Teledyne 3300ZB operating instructions LED Display on StartUp

Page 29

Trace Oxygen Analyzer

Operation

 

 

 

3.As the display is blinking, once the readout appears stable, press either the Δ or ∇ button to adjust the reading to coincide with the oxygen concentration of the span gas. Once set, the unit will enter Analyze Mode after approximately five seconds after the last key press.

Note: If no key presses are made during the 60 second period when the display is blinking, the analyzer will accept the display value as the true span concentration and automatically leave the span mode. The display will cease blinking.

The unit is now calibrated.

If the output of the sensor is outside of the expected range, this may be caused by:

a)Bad electrical connection between the controler and sensor.

b)Improperly analyzed or entered calibration gas value

c)Electronics failure

The unit will not accept the attempted calibration and will flash ‘0000’ on the LED display unit a valid calibration is performed.

4.7 LED Display on StartUp

When the unit is powered, it will initially display “8.8.8.8” for a short period of time. This is to confirm all LED segments are functioning. Afterward, the LED display will initiate a delay countdown, allowing the sensor’s heater to begin charging and capture a zero reading. Flowing “zero” gas or a sensor is not needed for this instrument adjustment.

Teledyne Analytical Instruments

19

Image 29
Contents Model 3300ZB Model 3300ZB Safety Messages Model 3300ZB Table of Contents Maintenance Front Panel List of FiguresCombustible GAS Usage This page intentionally left blank Page Overview Main Features of the AnalyzerIntroduction Front Panel Front Panel Description∙ Measurement Parameters Readout Rear Panel Description Power Connection Introduction Operational TheoryZirconium Oxide Sensors Principles of OperationGeneral ElectronicsEffect of Pressure Block Diagram of the Signal Processing Electronics Signal ProcessingOperational Theory Model 3300ZB Unpacking the Analyzer InstallationElectrical Connections Control Unit InstallationLocation and Mounting Rear Panel Electrical Connectors Contact ID for Failsafe Relay Operation Gas Connections Piping Diagram Installation ChecklistOperation Front Panel Controls and IndicatorsSetting the Analysis Ranges Using the Function and Data Entry ButtonsHI Alarm Setting the Alarm SetpointsLO Alarm HI RangeCalibration Sensor Fail AlarmSelecting a Fixed Range or Autoranging LED Display on StartUp This page intentionally left blank Replacing the Fuse MaintenanceWhen to Replace a Sensor Sensor Installation or ReplacementZrO2 Sensor and Cell Block This page intentionally left blank Appendix SpecificationsTeledyne Analytical Instruments Spare Parts ListDrawing List