Fluke 114, 115, 117, 116 specifications COM 60 Hz + and COM 600 μA, 60 Hz AC μamps

Page 15

Digital Multimeters

Calibration Adjustment

Notes

Set the calibrator to Standby prior to changing the function switch position and after completing adjustment of each function.

If the calibration adjustment procedure is not properly completed, the Meter will not operate correctly.

Table 4. Calibration Adjustment Steps

Rotary

 

 

Calibration Steps

 

Input

Calibrator

Switch

 

 

 

 

 

 

114

[1]

[1]

[1,2]

Terminals

Source Value

 

 

Position

 

115

116

117

 

 

 

 

 

 

 

 

 

e

 

N/A

C/01

C/01

C/01[2]

No leads

No leads

Ohms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

m

 

C/01

C/02

C/02

C/02

VΩ/+ and COM

0 V, 0 Hz

 

 

 

 

 

 

 

 

 

 

C/02

C/03

C/03

C/03

VΩ/+ and COM

300 mV, 0 Hz

 

 

 

 

 

 

 

 

 

 

C/03

C/04

C/04

C/04

VΩ/+ and COM

100 mV, 0 Hz

 

 

 

 

 

 

 

 

 

 

C/04

C/05

C/05

C/05

VΩ/+ and COM

-300 mV, 0 Hz

 

 

 

 

 

 

 

 

 

 

C/05

C/06

C/06

C/06

VΩ/+ and COM

60 mV, 0 Hz

 

 

 

 

 

 

 

 

 

 

C/06

C/07

C/07

C/07

VΩ/+ and COM

600 mV, 0 Hz

 

 

 

 

 

 

 

 

 

 

C/07

C/08

C/08

C/08

VΩ/+ and COM

600 mV, 60 Hz

 

 

 

 

 

 

 

 

e

 

C/08

C/09

C/09

C/09

VΩ/+ and COM

600 e, 2-wire comp

Ohms

 

 

 

 

 

 

 

 

C/09

C/10

C/10

C/10

VΩ/+ and COM

6 ke

 

 

 

 

 

 

 

 

 

 

 

 

C/10

C/11

C/11

C/11

VΩ/+ and COM

60 ke

 

 

 

 

 

 

 

 

 

 

C/11

C/12

C/12

C/12

VΩ/+ and COM

600 ke

 

 

 

 

 

 

 

 

 

 

C/12

C/13

C/13

C/13

VΩ/+ and COM

6 Me[3]

 

 

 

 

 

 

 

 

 

 

C/13

C/14

C/14

C/14

VΩ/+ and COM

Short[3]

 

 

 

 

 

 

 

 

 

 

C/14

C/15

C/15

C/15

VΩ/+ and COM

40 Me[3]

 

 

 

 

 

 

 

 

K

 

C/15

C/16

C/16

C/16

VΩ/+ and COM

6 V, 60 Hz

 

 

 

 

 

 

 

 

 

 

C/16

C/17

C/17

C/17

VΩ/+ and COM

60 V, 60 Hz

 

 

 

 

 

 

 

 

 

 

C/17

C/18

C/18

C/18

VΩ/+ and COM

600 V, 60 Hz

 

 

 

 

 

 

 

 

 

 

C/18

C/19

C/19

C/19

VΩ/+ and COM

6 V, 0 Hz

 

 

 

 

 

 

 

 

 

 

C/19

C/20

C/20

C/20

VΩ/+ and COM

60 V, 0 Hz

 

 

 

 

 

 

 

 

 

 

C/20

C/21

C/21

C/21

VΩ/+ and COM

600 V, 0 Hz

 

 

 

 

 

 

 

 

 

Set calibrator to standby, reconfigure leads, and program for amps output.

 

 

 

 

 

 

 

 

?

 

N/A

C/22

N/A

C/22

A and COM

6 A, 60 Hz[3]

 

 

 

 

 

 

 

 

A

 

N/A

C/23

N/A

C/23

A and COM

6 A, 0 Hz

 

 

 

 

 

 

 

 

$

 

N/A

N/A

C/22

N/A

+ and COM

600 μA, 60 Hz

AC μamps

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

15

Image 15
Contents Introduction Service InformationXWWarnings and Precautions Safety InformationSpecifications General Specifications International Electrical SymbolsFunction Range Resolution Accuracy Model1 Accuracy SpecificationsBasic Maintenance Input CharacteristicsTesting the Fuse 115 & 117 only Fuse Testing Replacing the Battery and FuseCleaning the Meter Performance TestsRequired Equipment Preparing for the Performance Tests Testing the DisplayBacklight Test Keypad TestTesting Temperature 116 only 000V Testing the VoltAlert Function 117 only Calibration Adjustment Password Calibration AdjustmentXW Warning Meter Buttons Used in the Calibration Steps Calibration Adjustment ProcedureCOM 60 Hz + and COM 600 μA, 60 Hz AC μamps 114, 115, 116, and 117 Calibration Information Lists the Meter’s replacable parts identified in Figure Replacement PartsReplacable Parts List Description Qty Number Warranty 114, 115, 116, and 117 Calibration Information