FEATURES AND FUNCTIONS
Functions setting
(1) Inflation level setting P-SET
AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected. The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure (approximately
Manual level setting: Inflation level can be set manually between 100 and 280 mmHg. Set the level to 30 to 40 mmHg higher than the expected systolic pressure.
•To set the
•In the "AUTO" setting, inflation level may not be set automatically when the systolic blood pressure is more than 220 mmHg. Use the unit in the manual setting.
•If the cuff has not been inflated to the necessary level, it may be
FEATURES AND FUNCTIONS
List of Functions for each Mode
MeasurementMode | Single (SINGLE) | Average | Auscultation | Check |
Function | Mode | (AVG) Mode | (MANU) Mode | (CHECK) Mode |
Inflation level set- |
|
|
|
|
ting |
|
|
|
|
tion
Manual deflation
control
(4) AVG Function setting
You can set the number of measurements, the waiting time until the 1st measurement, and the measurement interval for the AVG Mode.
Function # | Items to set | Set value |
F1 | Number of measurements | 2 times or 3 times |
F2 | Waiting time until the start of 1st measurement | 0 sec, 3 min, 5 min, or 10 min. |
F3 | Measurement interval | 5 sec, 30 sec, 1 min, or 2 min. |
(2) Non-display function
HIDE
Note: The bold letters represent the
Use to prohibit the display of measurement results. However, the cuff pressure during measurement is displayed.
This function can be used in the SINGLE and AVG Modes.
By pushing the | HIDE | Button, display or |
|
is switched alternatingly.
SYS
|
|
|
| ON/OFF |
|
| HIDE |
| START |
|
|
|
| |
|
|
| A | DEFLATION |
|
|
| d | |
MODE |
| VG /1s t/2nd/3r | ||
|
| |||
140 | 180 | SINGLE |
|
|
AVG. | MANU. |
| ||
|
|
| ||
100 |
| 220 | CHECK |
|
|
| 260 |
| STOP |
|
|
|
| |
AUTO |
| 280 |
|
|
XL
Procedure to change the set values
1) When the power is OFF, push the ON/OFF (power) Button for more than three seconds while holding the START Button;
ON/OFF
F1 is displayed.
AC
START
HIDE
|
| DEFLATION |
AVG /1s /t2nd/3rd | ||
MODE | ||
140 | 180 | SINGLE |
| AVG. | MANU. |
100 | 220 | CHECK |
STOP
| 260 |
AUTO | 280 |
Display DIA
mmHg
mmHg
2)Push the START Button and select the function to set from F1 to F3. Each time you push the START Button, the func-
tions change in the order of .
XL
PULSE
BPM AC | AC |
(3) Manual deflation control
Accelerate deflation by pushing the DEFLATION (deflation control) / Measurement Result Display Switch Button during deflation in the measurement by the MANU Mode.
With each push of the Button, cuff is deflated rapidly in increments of 5 to 10 mmHg.
|
|
|
| ON/OFF |
|
| HIDE |
| START |
|
|
|
| |
|
|
| A | DEFLATION |
|
|
| d | |
MODE |
| VG /1s t/2nd/3r | ||
|
| |||
140 | 180 | SINGLE |
|
|
|
| AVG. | MANU. |
|
100 |
| 220 | CHECK |
|
|
| 260 |
| STOP |
|
|
|
| |
AUTO |
| 280 |
|
|
3)Push the DEFLATION (deflation control)/Measurement Result Display Switch Button and change the set values.
4)When the setting is finished, push the ON/OFF (power) Button to turn off the power. The setting is changed.
min
sec
ON/OFF
AC
START
HIDE
|
| DEFLATION |
AVG /1s /t2nd/3rd | ||
MODE | ||
140 | 180 | SINGLE |
| AVG. | MANU. |
100 | 220 | CHECK |
STOP
| 260 |
AUTO | 280 |
XL
ON/OFF
AC
START
HIDE
|
| DEFLATION |
AVG /1s /t2nd/3rd | ||
MODE | ||
140 | 180 | SINGLE |
| AVG. | MANU. |
100 | 220 | CHECK |
STOP
| 260 |
AUTO | 280 |
12 | 13 |