Servicing and repairs
–Repairs may only be carried out by
–Only original Leica spare parts may be used
–Unplug the power cable before opening the supply unit.. Touching live circuits can cause injury..
Legal requirements
Adhere to general and local regulations relating to accident prevention and environmental protection..
Conformity with European Community directive
Leica
Supply unit
•Pull the power plug of the supply unit out from the power soc- ket:
–when assembling and dismantling the lamp housing
–before opening the lamp housing
–when changing the
–when servicing the supply unit
Lamp housing
•Never open the lamp housing while the burner is switched on (risk of explosion, UV irradiation, dazzling)
•Always allow the lamp housing to cool for at least 15 minutes before opening it (risk of burner exploding)
•Never cover the ventilator slots on the lamp housing (risk of fire)
|
|
| |
| • Read the user manual and safety directions provided by the | ||
| manufacturer of the | ||
| those relating to its breakage and to the release of mercury | ||
Light source: Safety regulations | • Before transporting the equipment, remove the | ||
| burner and place it in its original packaging.. Use the transport | ||
Safety measures introduced by manufacturer | peg in place of the burner, to secure movable parts within the | ||
lamp housing |
| ||
– UV protection screen in front of the specimen plane prevents | • When the | ||
direct UV radiation from reaching the eyes.. | nominal working life as indicated by the manufacturer’s | ||
– Dummy filter carriers in the unoccupied positions of the rapid | information and the time counter on the supply unit, change the | ||
discoloured burner (increased risk of explosion) | |||
filter changer prevent direct UV radiation from reaching | |||
• Leica declines all responsibility for injury and damage | |||
the eyes | |||
– UV filters in the observation beam paths protect the eyes | resulting from exploding, incorrectly |
against UV radiation.. | |
| |
– |
|
protects the user's hands against UV radiation.. |
|
Warning
UV can damage your eyes.. Therefore:
–Never look at the light spot on the specimen plane without the UV protection screen in position
–Never look into the eyepieces unless there is an excitation filter in the beam path
–Always keep dummy filter carriers in the unoccupied positions of the rapid filter changer (for MZ10 F)
–Do not place the specimen on a white or
5