SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

!DANGER

Improper filling or abuse of this cylinder, or failure to follow these safety instructions may cause serious bodily injury or death. High pressure oxidizing gas vigorously accelerates combustion.

DO NOT alter or modify cylinder or related components. DO NOT use Oil or Grease. Keep cylinder, cylinder valve, all equipment and connections free of oil and grease to avoid violent ignition resulting in serious bodily injury or death.

!WARNING

For medical applications. Use only as directed by a licensed practitioner. Uninterrupted use of high concentrations of oxygen over a long duration, without monitoring its effect on oxygen content of arterial blood, may be harmful.

Oxygen Cylinders filled by the Venture HomeFill

 

shall be used for personal use only. "Not to be

 

filled for resale or use by professional users".

OXYGEN

This cylinder is to be used with the Invacare Venture 2 HomeFill system. This cylinder contains 93 + 3% oxygen produced by an oxygen concentrator using the pressure swing adsorption process.

DO NOT handle cylinder or use contents until you are professionally trained, including emergency procedures.

Use in accordance with Venture HomeFill Operator’s Manual. Open valve slowly. Close valve after each use and when empty. Secure cylinder during storage and use. No smoking in cylinder area. Keep away from heat, flame and spark. Keep out of reach of children. DO NOT drop.

Transfilling of this gas is performed by Venture HomeFill Only.

Cylinders that have been refinished and/or exposed to elevated temperatures must be hydrostatically tested before filling. Cylinder temperature should not exceed 125°F (52°C). Aluminum cylinders subject to the action of fire or heated to temperatures in excess of 350°F must be withdrawn from service and condemned by trained personnel. DO NOT use caustic paint strippers. Valve and safety relief devices must be removed and replaced by trained and authorized personnel. DO NOT alter or change valve assembly.

Invacare Corporation

DO NOT REMOVE THIS PRODUCT LABEL

2101 E. Lake Mary Blvd.

Not for Commercial Resale

Sanford, FL 32773

Part Number 1125090 Rev B - 11/04

1-800-832-4707

FIGURE 4.1 External Examination

Oxygen Conserving Cylinder

12

Part No 1143247

Image 12
Contents HomeFill Oxygen Conserving Cylinder Oxygen Conserving Cylinder Table of Contents Special Notes Receiving InspectionIntended USE General Guidelines General Guidelines General Guidelines Features Product DiagramTypical Product Parameters Inspection and Compressor Interface Cylinder Prefill InspectionExternal Examination External Examination Connecting/Disconnecting the Cylinder to/from the Compressor Outer Ring Sleeve Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Fillport Covers Changing the Carrying Position of the Cylinder Changing the Carrying Position of the Cylinder Cleaning the Cylinder Bag Operating Instructions Operating Instructions Maintenance Troubleshooting Cylinder Fill TimesCylinder Fill Times Limited Warranty Limited Warranty USA