A&D UA-767NFC instruction manual Troubleshooting

Page 18

Troubleshooting

Problem

 

 

 

Possible Reason

Recommended Action

Nothing

Batteries are drained.

Replace all batteries with new

ones.

appears on the

 

 

 

 

 

 

 

 

 

 

 

 

Reinstall the batteries with

display, even

 

 

 

 

 

 

Battery terminals are not

negative and positive terminals

when the power

is turned on.

in the correct position.

matching those indicated on the

 

 

 

 

 

 

battery compartment.

 

 

 

 

 

 

 

 

Battery voltage is too low.

 

The cuff does

 

 

 

(LOW BATTERY mark)

Replace all batteries with new

 

 

 

 

 

 

 

 

not inflate.

blinks. If the batteries are

ones.

drained completely, the

 

 

 

mark does not appear.

 

 

The cuff is not applied

Apply the cuff correctly.

 

properly.

 

 

 

You moved your arm or

Make sure you remain very still

The device does

body during measurement.

and quiet during measurement.

 

 

 

Sit comfortably and still.

not measure.

 

 

 

 

 

 

The cuff position is not

Place your arm on a table with your

Readings are

correct.

palm facing upward and the cuff at

too high or too

 

 

 

 

 

 

the same level as your heart.

low.

 

 

 

 

 

 

 

 

 

 

 

 

If you have a very weak or

 

 

 

 

 

 

 

 

 

 

 

 

 

 

irregular heat beat, the device

 

 

 

 

 

 

 

may have difficulty in determining

 

 

 

 

 

 

 

 

 

 

 

 

 

 

your blood pressure.

 

The value is different

Refer to “Why Measure Blood

 

from that measured at a

 

Pressure at Home?”.

Other

clinic or doctor’s office.

 

 

 

 

 

 

 

Remove the batteries. Place them

 

 

 

 

 

 

 

 

 

 

 

 

 

 

back properly and take another

 

 

 

 

 

 

 

 

 

 

 

 

 

 

measurement.

Note: If the actions described above do not solve the problem, contact the dealer. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid.

17

Image 18
Contents Model UA-767NFC Contents Precautions Dear CustomersPreliminary Remarks Parts Identification DisplaySymbols Symbols that are printed on the device caseMeasurement Battery is low when it blinks Memory Function Operation ModeNormal Measurement Deleting all Data Stored in MemoryConnecting the Air Hose Using the MonitorInstalling / Changing the Batteries Connecting the AC AdapterSelecting the Clock Mode Adjusting the Clock Select the year using the buttonSelecting the Correct Cuff Size Applying the Arm CuffHow to Take Proper Measurements MeasurementAfter Measurement Measurements Button until a Memory Data Transmission Measurement Data TransmissionWhat is Blood Pressure? What is an Irregular HeartbeatPressure Bar Indicator About Blood PressureWHO Blood Pressure Classification What is Hypertension and How is it Controlled?Why Measure Blood Pressure at Home? Blood Pressure VariationsTroubleshooting Maintenance Technical DataENGINEERING, INC