SECTION 5—OPERATING INSTRUCTIONS

SECTION 5—OPERATING

INSTRUCTIONS

WARNING

NEVER smoke in an area where oxygen is being administered.

NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.

NOTE: For this procedure, refer to FIGURE 5.1 on page 22.

1.Prior to each use, inspect product for visible damage. DO NOT use if any damage is found.

NOTE: If for any reason, any label becomes illegible or lost, contact your dealer/home care equipment provider.

2.Be sure regulator is in the “Off” position. Align “Off” with the indicating pointer.

3.Position the cylinder so the oxygen cylinder valve outlet is pointing away from the user and any other person(s).

WARNING

DO NOT direct flow of oxygen at any person, or flammable material when adjusting the flow dial.

4.Attach a standard adult single lumen oxygen nasal cannula, no longer than seven feet to the product’s outlet according to the cannula manufacturer’s instructions. DO NOT use pediatric, low flow nasal cannulas or oxygen masks with this product.

NOTE: A unit without cannula attached may pulse if set at position 1-5.

Part No 1143247

21

Oxygen Conserving Cylinder

Image 21
Contents HomeFill Oxygen Conserving Cylinder Oxygen Conserving Cylinder Table of Contents Special Notes Receiving InspectionIntended USE General Guidelines General Guidelines General Guidelines Product Diagram FeaturesTypical Product Parameters Cylinder Prefill Inspection Inspection and Compressor InterfaceExternal Examination External Examination Connecting/Disconnecting the Cylinder to/from the Compressor Outer Ring Sleeve Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Fillport Covers Changing the Carrying Position of the Cylinder Changing the Carrying Position of the Cylinder Cleaning the Cylinder Bag Operating Instructions Operating Instructions Maintenance Cylinder Fill Times TroubleshootingCylinder Fill Times Limited Warranty Limited Warranty USA