3M 46, 44 user manual Power options, Volt Alkaline Battery Replacement, Approved 9-Volt Batteries

Page 28

23

Power options

9-Volt Alkaline Battery Replacement

Power options

There are 3 options for powering the QUESTempº 46: a 9-volt alkaline battery, a NiMH (Nickel Metal Hydride) rechargeable 6-cell battery pack, and an AC adapter. A door on the back of the unit allows the user access to the 9-volt battery. The rechargeable battery pack is located inside of the unit. If the rechargeable battery pack ever needs to be replaced, it can be accessed by removing the screws from the bottom panel of the unit.

The 2-position switch located in the battery compartment must be set by the user if the power supply method is changed. The up position is for the 9-volt battery. The down position allows for either the AC adapter or the rechargeable batteries. The AC adapter will trickle charge the rechargeable batteries if they are in place or it will simply allow for line power operation of the unit.

Figure 1-28: 9-volt battery

9-Volt Alkaline Battery Replacement

WARNING: Replace batteries only in a non-hazardous environment.

The 9-volt battery should be replaced or the NiMH battery pack should be recharged when the voltage drops below 6.4 volts. The battery voltage is displayed when the instrument is turned on. While turned on, the battery voltage can be displayed at any time by pressing the up or down arrow keys to move through the display until the battery voltage screen appears. If, while operating, the battery voltage drops below

6.4volts, the display will automatically switch to the display showing the battery voltage along with a low battery message. After a low battery occurs, the unit will continue to operate for approximately 8 hours. When the battery voltage falls to 6.2 volts or below, the unit will automatically turn off.

Replace only with an approved 9-volt alkaline battery.

Approved 9-Volt Batteries

Eveready: Energizer 522, EN22, 6LR61

Duracell: MN1604

Panasonic: 6LR61, 6AM6X

Rayovac: A1604

UltraLife: U9V

Image 28
Contents Market Leader Contents Displayed Items Iii List of Figures List of TablesIntroduction QUESTempº ModelsUp and Running overview Getting StartedKeypad Operation Enter key Up Arrow keyDown Arrow key Run Stop keyTurning on/off and basic operation Placing the QUESTempº 44/46 on the job siteSensors About the sensor barGlobe Thermometer Waterless Wetbulb Sensor and Relative Humidity SensorTri-sensor weighted average Remote, Sensors 2Waterless Wetbulb Computing Waterless wetbulb and wind speedWetBulb Globe Temperature Stay Times/Rest Times QTº46 onlyHeat Index/Humidex AirflowThermal Comfort Operating QUESTempº 44/46 SetupView Setup for Waterless Wetbulb measurement airflow rate Index Setting NotesPrint ResetRun Displayed Items Flow 0.3m/s Date Stay TimeData Logging Data LoggingElectronic sensor check Performing an electronic sensor checkSensor Alignment RH Sensor AlignmentLow RH Alignment screen Adjust % to 33% using up/down arrow PrintingPrinting PrintingSerial Parallel Heat Stress ReportAirflow functionality Airflow FunctionalityOperational Check Operating sequenceData Logging Airflow Batteries for Air ProbeVolt Alkaline Battery Replacement Power optionsApproved 9-Volt Batteries NiMH Battery Pack 29 NiMH rechargeable batteryAppendix a Specifications Appendix aCharge Time NiMH Battery Pack Appendix B Heat Exposure Tables Appendix BAcgih Clothing Corrections United States NavyUnited States Navy Electrical power research institute Epri Flag Conditions for U.S. Navy/Marine Corp. AshoreAir Probe Accessories Appendix C AccessoriesAppendix C Communication setup Appendix D 3MTM Detection Management Software DMSQUESTempo 44/46 Appendix D31 QT⁰44/46 downloading data Downloading data with the QT⁰44/46 32 Downloading QT⁰44/46 filesViewing Data in DMS Charts and Graphs in Panel layout view PLVSelecting a session/study 34 QT⁰44/46 data in panel layout view Remember setting Inserting on-site testing images Reports and Printing3M Service WarrantyService Department and Technical Support 1 800 CalibrationAbout 3M Personal Safety About Us