Fluke 115, 117, 116, 114 specifications Calibration Adjustment Password

Page 12

114, 115, 116, and 117

Calibration Information

NORMAL

AUX

 

 

V

RTD

 

SC

 

A SENSE A

UX V

 

 

 

OPE

 

 

 

 

O

 

 

 

 

UT

TRIG

STBY

7 8

4 5

.

2

+/- 0

.

1 1 7 T R U E R M S M U L T I M E T E R

VoltAlert

Insert Double Banana

Center Meter top on Hi terminal.

 

Note that LED is RED

 

 

 

 

erc013f.emf

 

Figure 4. VoltAlert Testing

8.Hold the Meter so that the top is vertically and horizontal centered to the banana plug’s Hi terminal. Verify that the Meter’s beeper is sounding continuously and the red LED at the top of the display lights.

9.Return the calibrator to standby mode and verify the Meter’s beeper is silent and the red LED is off.

Calibration Adjustment

The Meter features closed-case calibration adjustment using known reference sources. The Meter measures the applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory.

The following sections present the features and Meter pushbutton functions available during the Calibration Adjustment Procedure. Should the Meter fail any of the performance tests, perform the Calibration Adjustment Procedure.

Use the following steps to view the Meter’s calibration counter.

1.While pressing K, turn the rotary switch from OFF to Ω function. The Meter should display “ZCAL”.

2.Press g once to view the calibration counter. For example, “n001

3.Turn the rotary switch to OFF.

Calibration Adjustment Password

To start the Calibration Adjustment Procedure, the correct 4-digit password must be entered. The default password is “1234”. The password can be changed or reset to the default as described in following paragraphs.

Changing the Password

Use the following steps to change the Meter’s password:

1.While pressing K, turn the rotary switch from OFF to Ω function. The Meter should display “ZCAL”.

12

Image 12
Contents Service Information IntroductionSafety Information XWWarnings and PrecautionsInternational Electrical Symbols Specifications General SpecificationsAccuracy Specifications Function Range Resolution Accuracy Model1Basic Maintenance Input CharacteristicsTesting the Fuse 115 & 117 only Replacing the Battery and Fuse Fuse TestingCleaning the Meter Performance TestsRequired Equipment Testing the Display Backlight TestKeypad Test Preparing for the Performance TestsTesting Temperature 116 only 000V Testing the VoltAlert Function 117 only Calibration Adjustment Calibration Adjustment PasswordXW Warning Calibration Adjustment Procedure Meter Buttons Used in the Calibration StepsCOM 60 Hz + and COM 600 μA, 60 Hz AC μamps 114, 115, 116, and 117 Calibration Information Replacement Parts Lists the Meter’s replacable parts identified in FigureReplacable Parts List Description Qty Number Warranty 114, 115, 116, and 117 Calibration Information