HOW TO CHECK
HOW TO CLEAN THE UNIT AFTER USE
PRESSURE ACCURACY (IN CHECK MODE)
Accuracy of pressure display can be checked in the CHECK Mode.
What you need to prepare
(1)Calibrated mercury manometer (including inflation bulb), (2)
When cleaning this unit, unplug the AC adapter from the electric outlet.
After cleaning this unit, dry it well, before plugging the AC adapter in the electric outlet.
General advice
•Do not clean this unit with gasoline, paint thinner, or high concentration alcohol.
•Do not disinfect this unit by autoclave or gas steril- ization (EtO, gluteraldehyde, or high concentration ozone.)
How to check
1. Connect the manometer, inflation bulb, cuff, and the monitor with the
2. Tightly wrap the cuff over a sturdy cylinder.
3. Release the valve of inflation bulb to remove the air inside the cuff completely.
4. Push the ON/OFF (power) Button to turn on the monitor.
5. Set the MODE Selector to “CHECK”.
6. Close the valve of inflation bulb and inflate the cuff to the pressure to be checked, based on the manometer read.
7. Compare the pressure values displayed on the monitor to the one on the manometer.
Check result
Accuracy of the monitor is validated to be ±3 mmHg or 2% of standard manometer reading. If your result shows a difference exceeding the tolerance, contact Omron repair department
1.Wipe the monitor with a soft, damp cloth diluted with disinfectant alcohol, or diluted detergent.
2.Complete cleaning by wiping the monitor with a soft, dry cloth.