Omron Healthcare HEM-907XL manual Blood Pressure in Average Mode

Page 12

HOW TO MEASURE

BLOOD PRESSURE (IN AVERAGE MODE)

1. Push the ON/OFF (power) Button to turn on the power.

2. Set the MODE Selector to "AVG".

The factory-set values are set as follows:

• Number of measurements: 2

• Waiting time until the 1st measurement: 0 sec.

• Interval: 1 min.

To change these factory-set values, refer to

Page 13.

3. Set the P-SET (inflation level setting) Knob to "AUTO" or the target pressure value.

4.Measure the patient’s arm size and wrap appropriate cuff over the patient's arm. (Refer to Pages 14 and 15.)

HOW TO MEASURE

BLOOD PRESSURE (IN AVERAGE MODE)

5. Push the START Button to start the measurement.

After the pre-select waiting time, the unit takes the 1st measurement.

After displaying the results of 1st measurement, subsequent measures occur automatically at the specified intervals.

• For setting the number of measurements, the waiting time until the 1st measurement, and the interval, refer to Page 13.

• If you want to stop measurement, push the STOP Button. The unit will rapidly deflate.

If an error occurs during measurement, the monitor will automatically start measurement again. If a second error occurs, measurement will automatically stop.

Do not push the START Button without wrapping the cuff.

6. The measurement results are displayed.

After all the measurements are finished, average values will be displayed.

Each time the DEFLATION (deflation control) / Measurement Result Display Switch Button is pushed, the measurement results for each reading and the average value will be displayed.

While the battery is in use, the monitor will turn off after five minutes of inactivity and the display (measurement results) will disappear. (Automatic Power Off)

7. Push the ON/OFF (power) Button to turn off the power.

If the monitor determines that the pressure value is not correct, an error display appears (Er1 to 9). In this case, refer to Page 28 and start the measurement again.

22

23

Image 12
Contents How to Measure Blood Pressure SpecificationsBe Sure to Read This Section Know Your UnitExemptions General advice Features of the Product Maintenance and inspectionTroubleshooting Components of the Product IntelliSense Blood Pressure Monitor, Model HEM-907XLNot included Cuffs without bladder Bladders Included and also available separatelyFeatures and Functions Names of the PartsFunctions setting Inflation level setting P-SETAVG Function setting Non-display functionHOW to Apply the Cuff Select the cuff according to the arm sizeHOW to USE the Power Source How to use the AC adapterInstallation and Replacement of Battery Pack List of Measurement Modes HOW to Measure Blood Pressure in Single ModeBlood Pressure in Average Mode Blood Pressure in Manual Mode Unit does not automatically deflate in the Manu ModeHOW to Check Pressure Accuracy in Check ModeHow to check What you need to prepareList of Error Codes TroubleshootingWrap the cuff correctly, Five Year Limited Warranty SpecificationsFor Customer Service Call Toll Free