S A F E T Y
I
N S P E C T I O N
T R O U B L E S H O O T I N G
PROCEDURE 2 | SAFETY INSPECTION/TROUBLESHOOTING | |
|
|
|
This Procedure Includes the Following:
Troubleshooting
Checking Battery Charge Level
Safety Inspection Checklist
TROUBLESHOOTING - MECHANICAL
CHAIR VEERS | SLUGGISH TURN/ | CASTERS | SQUEAKS AND | LOOSENESS | CHAIR 3 | SOLUTIONS |
LEFT/RIGHT | PERFORMANCE | FLUTTER | RATTLES | IN CHAIR | WHEELS |
|
X | X | X |
|
| X | If pneumatic, check tires for cor- |
|
|
|
|
|
| rect and equal pressure. |
X | X | X | X | X |
| Check for loose stem nuts/bolts. |
X |
| X |
|
|
| Check that both casters contact |
|
|
|
|
|
| ground at the same time. |
|
|
|
|
|
|
|
TROUBLESHOOTING - ELECTRICAL
SYMPTOM | PROBABLE CAUSE | SOLUTIONS | |
Battery won't charge | Blown Fuse. (RANGER IIJR ONLY) | Replacefuse(PROCEDURE6).(RANGERIIJR ONLY) | |
Batteries draw excessive current | Battery failure. | Check batteries for shorted cell. Replace if nec- | |
when charging. |
| essary | |
|
| DURE 12). | |
| Electrical malfunction. | Contact Dealer/Invacare for Service. | |
Battery indicator flashes the charge | Battery failure. | Check batteries for shorted cell. Replace if nec- | |
level is |
| essary | |
charge. |
| DURE 12). | |
| Malfunctioning battery charger. | Contact Dealer/Invacare for Service. | |
| Electrical malfunction. | Poor connections between charger and wheel- | |
|
| chair. Contact Dealer/Invacare for Service. | |
|
|
| |
Battery indicator flashes the charge | Batteries not charged. | Have charger checked. | |
level is | Weak batteries. | Replace batteries if necessary | |
recharged. | |||
| DURE 11, | ||
|
| ||
| Electrical malfunction. | Contact Dealer/Invacare for Service. | |
Motor “chatters” or runs irregular. | Electrical malfunction. | Contact Dealer/Invacare for Service. | |
Wheelchair does not respond to | One (1) or both motor locks/clutches | Engage motor locks/clutches (PROCEDURE 8). | |
commands. Power "ON", battery | disengaged. |
| |
indicator flashes. |
|
| |
Only one (1) rear wheel turns. | Electrical malfunction. | Contact Dealer/Invacare for Service. | |
| One (1) motor lock/clutch is disengaged. | Engage motor lock/clutch (PROCEDURE 8). | |
|
|
| |
Joystick erratic or does not re- | Electrical malfunction. | Contact Dealer/Invacare for Service. | |
spond as desired. | Controller Programed improperly. | Contact Dealer/Invacare for Service. | |
| |||
|
|
| |
Wheelchair does not respond to | Poor battery terminal connection. | Clean terminals | |
commands. Power indicator |
| PROCEDURE 12). | |
Electrical malfunction. | Contact Dealer/Invacare for Service. | ||
| |||
|
|
|
NOTE: For additional troubleshooting information and explanation of error codes, refer to the individual CONTROL- LER MANUAL supplied with each wheelchair.
CHECKING BATTERY CHARGE LEVEL
The following “Do’s” and “Don’ts” are provided for your convenience and safety.
DON’T | DO |
Don’t perform any installation or maintenance | Read and understand this manual and any service information that accompa- |
without first reading this manual. | nies a battery and charger before operating the wheelchair. |
Don’t perform installation or maintenance of | Move the wheelchair to a work area before checking the fluid level, adding |
batteries in an area that could be damaged by | distilled water, cleaning terminals, or opening battery box. |
battery spills. |
|
|
|
|
|
18