will result in | 
 | 
| 
 | 
 | 100% dose over the criterion time and is typically set by a | 
 | 
| 
 | 
 | regulatory agency such as OSHA, EU directive.) | 
 | 
| Exchange Rate | 
 | Enter in (or select with up/down arrows) 3, 4, or 5 dB. ER refers to | 
 | 
| 
 | 
 | how sound energy is averaged over time. (OSHA uses a 5 dB | 
 | 
| 
 | 
 | doubling rate while most of the world uses 3 dB.) | 
 | 
Table 1-2:  Dosimeter fields explained for the Edge
| Page 25 | eg4 |