Fluke specifications 114, 115, 116, and 117 Calibration Information

Page 16

114, 115, 116, and 117 Calibration Information

Table 4. Calibration Adjustment Steps (cont)

Rotary

 

Calibration Steps

 

Input

Calibrator

Switch

 

 

 

 

114

[1]

[1]

[1,2]

Terminals

Source Value

 

 

Position

115

116

117

 

 

 

 

 

 

 

 

$

 

N/A

N/A

C/23

N/A

+ and COM

600 μA, 0 Hz

DC μamps

 

 

 

 

 

 

 

 

 

 

 

 

 

 

[1]

Models listed in this column also refer to the “C” version of the model. For example, model 115 steps are valid for the 115C.

[2]

Do not calibrate the 117 or 117C with a line-frequency power source nearby (e.g. fluorescent light, power strip, etc.). These devices

 

can produce errors in the VoltAlert calibration.

 

 

 

 

[3]

Wait an additional 5 seconds after calibrator has settled before pressing g.

 

 

 

 

 

 

 

 

 

 

16

Image 16
Contents Service Information IntroductionSafety Information XWWarnings and PrecautionsInternational Electrical Symbols Specifications General SpecificationsAccuracy Specifications Function Range Resolution Accuracy Model1Input Characteristics Basic MaintenanceTesting the Fuse 115 & 117 only Replacing the Battery and Fuse Fuse TestingPerformance Tests Cleaning the MeterRequired Equipment Testing the Display Backlight TestKeypad Test Preparing for the Performance TestsTesting Temperature 116 only 000V Testing the VoltAlert Function 117 only Calibration Adjustment Calibration Adjustment PasswordXW Warning Calibration Adjustment Procedure Meter Buttons Used in the Calibration StepsCOM 60 Hz + and COM 600 μA, 60 Hz AC μamps 114, 115, 116, and 117 Calibration Information Replacement Parts Lists the Meter’s replacable parts identified in FigureReplacable Parts List Description Qty Number Warranty 114, 115, 116, and 117 Calibration Information