3M 44, 46 user manual Electronic sensor check, Performing an electronic sensor check

Page 21

Electronic Sensor

check

16

Electronic sensor check

Performing an electronic sensor check

Electronic sensor check

A verification module, Quest model 053-923, may be used to check the operation of the QUESTemp’s wet bulb, dry bulb, and globe. The purpose is to verify that the electronic components are within a specific range with known values and a known source. The temperature tolerances should be within +/-0.5oC.

Example of Verification module

NOTE: If the sensors are outside of the tolerances, this indicates the sensor alignment should be serviced off-site for calibration.

Performing an electronic sensor check

1. Ensure the instrument is reading in Celsius prior to your electronic sensor check.

To change the temperature setting, select Setup from the main menu. Then,

either Fahrenheit or Celsius will appear on the screen. When selected, press I/O enter key to switch between settings. For more information, please refer to Operating, page 10.

2.Remove the top sensor bar, place to the side, and plug in the verification module into the center pins of the sensor housing.

3.Verify the measurement readings on the screen are within +/- 0.5°C tolerance to the

readings printed on the verification module label. Example below:

a. Wet Bulb (WB) – 11.1°C b. Dry Bulb (DB) - 45.3°C c. Globe (G) – 69.2°C

NOTE: Relative humidity (RH) is not valid on this instrument.

4.Once completed, remove verification module and place sensor bar back on the instrument. (Tighten down the two bolts.) Change the Celsius reading back to Fahrenheit. (Refer to step 1 a-b.)

Image 21 Contents
Market Leader Contents Displayed Items Iii List of Tables List of FiguresQUESTempº Models IntroductionGetting Started Up and Running overviewKeypad Operation Up Arrow key Enter keyDown Arrow key Run Stop keyPlacing the QUESTempº 44/46 on the job site Turning on/off and basic operationAbout the sensor bar SensorsGlobe Thermometer Waterless Wetbulb Sensor and Relative Humidity SensorRemote, Sensors 2 Tri-sensor weighted averageComputing Waterless wetbulb and wind speed Waterless WetbulbStay Times/Rest Times QTº46 only WetBulb Globe TemperatureHeat Index/Humidex AirflowThermal Comfort Setup Operating QUESTempº 44/46View Index Setting Notes Setup for Waterless Wetbulb measurement airflow rateReset PrintRun Flow 0.3m/s  Displayed ItemsStay Time DateData Logging Data LoggingPerforming an electronic sensor check Electronic sensor checkRH Sensor Alignment Sensor AlignmentLow RH Alignment screen Adjust % to 33% using up/down arrow Printing PrintingPrintingSerial Heat Stress Report ParallelAirflow Functionality Airflow functionalityOperating sequence Operational CheckData Logging Airflow Batteries for Air ProbePower options Volt Alkaline Battery ReplacementApproved 9-Volt Batteries 29 NiMH rechargeable battery NiMH Battery PackAppendix a Appendix a SpecificationsCharge Time NiMH Battery Pack Appendix B Appendix B Heat Exposure TablesUnited States Navy Acgih Clothing CorrectionsUnited States Navy Flag Conditions for U.S. Navy/Marine Corp. Ashore Electrical power research institute EpriAppendix C Accessories Air Probe AccessoriesAppendix C Appendix D 3MTM Detection Management Software DMS Communication setupQUESTempo 44/46 Appendix D31 QT⁰44/46 downloading data 32 Downloading QT⁰44/46 files Downloading data with the QT⁰44/46Charts and Graphs in Panel layout view PLV Viewing Data in DMSSelecting a session/study 34 QT⁰44/46 data in panel layout view Remember setting Reports and Printing Inserting on-site testing imagesWarranty 3M ServiceService Department and Technical Support 1 800 CalibrationAbout Us About 3M Personal Safety