Operational Theory

 

 

 

Operational Theory

2.1 Introduction

The analyzer is composed of three subsystems:

1.Analysis unit with zirconium oxide sensor(s)

2.Control unit with signal processing, display and controls

3.Interface unit, supplying signals between 1 and 2.

The analysis unit is designed to accept the sample gas and direct it to the sensitive surface of the zirconium oxide sensor.

The control unit processes the sensor output and translates it into an electrical concentration, which supplies the customer with range, alarm outputs, and a meter readout giving the trace oxygen level. It contains a microcontroller that manages all signal processing, input/output, and display functions for the analyzer.

The interface unit receives power from the control unit, and creates precision heater and bias voltages for the zirconium oxide sensor.

2.2 Zirconium Oxide Sensors

2.2.1 Principles of Operation

Teledyne’s zirconium oxide sensor is a miniature solid state

sensor. It uses a stabilized zirconia disc as an electrolyte with a sensing electrode (the cathode) and a counter electrode (the anode) on each side of the disc. The zirconia is heated to a temperature of about 500°C via a built-in heater. By applying a suitable reference voltage across the cathode and anode, the heated zirconia acts as an electrolyte which is capable of moving oxygen ions within its crystalline structure. This allows for the reduction of oxygen molecules reaching the cathode and the generation of a current equivalent to the amount of oxygen in a sample measured in parts per million (ppm).

Teledyne Analytical Instruments

5

Image 15
Contents Model 3300ZB Model 3300ZB Safety Messages Model 3300ZB Table of Contents Maintenance Front Panel List of FiguresCombustible GAS Usage This page intentionally left blank Page Main Features of the Analyzer IntroductionOverview Front Panel Front Panel Description∙ Measurement Parameters Readout Rear Panel Description Power Connection Principles of Operation Operational TheoryIntroduction Zirconium Oxide SensorsElectronics Effect of PressureGeneral Block Diagram of the Signal Processing Electronics Signal ProcessingOperational Theory Model 3300ZB Unpacking the Analyzer InstallationControl Unit Installation Location and MountingElectrical Connections Rear Panel Electrical Connectors Contact ID for Failsafe Relay Operation Gas Connections Piping Diagram Installation ChecklistOperation Front Panel Controls and IndicatorsSetting the Analysis Ranges Using the Function and Data Entry ButtonsHI Range Setting the Alarm SetpointsHI Alarm LO AlarmSensor Fail Alarm Selecting a Fixed Range or AutorangingCalibration LED Display on StartUp This page intentionally left blank Replacing the Fuse MaintenanceWhen to Replace a Sensor Sensor Installation or ReplacementZrO2 Sensor and Cell Block This page intentionally left blank Appendix SpecificationsTeledyne Analytical Instruments Spare Parts ListDrawing List