Leica 106Z user manual Observe the safety regulations on, Fuses

Page 14

Assembly

Supply unit for lamp housing 106Z with 50W mercury-vapour burner

1.On/off switch

2.Running-time meter

3.Selector for L1/L2 burners

4.Fuse

5.Selector for 50Hz/60Hz

6.

Connector for lamp

2

7.

Power cable

 

The supply unit is intended for power supplies above 220V..

To run it on lower voltages a step-down transformer is required (e..g.. 230V to 110V)..

Observe the safety regulations on page 5.

Connect the cable from the lamp housing to the supply unit

Engage “L1” or “L2” in accordance with the inscription

on the burner­ socket (see page 12)

Engage 50Hz or 60Hz

Connect the power cable to the supply unit and to the grid

Make a note of the running-time meter reading..

Supply unit for lamp housing 106Z with 100W mercury-vapour burner

1.On/off switch

2.Pilot lamp LAMP

3.Pilot lamp TEMP

4.Running-time meter

5.Pilot lamp SAFETY

3 4 5

1

6

7

Input type 220V/240V ±10% 220V ~50/60Hz Hg50W L1/L2 Osram no.. 0-958 Max.. power 320VA

1

4

2

5

3

Observe the safety regulations­ on page 5

Connect the cable from the lamp housing to the supply unit

Secure the lamp cable..

Connect the power cable to the supply unit and to the grid

Make a note of the running-time meter reading..

14

Type

ebq 100dc

Input voltage

90V to 250V AC, max 265VA

Input frequency

48Hz to 63 Hz

Protection class

IP20

Series no..

G 34240090

Fuses

The fuse holder is secured in the holder with spring clips on both sides..

Insert a screwdriver behind the spring clips and extract the fuse holder..

Insert the fuses (2T3,15A H250V F1/F2).

Refit the fuse holder

Image 14
Contents Leica Stereo Fluorescence Systems Page Contents Leica MZ10 F Fluorescence stereomicroscope Leica stereo-fluorescence moduleSafety concept DescriptionLight source Safety regulations Applications Natural science Application Fluorescence filter setIndustry Application Fluorescence filter set 3c 4a OverviewLeica MZ10 F Its components and controls Barrier filter Filter sets for Leica MZ10 FFilter sets Designation Assembly Double-iris diaphragm Leica MZ10 FAdditional components Video-/phototube UV protection screen microscope carrierFitting and changing the mercury-vapour burner AssemblyObserve the safety regulations­ on Removing the collector and the heat-protection filter Lamp housing assembleStray-light protection Replacing the collectorObserve the safety regulations on FusesAdjusting the mercury vapour lamp Figs.. a Bringing the discharge arc into focusFluoiii filter system Reflexes hot-spots UseDioptric correction Outfit with stereo-fluorescence module Components 5a, 5b 4a 4b, 4c, 4dFilter sets for stereo-fluorescence module 13,14 Video-/phototubes Stereo-fluorescence module Focusing drive base of standFluorescence module stereomicroscope Changing the filter setFilterholder fluorescence module Additional assembly instructionsUV-light excluder side-faced column With the fluorescence illumination Stereo-fluorescence module Focusing and observingFocusing Transmitted light ObservationLeica Microsystems the brand for outstanding products