Sunrise Medical F45 owner manual Safety and caution notes

Page 45

Batteries and charging

At that point the battery is about 85% - 90% charged. It is essential to provide a full recharge at the first opportunity.

5.Do not leave the charger connected to the battery with the mains disconnected or switched off. This could result in damage to your battery being caused by deep discharge over a period of time.

6.Always switch off at the mains before disconnecting the batteries.

Safety and caution notes

1.The charger is designed for indoor use. Do not use outdoors or expose to rain, snow, spray or moisture.

2.Use only with Sonnenschein dryfit Gel type batteries of the A200, A500 or A500C groups when employed in cyclic mode.

3.The charger may be used with other brands of Gel type batteries, subject to written confirmation from the Technical Department of Sunrise Medical.

4.Do not smoke, or create sparks or flames in the vicinity of the batteries.

5.Avoid standing the charger on carpets or rugs during use and do not cover the ventilation slots of the charger or allow objects to rest on the case.

6.Do not attempt to recharge other types of batteries such as car batteries, non-rechargeable

batteries, AGM sealed batteries, wet electrolyte batteries or batteries of the wrong voltage.

7.Repairs should only be carried out by Sunrise Medical.

8.The charger is designed for mains input voltages of 220V AC to 240V AC, for chargers to operate with 110V AC (in countries such as USA or North America), a different charger will be needed, please consult local Sunrise Medical Dealer.

9.Use of an attachment, not recommended or sold by Sunrise Medical, may result in a risk of fire, electric shock or injury to persons. To reduce the risk of damage to electric plug and cord, pull by the plug rather than the cord when disconnecting the charger. Make sure the cord is located so that it will not be stepped on, tripped over, or otherwise subjected to damage or stress. An extension cord should not be used unless absolutely necessary. Use of an improper extension cord could result in a risk of fire and electric shock. If any extension cord must be used, make sure the pins on the plug of the extension cord are the same number, size and shape as those of the plug on the charger; and that the extension cord is properly wired and in good electrical condition.

41

F45 Issue 6

 

 

Image 45
Contents F45 F45 Issue F45 How to use this manualContents Serial Number IntroductionYour local dealer GuaranteeFeatures and options Routine service Safety warning and user tipsKerbs Sharp turnsCleaning controls Cleaning seatingBatteries Wheelchair motorsWheelchair range Use on a slopeMobile telephones and two way radios Preparing your wheelchair for use Armrests-removingArmrests-replacing Control box Fitting footrestsDedocking procedure Emergency freewheel and dedockingIt is not intended for permanent use Dedocking Ensure the freewheel levers are in the drive position Refitting motor drive unitsAdjusting the armrest width Adjusting armrest heightArmrests Adjusting the footrest angle FootrestsFootrest Footplates Adjusting the footrest lengthSeating Seat sling tension adjustmentFirm seat board Seat cushionsOptima backrest upholstery Backrest removal and refittingChanging seat depth Changing the shape or tensionFully adjustable backrest Lumbar supportRemovable covers JAY backrest posts Manual adjustable backrest reclinePenny and Giles Pilot + control pod Power supplyImmobiliser plug To activate the immobiliserBattery level indicator Adjustable speed settingHorn button IndicatorsMain lights Hazard lightsMenu select Actuator mode functionLegrest left Backrest recline Seat tilt Legrest right Up/down movementOperating the control joystick Charging socketSummary Penny and Giles Integral control Adjustable speed setting Fault Type Red bars only lit or slow flashDescription Parallel Swing Away Arm Kerb climber Fitting and removal procedureMounting the kerb Front transfersDismounting the kerb Batteries and charging Maintenance free batteries Battery careGeneral information Maintenance free battery care plan Safety features Yellow mains On 24V 6A charger indicator lightsOrange charging Yellow bulk charge completeNo lights Fault finding guide 24V 6A chargerFlashing lights Yellow mains on light onlyProcedure for connecting the charger and charging Safety and caution notes Battery warranty Myths and legends Range of your vehicleUser tips Adverse conditionsRamps Transfer to and from the chair Gradients descents Gradients ascentsClamp points TransportationTransportation In vehicles Lap belt instructions Sling Back Option Sling back option MaintenanceMaintenance To order spare parts F45 maintenance and routine inspectionYear High Street Wollaston West Midlands DY8 4PS Sunrise Medical LimitedF45 Issue F45 Issue Sunrise Medical Limited