Teledyne 3300ZB operating instructions Rear Panel Electrical Connectors

Page 21

Trace Oxygen Analyzer

Installation

 

 

 

Figure 3-1: Rear Panel Electrical Connectors

Analog Outputs: There are three DC output signal connectors with screw terminals on the panel. There are two wires per output with the polarity noted. See Figure 3-1. The outputs are:

0–10 V Concentration: Voltage is proportional to oxygen reading, from 0 V at 0 ppm oxygen to 10 V at full scale ppm oxygen. (Full scale = 100% of programmed range).

0–10 V Range ID:

3.33 V = Low Range, 6.66 V = High Range

4–20 mA % Range:

Current is proportional to oxygen reading, from 4

 

mA at 0 ppm oxygen to 20 mA at full scale ppm

 

oxygen. (Full scale = 100% of programmed

 

range).

Alarm Relays: The three alarm circuit connectors are screw terminals for making connections to internal alarm relay contacts. There is one set of contacts for each type of alarm. Contacts are Form C, with normally open and normally closed contact connections capable of switching up to 0.5A at 125 VAC into a resistive load.

The alarm relay circuits are designed for failsafe operation, meaning the relays are energized during normal operation. If power fails the relays de-energize (alarms activated).

The contact connections are indicated diagrammatically on the rear panel as normally closed (NC), common (C), and normally open (NO). Figure 3-2 explains how these act in failsafe operation.

Teledyne Analytical Instruments

11

Image 21
Contents Model 3300ZB Model 3300ZB Safety Messages Model 3300ZB Table of Contents Maintenance Front Panel List of FiguresCombustible GAS Usage This page intentionally left blank Page Main Features of the Analyzer IntroductionOverview Front Panel Front Panel Description∙ Measurement Parameters Readout Rear Panel Description Power Connection Introduction Operational TheoryZirconium Oxide Sensors Principles of OperationElectronics Effect of PressureGeneral Block Diagram of the Signal Processing Electronics Signal ProcessingOperational Theory Model 3300ZB Unpacking the Analyzer InstallationControl Unit Installation Location and MountingElectrical Connections Rear Panel Electrical Connectors Contact ID for Failsafe Relay Operation Gas Connections Piping Diagram Installation ChecklistOperation Front Panel Controls and IndicatorsSetting the Analysis Ranges Using the Function and Data Entry ButtonsHI Alarm Setting the Alarm SetpointsLO Alarm HI RangeSensor Fail Alarm Selecting a Fixed Range or AutorangingCalibration LED Display on StartUp This page intentionally left blank Replacing the Fuse MaintenanceWhen to Replace a Sensor Sensor Installation or ReplacementZrO2 Sensor and Cell Block This page intentionally left blank Appendix SpecificationsTeledyne Analytical Instruments Spare Parts ListDrawing List