Fluke 117, 115, 116, 114 specifications Introduction, Service Information

Page 1

®

114, 115, 116, and 117

Digital Multimeters

Calibration Information

Introduction

XWWarning

To avoid electric shock or injury, do not perform the performance tests or calibration adjustment procedures unless qualified to do so.

The information provided in this document is for the use of qualified personnel only.

The 114, 115, 116, & 117 Calibration Information provides the information necessary to adjust and verify the performance of the Fluke Models 114, 115, 115C, 116, 116C, 117, and 117C True RMS Multimeters (hereafter known as the Meter). When specific models are noted in this manual, the “C” version is also included. For example, when the listed model is 115, the instructions are applicable to the 115C as well.

The following information is included in this document:

Safety Information and International Electrical Symbols (page 2)

Specifications (page 3)

Testing the Fuse (page 5)

Replacing the Fuse and the Battery (page 6)

Cleaning (page 7)

Performance Tests (page 7)

Calibration Adjustment (page 12)

Replacement Parts and Accessories (page 17)

Complete Warranty (page 19)

See the 114, 115, 117 Users Manual or the 115C, 117C Users Manual or the 116 Users Manuals or the 116C Users Manual for complete operating instructions.

Service Information

To contact Fluke, call one of the following telephone numbers:

USA: 1-888-99-FLUKE (1-888-993-5853)

Canada: 1-800-36-FLUKE (1-800-363-5853)

Europe: +31 402-675-200

Japan: +81-3-3434-0181

Singapore: +65-738-5655

Anywhere in the world: +1-425-446-5500

Or, visit Fluke's Web site at www.fluke.com.

To register your product, visit register.fluke.com

September 2006 ©2006 Fluke Corporation. All rights reserved.

1

Image 1
Contents Introduction Service InformationXWWarnings and Precautions Safety InformationSpecifications General Specifications International Electrical SymbolsFunction Range Resolution Accuracy Model1 Accuracy SpecificationsInput Characteristics Basic MaintenanceTesting the Fuse 115 & 117 only Fuse Testing Replacing the Battery and FusePerformance Tests Cleaning the MeterRequired Equipment Backlight Test Testing the DisplayKeypad Test Preparing for the Performance TestsTesting Temperature 116 only 000V Testing the VoltAlert Function 117 only Calibration Adjustment Password Calibration AdjustmentXW Warning Meter Buttons Used in the Calibration Steps Calibration Adjustment ProcedureCOM 60 Hz + and COM 600 μA, 60 Hz AC μamps 114, 115, 116, and 117 Calibration Information Lists the Meter’s replacable parts identified in Figure Replacement PartsReplacable Parts List Description Qty Number Warranty 114, 115, 116, and 117 Calibration Information