Fluke 114, 115, 117, 116 International Electrical Symbols, Specifications General Specifications

Page 3

Digital Multimeters

International Electrical Symbols

International Electrical Symbols

Table 1 lists the international symbols that appear in this document and on the Meter.

Table 1. Electrical Symbols

Symbol

Description

Symbol

Description

 

 

 

 

B

AC (Alternating Current)

I

Fuse

 

 

 

 

F

DC (Direct Current)

T

Double Insulated

 

 

 

 

X

Hazardous voltage

W

Important Information; Refer to manual

 

 

 

 

N

Battery (Low battery when shown on the

J

Earth ground

 

display.)

 

 

 

 

 

 

 

~

Do not dispose of this product as unsorted municipal waste. Contact Fluke or a qualified recycler for

 

disposal.

 

 

 

 

 

 

Specifications

Accuracy is specified for 1 year after calibration at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Extended specifcifications are available at www.fluke.com.

General Specifications

Maximum voltage between any

 

terminal and earth ground

600 V

Surge Protection

6 kV peak per IEC 61010-1 600V CAT III,

 

Pollution Degree 2

WFuse for A input (115 & 117 only):.................. 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293)

Display

Digital: 6,000 counts, updates 4/sec

 

Bar Graph: 33 segments, updates 32/sec

Temperature

Operating: -10 °C to +50 °C

 

Storage: -40 °C to +60 °C

Temperature Coefficient

0.1 x (specified accuracy)/°C (<18 °C or >28 °C)

Operating Altitude

2,000 meters

Battery

9 Volt Alkaline, NEDA 1604A / IEC 6LR61

Battery Life

Alkaline: 400 hours typical, without backlight

Safety Compliances

Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 No 61010-1-04, UL 6101B

 

(2003) and IEC/EN 61010-1 2nd Edition for measurement Category III, 600 V, Pollution

 

Degree 2, EMC EN61326-1

Certifications

UL, P, CSA, TÜV, ; (N10140), VDE

IP Rating (dust and water protection)

IP42

3

Image 3
Contents Introduction Service InformationXWWarnings and Precautions Safety InformationSpecifications General Specifications International Electrical SymbolsFunction Range Resolution Accuracy Model1 Accuracy SpecificationsBasic Maintenance Input CharacteristicsTesting the Fuse 115 & 117 only Fuse Testing Replacing the Battery and FuseCleaning the Meter Performance TestsRequired Equipment Preparing for the Performance Tests Testing the DisplayBacklight Test Keypad TestTesting Temperature 116 only 000V Testing the VoltAlert Function 117 only Calibration Adjustment Password Calibration AdjustmentXW Warning Meter Buttons Used in the Calibration Steps Calibration Adjustment ProcedureCOM 60 Hz + and COM 600 μA, 60 Hz AC μamps 114, 115, 116, and 117 Calibration Information Lists the Meter’s replacable parts identified in Figure Replacement PartsReplacable Parts List Description Qty Number Warranty 114, 115, 116, and 117 Calibration Information