Fluke 116, 115, 117, 114 specifications Replacing the Battery and Fuse, Fuse Testing

Page 6

114, 115, 116, and 117

Calibration Information

VoltAle rt

117 TRUE RMS MULTIMETER

 

 

Lo / H

HOLD

MIN MAX

RANGE

AUTO-V

OFF

Hz

LoZ

 

V

 

 

V

 

 

mV

Volt

 

A Hz

Alert

A

V

A COM

<.5

OK

OK

10 A FUSED

erc010f.emf

Figure 1. Fuse Testing

Replacing the Battery and Fuse

XWWarning

To avoid shock, injury, or damage to the Meter:

Remove test leads from the Meter before opening the case or battery door.

Use ONLY a fuse with the amperage, interrupt voltage, and speed ratings specified.

Battery

Fuse

erc011f.emf

Figure 2. Battery and Fuse Replacement

To remove the battery door for battery replacement, refer to Figure 2 while performing the following:

1.Remove the test leads from the Meter.

2.Remove the battery door screw.

3.Use the finger recess to lift the door slightly.

4.Lift the door straight up to separate it from the case.

5.The battery fits inside the battery door, which is then inserted into the case, bottom edge first, until it is fully seated. Do not attempt to install the battery directly into the case.

6.Install and tighten battery door screw.

To open the case for fuse replacement, refer to Figure 2 while performing the following: 1. Remove the test leads from the Meter.

6

Image 6
Contents Service Information IntroductionSafety Information XWWarnings and Precautions International Electrical Symbols Specifications General SpecificationsAccuracy Specifications Function Range Resolution Accuracy Model1Basic Maintenance Input CharacteristicsTesting the Fuse 115 & 117 only Replacing the Battery and Fuse Fuse TestingCleaning the Meter Performance TestsRequired Equipment Keypad Test Testing the DisplayBacklight Test Preparing for the Performance TestsTesting Temperature 116 only 000V Testing the VoltAlert Function 117 only Calibration Adjustment Calibration Adjustment PasswordXW Warning Calibration Adjustment Procedure Meter Buttons Used in the Calibration StepsCOM 60 Hz + and COM 600 μA, 60 Hz AC μamps 114, 115, 116, and 117 Calibration Information Replacement Parts Lists the Meter’s replacable parts identified in FigureReplacable Parts List Description Qty Number Warranty 114, 115, 116, and 117 Calibration Information