3M 44, 46 Operational Check, Operating sequence, Data Logging Airflow, Batteries for Air Probe

Page 27

22

Operational Check

Operational

Check

Operating sequence

 

Operating sequence

1.Turn Flow On in the setup menu of the QUESTemp°46.

2.Plug the Air Probe into the port labeled Flow on the side of the QUESTemp° 46.

3.Turn on the Air Probe ON and make sure the green lamp is lit.

In the View or Run modes, airflow is displayed on the fifth screen.

Data Logging Airflow

To data log airflow in the QUESTemp° 46, the following two conditions must be met.

1.First, turn Flow On in the setup menu.

2.Second, make sure that a temperature sensor bar is connected to the Sensor 1 location (top) of the QUESTemp°46. Flow prints out with the Sensor 1 data therefore airflow data will only be reported if a sensor bar is plugged in. Airflow is recorded during the run mode at the interval the QUESTemp° 46 is setup for.

Batteries for Air Probe

The Air Probe uses a single NiMH Black&DeckerVersaPakGold battery. Typical operating time of the battery is between 6 and 8 hours.

To change the battery, push in and twist, counterclockwise, the cap on the bottom of the Air Probe. Pull out the battery. Insert a fully charged battery and replace the cap.

To recharge the batteries, set the battery into the VersaPackcharger. The supplied charger accepts one or two batteries. A full charge takes 9 hours. An indicator light shows that the battery is properly charging and it will remain on as long as the battery is in the charger. Continuous charging is not a safety concern.

Operational Check

A verification module, 3M model 053-923, may be used to check the operation of the QUESTemp. Remove the top sensor bar and plug the verification module into the top of the unit. With the QUESTemp set to read in degrees Celsius, verify that the displayed readings match those printed on the module within +/-0.5°C.

If the readings are not within the +/-0.5°C tolerance, then have the unit serviced and calibrated.

Image 27
Contents Market Leader Contents Displayed Items Iii List of Tables List of FiguresQUESTempº Models IntroductionGetting Started Up and Running overviewKeypad Operation Run Stop key Enter keyUp Arrow key Down Arrow keyPlacing the QUESTempº 44/46 on the job site Turning on/off and basic operationWaterless Wetbulb Sensor and Relative Humidity Sensor SensorsAbout the sensor bar Globe ThermometerRemote, Sensors 2 Tri-sensor weighted averageComputing Waterless wetbulb and wind speed Waterless WetbulbAirflow WetBulb Globe TemperatureStay Times/Rest Times QTº46 only Heat Index/HumidexThermal Comfort Setup Operating QUESTempº 44/46View Index Setting Notes Setup for Waterless Wetbulb measurement airflow rateReset PrintRun Flow 0.3m/s  Displayed ItemsStay Time DateData Logging Data LoggingPerforming an electronic sensor check Electronic sensor checkRH Sensor Alignment Sensor AlignmentLow RH Alignment screen Adjust % to 33% using up/down arrow Printing PrintingPrintingSerial Heat Stress Report ParallelAirflow Functionality Airflow functionalityBatteries for Air Probe Operational CheckOperating sequence Data Logging AirflowPower options Volt Alkaline Battery ReplacementApproved 9-Volt Batteries 29 NiMH rechargeable battery NiMH Battery PackAppendix a Appendix a SpecificationsCharge Time NiMH Battery Pack Appendix B Appendix B Heat Exposure TablesUnited States Navy Acgih Clothing CorrectionsUnited States Navy Flag Conditions for U.S. Navy/Marine Corp. Ashore Electrical power research institute EpriAppendix C Accessories Air Probe AccessoriesAppendix C Appendix D Communication setupAppendix D 3MTM Detection Management Software DMS QUESTempo 44/4631 QT⁰44/46 downloading data 32 Downloading QT⁰44/46 files Downloading data with the QT⁰44/46Charts and Graphs in Panel layout view PLV Viewing Data in DMSSelecting a session/study 34 QT⁰44/46 data in panel layout view Remember setting Reports and Printing Inserting on-site testing imagesCalibration 3M ServiceWarranty Service Department and Technical Support 1 800About Us About 3M Personal Safety