NurseCall Relay Unit

Programming en 21

 

 

Manual selection of the NurseCall Main Unit

5.a. Scroll with and to set manually the identification number. Confirm with or the Green button. The Relay Unit signalizes that it is not connected to the RS-485 bus.

NurseCall

RELAY UNIT OFF !!!

5.b. Connect the NurseCall Relay Unit to the RS-485 bus. The identification number is now displayed at the bottom right.

4.3

Special settings

After pressing three times quickly, you can enter special codes.

Enter Code

4.3.1

Displaying the firmware version

 

1.

Type the code 194155.

 

2.

The version of the firmware will be displayed for a few seconds.

Software REV A

V2.02 BN111.242.00A

4.3.2

Resetting all the parameters and buffers

CAUTION!

Disconnect the RS-485 bus before performing a reset. Reconnect it when you are finished.

NOTICE!

The following procedure resets the NurseCall Relay Unit to its original factory settings. It erases all the parameters, as well as the alarms and events buffers. This reset is mandatory if you wish to change the locating mode or the identification number.

1.Type the code 194156.

2.When the unit resets, it plays a short melody and displays a temporary message.

Please Wait...

 

3.

After a few seconds, the unit goes back to the first use display.

 

4.

Select the locating mode and the identification number. See Section 4.2 First use.

4.3.3

Assistance and fire priority

 

This command sets the assistance call and the fire alarm as priority calls.

 

This means that alarms of these types will be displayed first.

 

1.

Type the code 123991. The unit plays a melody and a confirmation message is displayed.

 

 

 

 

 

 

ASSISTANCE & FIRE

 

 

 

PRIORITY

 

 

 

 

 

Bosch Security Systems

User Manual

F.01U.252.722 V1 2011.12

Image 21 Contents
NurseCall Relay Unit Page Table of Contents Operation MaintenanceDisposal AppendixImportance of safety instructions Safety instructionsDisregarding safety rules General safety instructions1Observation and information General description DescriptionTop view Yellow buttonBottom view Front view Relay Unit rear viewStatus NurseCall Relay Unit V2.02 BN111.242.00A Detailed descriptionLoudspeaker DisplayKeys Function 4RS-485 interface 5AntennaRelay output 3Installation List of contents1Unpacking Installation on a piece of furniture InstallationWall installation Installing the antennaConnecting to the mains Disassemble the unit see .5.1 Disassembling the unit, 7Connecting the RS-485 Incorrect connection Correct connection Setting the 100 Ohm termination jumper 8Connecting the RS-232 to the Relay Plus UnitGeneralities ProgrammingDate TimeLocating ? First useOFF / on Relay Unit ID · +Special settings Relay Unit OFFAssistance & Fire Priority Assistance & Fire Nonpriority Printer/display setup on the Relay Plus UnitMultitone Texte BAD/WC + HILFE-2 Standardtexte Technik + NOTRUF2Display Zettler ¬ Display ZettlerCriter + 1 Spaces Display Norsk ¬ Display Norsk Rota.of AlarmsYES ¡Display ffrrrb FL 2 +RO 3 +BE1 Selection Display NorskAdjusting the loudspeaker volume Consulting the alarm or event buffer5Operation Display indications Switching between alarm and event buffers indicationFirst information block Second information blockHelp Local ACKMonitoring the power supply 6MaintenanceMonitoring the backup battery Checking the systemCleaning Parts replacementDisassembling the unit Backup battery replacing Important Safety InstructionsProcedure Battery DisassemblyDisposal Returning to the manufacturerElectrical specifications Dimensions and weightAAppendix List of criteria Criterion Number Alarm a or Event E CommentPower socket unit bottom ConnectorsLine socket unit bottom RS-232 socket Relay Plus Unit rearPage Bosch Security Systems