Emerson Process Management 3051N manual Function Keys, Alphanumeric and Shift Keys, Data Entry

Page 84

Rosemount 3051N

Reference Manual

00809-0100-4808, Rev CA June 2008

Hot Key

Function Keys

Use this key to quickly access important, user-selectable options when connected to a HART-compatible device. Pressing the HOT KEY turns the HART Communicator on and displays the Hot Key Menu. See “Customizing the Hot Key Menu” in the Rosemount 275 HART Communicator manual for more information.

Use the four software-defined function keys, located below the LCD, to perform software functions. On any given menu, the label appearing above a function key indicates the function of that key for the current menu. As you move among menus, different function key labels appear over the four keys. For example, in menus providing access to on-line help, the HELP label may appear above the F1 key. In menus providing access to the ONLINE MENU, the HOME label may appear above the F3 key. Simply press the key to activate the function. See your Rosemount 275 HART Communicator manual for details on specific function key definitions.

Alphanumeric and Shift Keys

FIGURE A-7. HART

Communicator Alphanumeric

and Shift Keys.

Data Entry

FIGURE A-8. Data Entry Key Sequence.

Some menus require data entry. Use the alphanumeric and shift keys to enter all alphanumeric information into the HART Communicator. If you press an alphanumeric key alone from within an edit menu, the bold character in the center of the key appears. These large characters include the numbers zero through nine, the decimal point (.), and the dash symbol (–).

To enter an alphabetic character, first press the shift key that corresponds to the position of the letter you want on the alphanumeric key. Then press the alphanumeric key. For example, to enter the letter “R”, first press the right shift key, then the “6” key (see Figure A-8). Do not press these keys simultaneously, but one after the other.

A-8

Image 84
Contents Reference Manual Page Rosemount 3051N Return of Material Changes from June 2006 to June OldChanges Rosemount 3051N Table of Contents Installation TroubleshootingSpecifications Reference Data Options Appendix aHart Communicator Glossary Index Using this Manual Section IntroductionRosemount 3051N Overview Safety Messages Transmitter FunctionsAlarm Level Verification TransmitterSecurity Configuring Transmitter Alarm and Security Jumper Procedure Physical Removal Local Zero and Span Local KeysElectronics Board Setting the Loop to Manual Wiring Diagrams Bench Hook-up Wiring Diagrams Field Hook-up Review Configuration Data Check OutputProcess Variables Sensor Temperature Basic SetupSet Process Variable Units Set OutputRerange Rerange with a Communicator OnlyRerange with a Pressure Input Source and a Communicator Span and Zero Adjustment Buttons Detailed Setup Diagnostics ServiceTransmitter Test Loop TestCalibration Transmitter Bench Calibration Tasks Field Calibration Tasks Configure the Analog Output Parameters Calibration OverviewSensor Trim URVZero Trim Typical Zero vs Non-zero-Based Application IllustrationsFull Trim Digital Trim Connection Drawing 4-20 mA TransmittersDigital-to-Analog Trim Using Other Scale Recall Factory Trim Recall Factory Trim- Sensor TrimRecall Factory Trim- Analog Output Analog Output TrimZero Effect HTc = URV + S URV Ps Rosemount 3051N Section Installation Rosemount 3051N Typical Installation Flowchart General Considerations Special Draft Range ConsiderationsMechanical Considerations 156 181 16 x 11/4 Bolts for Mounting to 54.1 Transmitter 120 127 66.8 Traditional Flange Option Code H2 Dimensional Drawing Mounting Rosemount 3051N GAS or Liquid Service GAS Service Steam Service Process Connections Housing Rotation Power SupplyElectrical Considerations Wiring Power Supply Load LimitationsAccess Requirements Cover Installation Grounding the Transmitter CaseEnvironmental Considerations Troubleshooting Symptom Corrective Actions Returning Rosemount Products and MaterialsNuclear Specifications SeismicModel Range Code Hydrostatic Test Pressure Performance Specifications Ambient Temperature Effect12Reference Accuracy DriftPower Supply Effect Load Effect Mounting Position Effect Overpressure Effect Static Pressure EffectRange Code Overpressure Effect Range Code Static Pressure Zero Effect1Functional Specifications ServiceTemperature Limits OutputSpan and Zero, Zero Elevation, Suppression Range Minimum Span Upper URL Lower LRLRange Code Minimum Time Constant Tc Including Dead Time Td Maximum Working Pressure12Static Pressure Limits Overpressure Limits Burst PressurePhysical Specifications Materials of Construction Option Code Description Add Ordering Information Configuration Information Housing Material Conduit Entry SizeMounting Bracket Options Meters optionalDate  On Rosemount 3051N Exploded View with Coplanar Process Flange Rosemount 3051N Rosemount 3051N Overview Safety Messages LCD Meter Section OptionsCustom Meter Configuration Installing the Meter Exploded View of the Rosemount 3051N with Optional LCD MeterRosemount 3051N Diagnostic Messages ErrorPress Limit Operation Mounting BracketsTraditional Flange H2 Transient Protection Terminal Block T1Transient Protection Terminal Block T1 Rosemount 3051N Introduction Safety Messages Appendix a Hart CommunicatorRosemount 3051N Online Menu Function Hart Fast Key Sequence Figure A-3. Rosemount Hart terminal access door Connections and HardwareFigure A-4. Bench Hook-up MA Transmitters Rosemount Communicator KeysAction Keys Function Keys Alphanumeric and Shift KeysData Entry Fast Key Sequences Fast Key Sequence Conventions Fast Key Sequence ExampleMenus Functions Main MenuOnline Menu Touch Alphanumeric Keys Multifunction LEDTouch Screen Multifunction LED Processing IndicationRosemount 3051N Diagnostic Messages Fast Key Sequences Hot Key optionsMessage Description Rosemount 3051N Uploadvariable Rosemount 3051N Rosemount 3051N Glossary Smart Family Index 00809-0100-4808, Rev CA June Security Page ¢00809-0100-4801V¤

3051N specifications

The Emerson Process Management 3051N is a state-of-the-art pressure transmitter that exemplifies precision and reliability in process automation. Designed for a wide range of industrial applications, the 3051N is celebrated for its advanced features, innovative technologies, and robust construction, catering to the evolving needs of process industries.

One of the standout features of the 3051N is its exceptional accuracy. With a standard accuracy rating of ±0.075% of span, it ensures precise pressure measurements, critical for maintaining optimal process conditions. This high level of accuracy makes it suitable for applications in industries such as oil and gas, chemicals, and water treatment, where even minor deviations can lead to significant operational issues.

The 3051N employs Smart Pressure Technology, integrating digital communication capabilities with traditional analog systems. This technology allows for seamless interoperability with existing process control systems, providing users with vital data for diagnostics, calibration, and configuration via HART, FOUNDATION fieldbus, or Profibus protocols. These capabilities lead to enhanced process efficiency and reduced downtime due to effective predictive maintenance.

Another notable characteristic of this transmitter is its robust design. The 3051N is built to endure harsh industrial environments, with a rugged housing that is resistant to corrosive and hazardous conditions. Its wide operating temperature range, typically from -40°C to 85°C, coupled with various diaphragm materials, ensures that it can function effectively across diverse applications.

Moreover, the Emerson 3051N incorporates advanced predictive diagnostics, enabling users to monitor the health of the transmitter continuously. This feature alerts operators to potential issues before they escalate, facilitating proactive maintenance strategies that can significantly lower operational costs.

The flexible configuration options offered by the 3051N also enhance its usability. Users can select from a variety of pressure ranges, process connection types, and electrical connections, making it easy to customize the transmitter to fit specific application needs.

In conclusion, the Emerson Process Management 3051N pressure transmitter is a versatile and highly efficient device that integrates advanced technology with robust design and exceptional accuracy. Its ability to adapt to various industrial demands while providing predictive diagnostics makes it a preferred choice for professionals seeking reliability and precision in process management.