Bayer HealthCare TS manual Error Codes and Symbols

Page 24

Solving

Error Codes and Symbols Problems

Error Codes and Symbols:

WHAT YOU SEE

WHAT IT MEANS

WHAT YOU SHOULD DO

 

Low Battery.

Replace the battery as soon

 

 

as possible (see page 26).

Temperature

Move to an area that is

out-of-range.

within the operating range

 

for the meter: 41°F–113°F.

 

Let the meter and test strips

 

adjust to this temperature

 

for twenty minutes before

 

testing.

The test strip has

Remove the test strip and

not filled enough

repeat the test using a new

to give an accurate

test strip.

result.

 

You may have

Do not attempt to apply

applied blood

blood more than once per

more than once

strip.

on the same

 

strip.

 

Blocked sample

Do not push strip into skin

tip.

while filling strip.

Blood drop too

See suggested drop size

small.

on page 10.

 

 

The meter is

Remove the test strip and

sensing a used

retest using a new test strip.

test strip.

Wait until you see the flash-

 

ing blood drop in the display

 

before adding the blood

 

sample.

22

Need Help? Call toll free: 1-800-348-8100

Image 24
Contents User Guide Welcome to Simplicity and Accuracy Table of Contents Important Your Contour TS Meter Display Meter Display ImportantMeter Features Important Meter FeaturesSymptoms of High or Low Blood Sugar Symptoms of High or Low Blood Sugar ImportantImportant Things to Know Important Important Things to KnowBefore You Begin Before You Begin ImportantTesting Preparing the Lancing Device Preparing the Lancing DeviceInserting the Test Strip Testing Inserting the Test StripTesting Getting the Blood Drop Getting the Blood DropTesting Your Blood Testing Your Blood TestingTesting Testing Your Blood Alternative Site Testing Alternative Site Testing TestingTesting Viewing Stored Test Results Viewing Stored Test ResultsTesting with a Control Solution Testing with a Control Solution TestingTesting Testing with a Control Solution Testing with a Control Solution Testing Setting the Date, Time and Sound Setting the Date Setting the Date, Time and Sound SetupSetting the Sound Setting the Month and Day Format Error Codes and Symbols Solving Problems Error Codes and Symbols Error Codes If you do not resolve the problem, contact BayerDiabetes Care Customer Service at Replacing the Battery Caring for the System Store meter in the carrying case provided whenever possibleTransferring Results to a Computer Weight 2 oz SpecificationsService Information Check ListWarranty Care and Extras WarrantyReferences Replacement Parts Supplies