Invacare HF2PCM2KIT manual Connecting/Disconnecting the Cylinder to/from the Compressor

Page 13

SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

Connecting/Disconnecting the Cylinder to/from the Compressor

Connecting the Cylinder to the Compressor

WARNING

NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur.

DO NOT drop oxygen cylinders. Use two hands when handling/transporting oxygen cylinders. Otherwise, injury or damage may occur.

DO NOT use a liquid leak detector to test for leaks.

NOTE: For this procedure, refer to FIGURE 4.2 on page 14 and FIGURE 4.3 on page 15.

1.Examine the cylinder gauge. If the cylinder pressure is less than 1500 psig., proceed to STEP 2 (FIGURE 4.3 on page 15).

NOTE: If the cylinder pressure is greater than 1500 psig., DO NOT attempt to top off this cylinder. It may not fill.

2.Set the regulator flow dial on the cylinder to OFF (FIGURE 4.3 on page 15).

3.Remove the cylinder and connector fillport covers (if present).

CAUTION

DO NOT connect the outlet (cannula) connection to the HomeFill unit, otherwise the conserving cylinder will not work properly.

Part No 1143247

13

Oxygen Conserving Cylinder

Image 13
Contents HomeFill Oxygen Conserving Cylinder Oxygen Conserving Cylinder Table of Contents Receiving Inspection Special NotesIntended USE General Guidelines General Guidelines General Guidelines Product Diagram FeaturesTypical Product Parameters Cylinder Prefill Inspection Inspection and Compressor InterfaceExternal Examination External Examination Connecting/Disconnecting the Cylinder to/from the Compressor Outer Ring Sleeve Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Fillport Covers Changing the Carrying Position of the Cylinder Changing the Carrying Position of the Cylinder Cleaning the Cylinder Bag Operating Instructions Operating Instructions Maintenance Cylinder Fill Times TroubleshootingCylinder Fill Times Limited Warranty Limited Warranty USA