Fluke 117, 115, 114 specifications Testing Temperature 116 only

Page 9

Digital Multimeters

Performance Tests

2.Warm up the calibrator as required by its specifications.

3.Allow the temperature of the UUT to stabilize at room temperature (23 °C ± 5 °C [73 °F ± 9 °F] ).

4.Check the fuses and Battery, and replace them if necessary. Refer to “Testing the Fuses”, and “Replacing the Battery and Fuse”.

To verify the accuracy of the DMM functions, do the following:

1.Connect the Calibrator to the VΩ and COM input terminals on the Meter.

2.Turn the rotary switch to the function listed in each step of Table 3.

3.Apply the input level for each step listed in Table 3.

4.Compare the reading on the Meter display with the Display Reading in Table 3.

5.If the display reading falls outside of the range shown in Table 3, the Meter requires calibration adjustment or repair.

Testing Temperature (116 only)

Connect the K-type thermocouple to the temperature input of the Meter and temperature calibrator. To ensure an accurate measurement, the Meter and the thermocouple connector must be at the same temperature. After connecting the thermocouple to the Meter, allow the junctions to stabilize before recording the displayed reading. This can take several minutes, depending on temperature gradients.

Table 3. DMM Performance Tests

Step

Function

Range

Applied

 

Display Reading

 

 

 

 

 

 

 

114[1]

115[1]

 

116[1]

117[1]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

e

600.0

 

0.0 Ω

 

 

0.0 to 0.2

 

 

Ohms

 

 

 

 

 

 

 

 

 

2

600.0

 

500 Ω

 

495.3 to 504.7

 

 

 

 

 

 

 

 

 

 

 

 

3

 

6.000 k

5 kΩ

 

4.954 to 5.046

 

 

 

 

 

 

 

 

4

 

60.00 k

50 kΩ

 

49.54 to 50.46

 

 

 

 

 

 

 

 

5

 

600.0 k

500 kΩ

 

495.4 to 504.6

 

 

 

 

 

 

 

 

 

6

 

6.000

M

5 MΩ

 

4.954 to 5.046

 

 

 

 

 

 

 

 

 

7

 

40.00

M

10 MΩ

 

9.83 to 10.17

 

 

 

 

 

 

 

 

 

8

 

40.00

M

30 MΩ

 

29.53 to 30.47

 

 

 

 

 

 

 

 

 

 

9

R

600 Ω

 

20 Ω

 

 

Beeper On

 

 

Continuity

 

 

 

 

 

 

 

 

 

10

600 Ω

 

250 Ω

 

 

Beeper Off

 

 

 

 

 

 

 

 

 

 

 

 

 

 

11

K

6.000

V

5V, 45 Hz

 

4.947 to 5.053

 

 

AC Volts

 

 

 

 

 

 

 

 

 

12

6.000

V

5V, 1 kHz

 

4.897 to 5.103

 

 

 

 

 

 

 

 

 

 

 

 

13

 

60.00

V

50 V, 45 Hz

 

49.47 to 50.53

 

 

 

 

 

 

 

 

 

14

 

60.00

V

50V, 1 kHz

 

48.97 to 51.03

 

 

 

 

 

 

 

 

 

15

 

600.0

V

600V, 45 Hz

 

593.7 to 606.3

 

 

 

 

 

 

 

 

 

16

 

600.0

V

600V, 1kHz

 

587.7 to 612.3

 

 

 

 

 

 

 

 

 

 

 

17

e

6.000V

2 v, 45 kHz[1]

NA

 

 

 

49.93 to 50.07

 

 

AC Volts + Hz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

9

Image 9
Contents Introduction Service InformationXWWarnings and Precautions Safety InformationSpecifications General Specifications International Electrical SymbolsFunction Range Resolution Accuracy Model1 Accuracy SpecificationsBasic Maintenance Input CharacteristicsTesting the Fuse 115 & 117 only Fuse Testing Replacing the Battery and FuseCleaning the Meter Performance TestsRequired Equipment Backlight Test Testing the DisplayKeypad Test Preparing for the Performance TestsTesting Temperature 116 only 000V Testing the VoltAlert Function 117 only Calibration Adjustment Password Calibration AdjustmentXW Warning Meter Buttons Used in the Calibration Steps Calibration Adjustment ProcedureCOM 60 Hz + and COM 600 μA, 60 Hz AC μamps 114, 115, 116, and 117 Calibration Information Lists the Meter’s replacable parts identified in Figure Replacement PartsReplacable Parts List Description Qty Number Warranty 114, 115, 116, and 117 Calibration Information