24 en Programming

NurseCall Relay Unit

 

 

c.Choose 0 to 9 characters for the criterion, followed by 0 to 9 spaces. Default values are 2 characters + 1 space.

7.Repeat the operation for

a.the floor (0 to 3 characters and 0 to 9 spaces), by default 2 characters and 1 space

b.the room (0 to 3 characters and 0 to 9 spaces), by default 2 characters and 1 space

c.the bed (0 or 1 character), by default 1 character.

NOTICE!

Configuration of the display is possible only in display mode FL,RO,BE (floor, room, bed).

If the NurseCall system is in display mode NUMBER, there is no need of further programming.

4.4.5

Selection "DISPLAY NORSK"

NOTICE!

The display mode NUMBER ZZZZ (alphanumeric display) is not compatible with the Display Norsk. Make sure to choose another display mode when connecting a Display Norsk to the Relay Plus Unit. See the NurseCall Main Unit user manual for more information.

if you have selected DISPLAY NORSK, press to confirm.

¬ RS-232 Output

 

DISPLAY NORSK

¢

1.Set the rotational display of alarms.

¬DISPLAY NORSK Rota.of Alarms:YES ¡

2.Select YES or NO with and . Default value is YES. Confirm with .

3.Configure the display of the 6 identification positions. The outcome is visible at the top.

DISPLAY : ffrrrb

FL: 2 +RO: 3 +BE:1

– Select a field with

and

, change a value with

and

.

Confirm with .

Choose 0 to 9 characters for the criterion, followed by 0 to 9 spaces. Default values are 2 characters and 1 space.

4.Repeat the operation for

the floor (0 to 3 characters), by default 2 characters

the room (0 to 3 characters), by default 3 characters

the bed (0 or 1 character), by default 1 character.

NOTICE!

Configuration of the display is possible only in display mode FL,RO,BE (floor, room, bed).

If the NurseCall system is in display mode NUMBER, there is no need of further programming.

F.01U.252.722 V1 2011.12

User Manual

Bosch Security Systems

Image 24 Contents
NurseCall Relay Unit Page Table of Contents Maintenance OperationDisposal AppendixSafety instructions Importance of safety instructionsDisregarding safety rules General safety instructions1Observation and information Description General descriptionTop view Yellow buttonBottom view Front view Relay Unit rear viewStatus Detailed description NurseCall Relay Unit V2.02 BN111.242.00ALoudspeaker DisplayKeys Function 4RS-485 interface 5AntennaRelay output 3Installation List of contents1Unpacking Installation Installation on a piece of furnitureWall installation Installing the antennaConnecting to the mains Disassemble the unit see .5.1 Disassembling the unit, 7Connecting the RS-485 Incorrect connectionCorrect connection 8Connecting the RS-232 to the Relay Plus Unit Setting the 100 Ohm termination jumperProgramming GeneralitiesDate Time First use Locating ? OFF / on Relay Unit ID · +Special settings Relay Unit OFFAssistance & Fire Priority Printer/display setup on the Relay Plus Unit Assistance & Fire NonpriorityMultitone Texte BAD/WC + HILFE-2 Standardtexte Technik + NOTRUF2Display Zettler ¬ Display ZettlerCriter + 1 Spaces ¬ Display Norsk Rota.of AlarmsYES ¡ Display NorskDisplay ffrrrb FL 2 +RO 3 +BE1 Selection Display NorskAdjusting the loudspeaker volume Consulting the alarm or event buffer5Operation Switching between alarm and event buffers indication Display indicationsFirst information block Second information blockLocal ACK Help6Maintenance Monitoring the power supplyMonitoring the backup battery Checking the systemCleaning Parts replacementDisassembling the unit Backup battery replacing Important Safety InstructionsProcedure Disassembly BatteryDisposal Returning to the manufacturerElectrical specifications Dimensions and weightAAppendix Criterion Number Alarm a or Event E Comment List of criteriaConnectors Power socket unit bottomLine socket unit bottom RS-232 socket Relay Plus Unit rearPage Bosch Security Systems