Omron Healthcare HEM-907XL manual List of Error Codes, Troubleshooting, Wrap the cuff correctly

Page 15

LIST OF ERROR CODES

 

Error code

Explanation

How to correct

 

 

 

 

 

 

Inflation error

• Confirm that the air tube connecting the cuff and

 

 

 

 

• When the pressure does not exceed 12

the main unit is connected securely.

 

 

mmHg within the set time after the start of

 

 

 

inflation

• Confirm that the air flow in the air tube

 

 

• When the inflation does not reach the set

connecting the cuff and the main unit isn’t

 

 

cuff pressure within the specified time

being restricted.

 

 

after the start of inflation

• Confirm that the cuff is wrapped correctly (refer

 

 

 

 

 

 

 

 

Deflation error

to pages 14 and 15).

 

 

• When the deflation speed is too fast during

 

 

 

the measurement

• Check bladder for leaks and, if necessary,

 

 

• When the deflation speed is too slow

replace the bladder with new one (option).

 

 

during the measurement

 

When the measurement does not finish within the specified time after starting the measurement

 

 

Overpressure error

• Confirm that air flow in the air tube connecting

 

 

• The cuff pressure exceeded 299 mmHg.

the cuff and the main unit isn’t being restricted.

 

 

 

 

 

 

Insufficient inflation error

• If the measurement is made by setting the P-SET

 

 

 

 

• Blood pressure could not be measured

to "AUTO", ask the patient not to move during

 

 

due to insufficient inflation level.

the inflation.

 

 

 

• Confirm that the P-SET is securely set to

 

 

 

"AUTO". Turn the Knob counterclockwise as far

 

 

 

as it goes until you can hear a click sound.

 

 

 

• If the measurement is made by manual inflation

 

 

 

level setting, set the value to 30 to 40 mmHg

 

 

 

higher.

 

 

 

 

 

 

 

 

 

 

Indeterminable blood pressure error

 

 

 

 

 

 

• Blood pressure could not be measured

• Confirm that the cuff is wrapped correctly (refer

 

 

even when the cuff pressure reached the

to pages 14 and 15).

 

 

specified pressure.

 

 

 

 

 

 

 

Low pulse level error

 

 

 

 

 

 

• Pulse wave was too small.

• Confirm that the cuff is wrapped correctly (refer

 

 

 

to pages 14 and 15).

 

 

 

 

 

 

Blood pressure error

 

 

 

• Relationship between systolic and dias-

• Ask the patient not to move during the measure-

 

 

tolic pressures was abnormal.

ment.

 

 

 

 

 

 

Pulse rate error

• Check the patient for arrhythmia.

 

 

 

 

• Pulse rate did not stay within the range of

 

 

 

30 to 199 beats/min.

 

 

 

 

 

 

 

Device error

• Contact Omron’s repair department toll-free at

 

 

 

 

• Main unit malfunction.

(1-877-216-1336).

 

 

 

 

TROUBLESHOOTING

If the unit malfunctions during use, please check the following:

Trouble

What to inspect

How to correct

 

 

Wrap the cuff correctly, and

 

Is the cuff wrapped correctly?

measure again.

 

 

(Refer to Page 14 and 15.)

 

 

 

 

Is the patient moving during

Ask the patient not to move

The unit inflates to

during measurement, and

inflation?

abnormally high (low)

measure again.

 

pressure.

 

 

 

 

Set the P-SET to 30 to 40 mmHg

 

Does this patient have arrhythmia?

higher than estimated systolic

 

pressure of the patient, then

 

 

 

 

measure.

 

 

 

 

 

After checking the patient with

 

Check the patient's condition.

the stethoscope refer to the "list

 

of error codes".

 

 

 

 

(Refer to Page 28.)

 

 

 

 

Is the patient moving during the

Ask the patient not to move

 

during measurement, and

 

measurement?

 

measure again.

 

 

The monitor cannot

 

 

 

Set the P-SET to 30 to 40 mmHg

measure blood

Does the patient have an

higher than estimated systolic

pressure.

arrhythmia?

pressure of the patient, then

Measured values are

 

measure.

 

 

 

 

abnormally high (low).

Is the size of the cuff correct and

Select the cuff according to the

 

patient's arm circumference, wrap

 

is it wrapped correctly?

 

it correctly, then measure again.

 

 

 

 

(Refer to Pages 14 and 15.)

 

 

 

 

 

Keep the level of the brachium to

 

Is the level of the brachium to

which the cuff is wrapped at the

 

which the cuff is wrapped at the

same level as the heart, then

 

same level as the heart?

measure again.

Are the patient’s clothes restricting

Remove the clothing and

normal blood flow to the arm?

measure again.

28

29

Image 15
Contents Know Your Unit How to Measure Blood PressureSpecifications Be Sure to Read This SectionExemptions General advice Features of the Product Maintenance and inspectionTroubleshooting Included and also available separately Components of the ProductIntelliSense Blood Pressure Monitor, Model HEM-907XL Not included Cuffs without bladder BladdersNames of the Parts Features and FunctionsNon-display function Functions settingInflation level setting P-SET AVG Function settingSelect the cuff according to the arm size HOW to Apply the CuffHOW to USE the Power Source How to use the AC adapterInstallation and Replacement of Battery Pack List of Measurement Modes Blood Pressure in Single Mode HOW to MeasureBlood Pressure in Average Mode Unit does not automatically deflate in the Manu Mode Blood Pressure in Manual ModeWhat you need to prepare HOW to CheckPressure Accuracy in Check Mode How to checkList of Error Codes TroubleshootingWrap the cuff correctly, Specifications Five Year Limited WarrantyFor Customer Service Call Toll Free