NOTES ON SAFETY
Maintenance and inspection
1.Check the unit operation on regular basis.
2.If this unit has not been used for more than three months, be sure to check that this unit operates normally and safely before use.
Troubleshooting
If device error 9 (Er9) occurs, take the following procedure promptly:
(1)Remove the cuff from the patient's arm.
(2)Turn off the power of the unit and unplug the AC adapter from the electric outlet.
(3)Display "Out of use" on this unit so that it cannot be used.
(4)Contact Omron for repair service
FEATURES OF THE PRODUCT
OMRON IntelliSense™ Blood Pressure unit, Model
or patient bedside.
• One-button operation
Simply wrap the cuff and push the START Button. Blood pressure and pulse rate are automatically measured by the oscillometric method.
• Automatic pressure setting
When the
• Noiseless operation
This unit operates so quietly that it can be used in the patient room at night.
• Average Mode (AVG Mode)
In the AVG Mode, this unit will automatically measure for two or three times. The average of systolic and diastolic blood pressures and pulse rate are displayed. Each measurement can also be shown individually. The number of measurements, waiting time before first measurement, and the interval can be changed.
• Auscultation Mode (MANU Mode)
You can measure auscultatory blood pressure by using a stethoscope, with automatic cuff inflation and deflation by this unit. Because the cuff pressures during deflation are displayed digitally and synchronized with the heart beat, they can be read with accuracy. After taking systolic reading you can accelerate cuff deflation to shorten measurement time.
• Large and easy to read display
Large and easy to read figures are displayed on the LCD display.
The IntelliSenseTM Monitor inflates the cuff to the ideal level with each use. No adjustments are required by the user to select an inflation level. This is especially convenient for hypertensive users and for people with certain arrhythmia or heart disorders, because their blood pressure is likely to fluctuate. The advantage is Personalized Inflation for maximum comfort.