Scott 2.2, 4.5, 3 manual Other Operational Features END of Service Time Indicator, Battery Test

Page 11

OTHER OPERATIONAL FEATURES

END OF SERVICE TIME INDICATOR

The SEMS PDA distress alarm also provides an electronic end of ser- vice time indicator and cylinder pressure display as follows:

1/2 full cylinder

When the air supply cylinder has reached 1/2 full capacity, the audible signal will sound and the display will read “1/2 AIR” for 10 seconds.

1/4 full cylinder

When the air supply cylinder has reached 1/4 full capacity, the audible signal will sound and the display will read “LOW AIR” and then flash the remaining cylinder air pressure in psig until the user responds by pressing the RESET button. However, the alarm will continue to repeat actuation until the system is shut down or until the unit detects no movement by the user at which time the SEMS PDA distress alarm sequence will begin with PRE ALARM and FULL ALARM as stated above.

Cylinder pressure display

Pressing and holding the RESET at any time will show the ap- proximate cylinder pressure remaining in psig on the display.

BATTERY TEST

When the SEMS PDA distress alarm is in the off condition (cylinder valve closed with no flashing green lights, the batteries in the Sensor Module can be checked by depressing and holding the reset button on the console.

A green LED will illuminate on the console to indicate sufficient battery power remaining; a red light indicates that the batteries must be re- placed before the respirator is to be used again.

WARNING

THE RESPIRATOR USER MUST IMMEDI - ATELY LEAVE THE AREA REQUIRING RESPIRATORY PROTECTION WHEN AN E N D O F S E RV I C E I N D I C AT O R A L A R M ACTUATES . ACTUATION OF AN END OF S E RV I C E I N D I C ATO R A L A R M WA R N S T H AT A P P R O X I M AT E LY 2 5 % O F F U L L PRESSURE REMAINS IN THE AIR SUPPLY C Y L I N D E R ( T H AT I S , A P P R O X I M AT E LY 3 / 4 O F T H E TO TA L A I R S U P P LY H A S BEEN USED) . A DELAY IN LEAVING THE A R E A A F T E R A L A R M A C T U AT I O N M AY R E S U LT I N S E R I O U S I N J U RY O R DEATH .

WARNING

FAILURE TO REPLACE THE BATTERIES AND/ OR CONTINUING WITH MULTIPLE USES OF THE RESPIRATOR AFTER THE LOW BAT- TERY CONDITION HAS BEEN INDICATED BY THE SEMS PDA DISTRESS ALARM MAY RE- SULT IN FAILURE OF THE SEMS PDA DIS- TRESS ALARM DURING USE AND POS - SIBLE INJURY OR DEATH OF THE USER.

11

Image 11
Contents Sems Description Base Station Sensor ModulePage Detecting and Avoiding Radio Frequency Interference RFI Warning Areas Keep Radio Antennas AT LeastSIX 6 Inches from These Areas Programming FCC NoticeRadio Frequency Interference RFI Intrinsically Safe ListingBasic Functions of the Sems Equipment Portable Unit Operation of the Sems PDA Distress AlarmFull Alarm PRE AlarmTurning OFF the Sems PDA Distress Alarm Manual AlarmBattery Test Other Operational Features END of Service Time IndicatorUSE of the Sems PDA Portable Unit Evacuation Base Station USE of the Sems Base StationPage Event LOG Screen Event LOGSample Event LOG Regular Operational Inspection Regular Operational Inspection On Next Inspection of Sems PDA Distress AlarmInspection of Sems PDA Portable Unit Control Console Battery Cover Battery ReplacementSems PDA Distress Alarm Portable Unit Check Your Work Battery Compartment and Cover SealBattery Replacement on Next Control Console Battery Maintenance Battery Orientation AS ShownBattery Compartments on Base Station Does When the Sems PDA IT Indicates That Distress Alarm isSound Levels Performance SpecificationsRev a 6/03