VXI VT1538A user manual OUTPutPOLarity?, Send with VXIplug&play Function hpe14XXcmdReal64Q

Page 50

Note Because the E1538 rounds <level> to the nearest multiple of 0.375, the returned value can be different from the value sent.

INP:THR:LEV? returns a numeric value between -46 and +46. The C-SCPI type is float32.

Related Commands: INPut:POLarity, INP:THR:LEVel

*RST Condition: INP:THR:LEV = 1.875

Send with VXIplug&play Function: hpe14XX_cmdReal64_Q(...)

Usage To query the threshold level on the second channel at SCP position 4 send:

INP:THR:LEV? (@133)

query 2nd chan on SCP pos. 4

enter statement here

 

OUTPut:POLarity

OUTPut:POLarity <select>,(@<ch_list>) sets the polarity on digital output channels in <ch_list>.

Parameters

 

 

 

 

 

 

 

 

 

 

 

Parameter

 

Parameter

Range of

Default

 

 

 

Name

 

 

Type

Values

Units

 

 

 

 

 

 

 

 

 

 

 

 

select

 

 

discrete (string)

NORMal INVerted

none

 

 

 

 

 

 

 

 

 

 

 

 

ch_list

 

 

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments

If the channels specified do not support this function, an error will be

 

 

 

 

 

generated.

 

 

 

Related Commands: INPut:POLarity, OUTPut:POLarity?

*RST Condition: OUTP:POL NORM for all digital channels

Send with VXIplug&play Function: hpe14XX_cmd(...)

Usage OUTP:POL INV,(@144)

invert output logic sense on

 

channel 44

OUTPut:POLarity?

OUTPut:POLarity? (@<channel>) returns the polarity on the digital output channel in <channel>.

46 Agilent E1538A Enhanced Frequency/Totalize/PWM SCP

Image 50
Contents VT1538A VXI Technology, Inc Introduction Support Resources User’s and Scpi Programming Manual Page Agilent E1538A Enhanced Frequency/Totalize/PWM SCP About this ManualIntroduction Identifying the Plug-on Important Setting Configuration Switches Switch Location and Example SettingsInstallation Connecting To The Terminal ModuleRecommended Signal Connections E1419A Option 11 Terminal Module ConnectionsInput and Output Characteristics Recommended Connections and Voltage LimitsVRS Mode Input Operation SCP Channels 0 & 1 only E1538A Input/Output CharacteristicsProgramming With Scpi Commands Shows the VRS mode input characteristicsConfiguring I/O Direction Setting the Input Threshold LevelDetermining the Input Threshold Level Inppol NormReading Static Digital State To set channels 40 through 43 to input digital statesTotalize Positive or Negative Edge State Changes Init SENSDATACVT? @44About Period Frequency Measurements SENSDATACVT? @45 SENSDATACVT? @45 Agilent E1538A Enhanced Frequency/Totalize/PWM SCP Agilent E1538A Enhanced Frequency/Totalize/PWM SCP Sense Quadrature Position Sensfuncquad 0,@148,149RPS Example of Rotational Velocity Sense Sense Rotational VelocityControlling Output Polarity Output Static Digital StateVariable Width Pulse Per Trigger Variable Width Pulse Train PWM Variable Frequency Fixed Width Pulse Train FM Variable Frequency Square-Wave Pulse Train Rotationally Positioned Pulse Output Rotational Pulse Command Usage Algupdate Init Example of variable position, preset width RST Sensfuncrvel 12,MISS,@140Variable Position, Width Preset by Time Example of variable width by angle, preset position Example of variable width by time, preset position Fixed Position, Variable Width by AngleALGSCALAR’ALG1’,’Width2’,.010 Stepper Motor Control Use the command Stepping mode valuesSensfuncvolt @100 Full Step Mode, Full and Half Speed, 4-Channel Page RST and *TST important Scpi Command Reference Command Syntax For newer Units onlyCommand Syntax For newer Units only Discussed Parameters INPutPOLarityINPutPOLarity? CommentsINPutTHResholdLEVel? Level can be set to a resolutionINPutTHResholdLEVel Send with VXIplug&play Function hpe14XXcmdReal64Q OUTPutPOLarityOUTPutPOLarity? SENSeFREQuencyAPERture SENSeFREQuencyLIMitLOWer Related Commands SensfreqaperSENSeFREQuencyAPERture? Init SENSeFUNCtionCONDition Related Commands INPutPOLaritySENSeFREQuencyLIMitLOWer? SENSeFUNCtionFREQuency SENSeFUNCtionPERiodSENSeFUNCtionPWIDth SENSeFUNCtionQUADratureSENSeFUNCtionRVELocity RPS SENSeFUNCtionTOTalize SENSePERiodAPERture SENSePERiodAPERture? SENSePERiodLIMitUPPerInit SENSePERiodLIMitUPPer? SENSePERiodMODESENSePERiodNPERiods Related Commands SENSePERiodMODESENSePERiodMODE? Send with VXIplug&play Function hpe14XXcmdInt32Q SENSePERiodNPERiods?SENSePERiodRANGeUPPer RST Condition is Send with VXIplug&play Function hpe14XXcmd SENSeTOTalizeRESetMODESENSePERiodRANGeUPPer? SENSeTOTalizeRESetMODE? SOURceFMSTATeSOURceFMSTATe? Send with VXIplug&play Function hpe14XXcmdInt16QSOURceFUNCtionRPULse SOURceFUNCtionSHAPeCONDition SOURceFUNCtionSHAPePULSeComments Send with VXIplug&play Function hpe14XXcmd SOURceFUNCtionSHAPeSQUareSOURceFUNCtionSTEPper Relationship of minvel, and maxvel SOURcePULMSTATe Usage See Stepper Motor Control onSOURcePULMSTATe? SOURcePULSePERiodSOURcePULSePERiod? SOURcePULSeWIDTh33,554,430 to 33,554,430 SOURcePULSeWIDTh?SOURceRPULsePOSitionANGLe Usage SOURceRPULsePOSitionANGLe?SOURRPULseVARType SOURRPULseVARType? SOURceRPULseWIDThANGLe00000787 to SOURceRPULseWIDThANGLe?SOURceRPULseWIDThTIME SOURceRPULseWIDThTIME? Specifications General SpecificationsFrequency Measurement Pulse Width Measure Rotational Pulse Source