LIST OF ERROR CODES
| Error code | Explanation | How to correct |
|
|
|
|
|
| Inflation error | • Confirm that the air tube connecting the cuff and |
|
| ||
|
| • When the pressure does not exceed 12 | the main unit is connected securely. |
|
| mmHg within the set time after the start of |
|
|
| inflation | • Confirm that the air flow in the air tube |
|
| • When the inflation does not reach the set | connecting the cuff and the main unit isn’t |
|
| cuff pressure within the specified time | being restricted. |
|
| after the start of inflation | • Confirm that the cuff is wrapped correctly (refer |
|
|
| |
|
|
| |
|
| Deflation error | to pages 14 and 15). |
|
| • When the deflation speed is too fast during |
|
|
| the measurement | • Check bladder for leaks and, if necessary, |
|
| • When the deflation speed is too slow | replace the bladder with new one (option). |
|
| during the measurement |
|
•When the measurement does not finish within the specified time after starting the measurement
|
| Overpressure error | • Confirm that air flow in the air tube connecting |
|
| • The cuff pressure exceeded 299 mmHg. | the cuff and the main unit isn’t being restricted. |
|
|
|
|
|
| Insufficient inflation error | • If the measurement is made by setting the |
|
| ||
|
| • Blood pressure could not be measured | to "AUTO", ask the patient not to move during |
|
| due to insufficient inflation level. | the inflation. |
|
|
| • Confirm that the |
|
|
| "AUTO". Turn the Knob counterclockwise as far |
|
|
| as it goes until you can hear a click sound. |
|
|
| • If the measurement is made by manual inflation |
|
|
| level setting, set the value to 30 to 40 mmHg |
|
|
| higher. |
|
|
|
|
|
|
|
|
|
| Indeterminable blood pressure error |
|
|
|
| |
|
| • Blood pressure could not be measured | • Confirm that the cuff is wrapped correctly (refer |
|
| even when the cuff pressure reached the | to pages 14 and 15). |
|
| specified pressure. |
|
|
|
|
|
|
| Low pulse level error |
|
|
|
| |
|
| • Pulse wave was too small. | • Confirm that the cuff is wrapped correctly (refer |
|
|
| to pages 14 and 15). |
|
|
|
|
|
| Blood pressure error |
|
|
| • Relationship between systolic and dias- | • Ask the patient not to move during the measure- |
|
| tolic pressures was abnormal. | ment. |
|
|
|
|
|
| Pulse rate error | • Check the patient for arrhythmia. |
|
| ||
|
| • Pulse rate did not stay within the range of |
|
|
| 30 to 199 beats/min. |
|
|
|
|
|
|
| Device error | • Contact Omron’s repair department |
|
| ||
|
| • Main unit malfunction. | |
|
|
|
|
TROUBLESHOOTING
If the unit malfunctions during use, please check the following:
Trouble | What to inspect | How to correct | |
|
| Wrap the cuff correctly, and | |
| Is the cuff wrapped correctly? | measure again. | |
|
| (Refer to Page 14 and 15.) | |
|
|
| |
| Is the patient moving during | Ask the patient not to move | |
The unit inflates to | during measurement, and | ||
inflation? | |||
abnormally high (low) | measure again. | ||
| |||
pressure. |
|
| |
|
| Set the | |
| Does this patient have arrhythmia? | higher than estimated systolic | |
| pressure of the patient, then | ||
|
| ||
|
| measure. | |
|
|
| |
|
| After checking the patient with | |
| Check the patient's condition. | the stethoscope refer to the "list | |
| of error codes". | ||
|
| ||
|
| (Refer to Page 28.) | |
|
|
| |
| Is the patient moving during the | Ask the patient not to move | |
| during measurement, and | ||
| measurement? | ||
| measure again. | ||
|
| ||
The monitor cannot |
|
| |
| Set the | ||
measure blood | Does the patient have an | higher than estimated systolic | |
pressure. | arrhythmia? | pressure of the patient, then | |
Measured values are |
| measure. | |
|
| ||
|
| ||
abnormally high (low). | Is the size of the cuff correct and | Select the cuff according to the | |
| patient's arm circumference, wrap | ||
| is it wrapped correctly? | ||
| it correctly, then measure again. | ||
|
| ||
|
| (Refer to Pages 14 and 15.) | |
|
|
| |
|
| Keep the level of the brachium to | |
| Is the level of the brachium to | which the cuff is wrapped at the | |
| which the cuff is wrapped at the | same level as the heart, then | |
| same level as the heart? | measure again. |
Are the patient’s clothes restricting | Remove the clothing and |
normal blood flow to the arm? | measure again. |