OPERATION OF THE SEMS PDA DISTRESS ALARM
With proper batteries installed and a charged air cylinder, the SEMS PDA distress alarm (PASS) is automatically activated when the respi- rator is pressurized by opening the cylinder valve of the respirator. If batteries are completely discharged or have not been installed, there will be no light or sound indicating that the unit will not work at all.
Activation is indicated by 3 quick audible chirps from the Sensor Mod- ule. On activation, a green light located on the Control Console, shown in FIGURE 3, will begin to flash approximately once a second. The SEMS PDA distress alarm is now in the automatic mode.
WARNING
THE SEMS PDA DISTRESS ALARM IS IN- TENDED TO ASSIST IN LOCATING A PER- SON WHO MAY BE IN A LIFE THREATEN- ING SITUATION . FAILURE TO FOLLOW
WITHDRAW
BUTTON (BLUE
INDICATOR)
WITHDRAW
MANUAL
ALARM BUTTON
RESET
(RED INDICATOR)
BUTTON (YELLOW INDICATOR)
MANUAL | RESET |
ALARM |
T H E I N S T R U C T I O N S F O R O P E N I N G , CHANGING ALL BATTERIES AND
RED SIGNAL
LIGHT
CONSOLE
GREEN SIGNAL LIGHT
DISPLAY
WARNING
USERS OF RESPIRATORS EQUIPPED WITH THE SEMS PDA DISTRESS ALARM MUST BE AWARE OF THE PROPER OPERATION OF THE DISTRESS ALARM. FAILURE TO RECOG- NIZE A MALFUNCTION OF THE SEMS PDA DIS- TRESS ALARM AND TAKE PROPER COR- RECTIVE ACTION MAY RESULT IN SERIOUS IN- JURY OR DEATH.
FIGURE 3
THE PORTABLE UNIT
If the any of the batteries needs replacement, the SEMS PDA distress alarm will indicate as follows:
•SENSOR MODULE – A single audible chirp from the Sensor Mod- ule once every 2 seconds and the green light on the control mod- ule will not flash.
•CONTROL CONSOLE – “BATT” will appear in the display but all other functions will operate normally.
If any low battery condition appears at
8