Fluke 115, 117, 116, 114 Accuracy Specifications, Function Range Resolution Accuracy Model1

Page 4

114, 115, 116, and 117 Calibration Information

Accuracy Specifications

Function

 

Range

 

Resolution

Accuracy

Model[1]

 

 

 

 

 

± ([% of Reading] + [Counts])

 

DC millivolts

600.0 mV

 

0.1 mV

0.5 % + 2

114, 115, 116, 117

 

6.000

V

 

0.001 V

 

 

 

 

 

DC Volts

60.00

V

 

0.01 V

0.5 % + 2

114, 115, 116, 117

 

600.0

V

 

0.1 V

 

 

 

 

 

 

 

 

 

 

DC, 45 to 500 Hz

 

500 Hz to 1 kHz

 

Auto-V LoZ[2] True-rms

600.0

V

 

0.1 V

2.0 % + 3

 

 

4.0 % + 3

114, 116, 117

 

 

 

 

 

45 to 500 Hz

 

500 Hz to 1 kHz

 

AC millivolts[2] True-rms

600.0 mV

 

0.1 mV

1.0 % + 3

 

 

2.0 % + 3

114, 115, 116, 117

 

6.000

V

 

0.001 V

 

 

 

 

 

AC Volts[2] True-rms

60.00

V

 

0.01 V

1.0 % + 3

 

 

2.0 % + 3

114, 115, 116, 117

 

600.0

V

 

0.1 V

 

 

 

 

 

Continuity

600 Ω

 

 

1 Ω

Beeper on <20 Ω, off >250 Ω; detects opens

114, 115, 116, 117

 

 

 

 

 

or shorts of 500 μs or longer.

 

 

600.0

Ω

 

0.1 Ω

0.9 % + 2

 

 

6.000 kΩ

 

0.001 kΩ

0.9 % + 1

 

Ohms

60.00 kΩ

 

0.01 kΩ

0.9 % + 1

114, 115, 116, 117

600.0 kΩ

 

0.1 kΩ

0.9 % + 1

 

 

 

 

6.000 MΩ

 

0.001 MΩ

0.9 % + 1

 

 

40.00 MΩ

 

0.01 MΩ

1.5 % + 2

 

Diode test

2.000

V

 

0.001 V

0.9 % + 2

115, 116, 117

Capacitance

1000 nF

 

1 nF

1.9 % + 2

115, 116, 117

 

10.00

μF

 

0.01 μF

1.9 % + 2

 

 

100.0

μF

 

0.1 μF

1.9 % + 2

 

 

9999 μF

 

1 μF

100 μF - 1000 μF: 1.9 % + 2

 

 

 

 

 

 

>1000 μF: 5 % + 20

 

Lo-Z Capacitance (Power-

 

1 nF to 500

μF

20% + 2

115, 116, 117

up option)

 

(10% +2

typical)

 

 

 

 

 

 

 

6.000

A

0.001 A

 

 

 

 

 

AC Amps True-rms[2]

10.00

A[4]

0.01 A

 

 

 

 

 

20 A overload for 30

 

1.5 % + 3

115, 117

(45 Hz to 500 Hz)

 

seconds maximum, 10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

minutes rest minimum.

 

 

 

 

 

 

 

6.000

A

0.001 A

 

 

 

 

 

 

10.00

A[4]

0.01 A

 

 

 

 

 

DC Amps

20 A overload for 30

 

1.0 % + 3

115, 117

 

seconds maximum, 10

 

 

 

 

 

 

 

minutes rest minimum.

 

 

 

 

 

 

Temperature (Type K

-40 °C to 400 °C

0.1 °C

1.0 % + 10[5]

116

thermocouple)

-40 °F to 752 °F

0.2°F

1.0 % + 18[5]

 

AC μAmps True-rms[2]

600.0

μA

0.1 μA

1.5 % + 3 (2.5 % + 3 > 500 Hz)

116

(45 Hz to 1 kHz)

 

 

 

 

 

 

 

 

 

 

 

 

 

DC μAmps

600.0

μA

0.1 μA

1.0 % + 2

116

4

Image 4
Contents Service Information IntroductionSafety Information XWWarnings and PrecautionsInternational Electrical Symbols Specifications General SpecificationsAccuracy Specifications Function Range Resolution Accuracy Model1Input Characteristics Basic MaintenanceTesting the Fuse 115 & 117 only Replacing the Battery and Fuse Fuse TestingPerformance Tests Cleaning the MeterRequired Equipment Testing the Display Backlight TestKeypad Test Preparing for the Performance TestsTesting Temperature 116 only 000V Testing the VoltAlert Function 117 only Calibration Adjustment Calibration Adjustment PasswordXW Warning Calibration Adjustment Procedure Meter Buttons Used in the Calibration StepsCOM 60 Hz + and COM 600 μA, 60 Hz AC μamps 114, 115, 116, and 117 Calibration Information Replacement Parts Lists the Meter’s replacable parts identified in FigureReplacable Parts List Description Qty Number Warranty 114, 115, 116, and 117 Calibration Information