Operating QUESTempº 44/46

Print

Print

Allows printing to a parallel or serial printer or to a computer. The QuesTempº 44/46 will recognize the cable plugged in and configure itself for serial or parallel. If no cable is plugged in, it will default to serial. Press I/O Enter key to begin printing. Press Run/Stop key to return to the menu.

NOTE: if you wish to stop the printing, press I/O enter key until you return to the main menu. When the printer has stopped printing, remove the cable from the printer to the instrument.

Reset

Resetting enables you to clear the logged data from memory. Press the I/O Enter key to enter the Reset mode. Clear the memory by holding down the I/O Enter key while the display counts down from three.

Run

The run mode begins a session in memory and logs the data.

1.Begin a session by pressing the Run/Stop key from the view mode (or measurement view). An asterisk in the lower right corner indicates the run mode.

To toggle through the views, press the up or down arrow.

WBGTi

66.7º F

1

 

WBGTo

68.6º F

 

 

 

 

 

 

 

 

 

Figure 1-8 Run mode indicator

Run indicator

2.End the session by pressing the Run/Stop key again. (The session will stop recording when the asterisk is no longer displayed.)

NOTE: If the logging memory is full or if there are no sensors plugged into the unit, attempting to enter the Run mode will result in an error message. If the memory capacity is exceeded, the asterisk in the lower right corner of the display will turn into an “F” and the memory remaining screen will show “0.0”.

Image 17 Contents
Market Leader Contents Displayed Items Iii List of Tables List of FiguresQUESTempº Models IntroductionKeypad Operation Getting StartedUp and Running overview Up Arrow key Enter keyDown Arrow key Run Stop keyPlacing the QUESTempº 44/46 on the job site Turning on/off and basic operationAbout the sensor bar SensorsGlobe Thermometer Waterless Wetbulb Sensor and Relative Humidity SensorRemote, Sensors 2 Tri-sensor weighted averageComputing Waterless wetbulb and wind speed Waterless WetbulbStay Times/Rest Times QTº46 only WetBulb Globe TemperatureHeat Index/Humidex Airflow Thermal Comfort View SetupOperating QUESTempº 44/46 Index Setting Notes Setup for Waterless Wetbulb measurement airflow rateRun ResetPrint Flow 0.3m/s  Displayed ItemsStay Time DateData Logging Data LoggingPerforming an electronic sensor check Electronic sensor checkRH Sensor Alignment Sensor AlignmentLow RH Alignment screen Adjust % to 33% using up/down arrow Serial PrintingPrintingPrinting Heat Stress Report ParallelAirflow Functionality Airflow functionalityOperating sequence Operational CheckData Logging Airflow Batteries for Air ProbeApproved 9-Volt Batteries Power optionsVolt Alkaline Battery Replacement 29 NiMH rechargeable battery NiMH Battery PackAppendix a Appendix a SpecificationsCharge Time NiMH Battery Pack Appendix B Appendix B Heat Exposure TablesUnited States Navy Acgih Clothing CorrectionsUnited States Navy Flag Conditions for U.S. Navy/Marine Corp. Ashore Electrical power research institute EpriAppendix C Appendix C AccessoriesAir Probe Accessories Appendix D 3MTM Detection Management Software DMS Communication setupQUESTempo 44/46 Appendix D31 QT⁰44/46 downloading data 32 Downloading QT⁰44/46 files Downloading data with the QT⁰44/46Selecting a session/study Charts and Graphs in Panel layout view PLVViewing Data in DMS 34 QT⁰44/46 data in panel layout view Remember setting Reports and Printing Inserting on-site testing imagesWarranty 3M ServiceService Department and Technical Support 1 800 CalibrationAbout Us About 3M Personal Safety