NOTES ON SAFETY

Maintenance and inspection

1.Check the unit operation on regular basis.

2.If this unit has not been used for more than three months, be sure to check that this unit operates normally and safely before use.

Troubleshooting

If device error 9 (Er9) occurs, take the following procedure promptly:

(1)Remove the cuff from the patient's arm.

(2)Turn off the power of the unit and unplug the AC adapter from the electric outlet.

(3)Display "Out of use" on this unit so that it cannot be used.

(4)Contact Omron for repair service (1-877-216-1336).

FEATURES OF THE PRODUCT

OMRON IntelliSense™ Blood Pressure unit, Model HEM-907XL is developed to measure blood pressure and pulse rate accurately and simply in a doctor's office, examination room,

or patient bedside.

One-button operation

Simply wrap the cuff and push the START Button. Blood pressure and pulse rate are automatically measured by the oscillometric method.

• Automatic pressure setting

When the P-SET (Pressure Setting) Knob is set to "AUTO," the unit will automatically inflate the cuff to the optimal pressure according to each patient's blood pressure. Pre-setting inflation level is not necessary.

• Noiseless operation

This unit operates so quietly that it can be used in the patient room at night.

• Average Mode (AVG Mode)

In the AVG Mode, this unit will automatically measure for two or three times. The average of systolic and diastolic blood pressures and pulse rate are displayed. Each measurement can also be shown individually. The number of measurements, waiting time before first measurement, and the interval can be changed.

• Auscultation Mode (MANU Mode)

You can measure auscultatory blood pressure by using a stethoscope, with automatic cuff inflation and deflation by this unit. Because the cuff pressures during deflation are displayed digitally and synchronized with the heart beat, they can be read with accuracy. After taking systolic reading you can accelerate cuff deflation to shorten measurement time.

• Large and easy to read display

Large and easy to read figures are displayed on the LCD display.

The IntelliSenseTM Monitor inflates the cuff to the ideal level with each use. No adjustments are required by the user to select an inflation level. This is especially convenient for hypertensive users and for people with certain arrhythmia or heart disorders, because their blood pressure is likely to fluctuate. The advantage is Personalized Inflation for maximum comfort.

6

7

Image 4
Contents How to Measure Blood Pressure SpecificationsBe Sure to Read This Section Know Your UnitExemptions General advice Maintenance and inspection Features of the ProductTroubleshooting Components of the Product IntelliSense Blood Pressure Monitor, Model HEM-907XLNot included Cuffs without bladder Bladders Included and also available separatelyFeatures and Functions Names of the PartsFunctions setting Inflation level setting P-SETAVG Function setting Non-display functionHOW to Apply the Cuff Select the cuff according to the arm sizeHow to use the AC adapter HOW to USE the Power SourceInstallation and Replacement of Battery Pack List of Measurement Modes HOW to Measure Blood Pressure in Single ModeBlood Pressure in Average Mode Blood Pressure in Manual Mode Unit does not automatically deflate in the Manu ModeHOW to Check Pressure Accuracy in Check ModeHow to check What you need to prepareTroubleshooting List of Error CodesWrap the cuff correctly, Five Year Limited Warranty SpecificationsFor Customer Service Call Toll Free