Omron Healthcare HEM-907XL Functions setting, Inflation level setting P-SET, AVG Function setting

Page 7

FEATURES AND FUNCTIONS

Functions setting

(1) Inflation level setting P-SET

AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected. The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure (approximately 30-40 mmHg above the patient’s systolic pressure).

Manual level setting: Inflation level can be set manually between 100 and 280 mmHg. Set the level to 30 to 40 mmHg higher than the expected systolic pressure.

To set the P-SET to "AUTO," turn the P-SET Knob counterclockwise as far as it goes until you can hear a click.

In the "AUTO" setting, inflation level may not be set automatically when the systolic blood pressure is more than 220 mmHg. Use the unit in the manual setting.

If the cuff has not been inflated to the necessary level, it may be re-inflated automatically.

FEATURES AND FUNCTIONS

List of Functions for each Mode

MeasurementMode

Single (SINGLE)

Average

Auscultation

Check

Function

Mode

(AVG) Mode

(MANU) Mode

(CHECK) Mode

Inflation level set-

 

 

 

 

ting

 

 

 

 

Non-display func-

tion

Manual deflation

control

(4) AVG Function setting

You can set the number of measurements, the waiting time until the 1st measurement, and the measurement interval for the AVG Mode.

Function #

Items to set

Set value

F1

Number of measurements

2 times or 3 times

F2

Waiting time until the start of 1st measurement

0 sec, 3 min, 5 min, or 10 min.

F3

Measurement interval

5 sec, 30 sec, 1 min, or 2 min.

(2) Non-display function

HIDE

Note: The bold letters represent the factory-set values.

Use to prohibit the display of measurement results. However, the cuff pressure during measurement is displayed.

This function can be used in the SINGLE and AVG Modes.

By pushing the

HIDE

Button, display or non-display of status

 

is switched alternatingly.

SYS

 

 

 

 

ON/OFF

 

 

HIDE

 

START

 

 

 

 

 

 

 

A

DEFLATION

 

 

 

d

P-SET

MODE

 

VG /1s t/2nd/3r

 

 

140

180

SINGLE

 

 

AVG.

MANU.

 

 

 

 

100

 

220

CHECK

 

 

 

260

 

STOP

 

 

 

 

AUTO

 

280

 

 

XL

Procedure to change the set values

1) When the power is OFF, push the ON/OFF (power) Button for more than three seconds while holding the START Button;

ON/OFF

F1 is displayed.

AC

START

HIDE

 

 

DEFLATION

P-SET

AVG /1s /t2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

Display DIA

mmHg

mmHg

Non-display

2)Push the START Button and select the function to set from F1 to F3. Each time you push the START Button, the func-

tions change in the order of .

XL

PULSE

BPM AC

AC

(3) Manual deflation control

Accelerate deflation by pushing the DEFLATION (deflation control) / Measurement Result Display Switch Button during deflation in the measurement by the MANU Mode.

With each push of the Button, cuff is deflated rapidly in increments of 5 to 10 mmHg.

 

 

 

 

ON/OFF

 

 

HIDE

 

START

 

 

 

 

 

 

 

A

DEFLATION

 

 

 

d

P-SET

MODE

 

VG /1s t/2nd/3r

 

 

140

180

SINGLE

 

 

 

 

AVG.

MANU.

 

100

 

220

CHECK

 

 

 

260

 

STOP

 

 

 

 

AUTO

 

280

 

 

3)Push the DEFLATION (deflation control)/Measurement Result Display Switch Button and change the set values.

4)When the setting is finished, push the ON/OFF (power) Button to turn off the power. The setting is changed.

min

sec

ON/OFF

AC

START

HIDE

 

 

DEFLATION

P-SET

AVG /1s /t2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

XL

ON/OFF

AC

START

HIDE

 

 

DEFLATION

P-SET

AVG /1s /t2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

12

13

Image 7
Contents Know Your Unit How to Measure Blood PressureSpecifications Be Sure to Read This SectionExemptions General advice Maintenance and inspection Features of the ProductTroubleshooting Included and also available separately Components of the ProductIntelliSense Blood Pressure Monitor, Model HEM-907XL Not included Cuffs without bladder BladdersNames of the Parts Features and FunctionsNon-display function Functions settingInflation level setting P-SET AVG Function settingSelect the cuff according to the arm size HOW to Apply the CuffHow to use the AC adapter HOW to USE the Power SourceInstallation and Replacement of Battery Pack List of Measurement Modes Blood Pressure in Single Mode HOW to MeasureBlood Pressure in Average Mode Unit does not automatically deflate in the Manu Mode Blood Pressure in Manual ModeWhat you need to prepare HOW to CheckPressure Accuracy in Check Mode How to checkTroubleshooting List of Error CodesWrap the cuff correctly, Specifications Five Year Limited WarrantyFor Customer Service Call Toll Free