Heat from the process is transferred to the transmitter housing. If the process temperature is high, the ambient temperature will need to be lower to account for heat transferred to the transmitter housing. See Table 1-1for process temperature derating.

Table 1-1. Temperature Derating

Process Temperature (°C)

Process Temperature (°F)

Max Ambient (°C)

Max Ambient (°F)

260

500

41

105.8

240

464

45

113

220

428

49

120.2

200

392

53

127.4

180

356

57

134.6

160

320

61

141.8

140

284

64

147.2

120

248

68

154.4

100

212

72

161.6

85

185

75

167

To expedite the return process outside of North America, contact your Emerson Process Management representative.

Within the United States, call the Emerson Process Management Response Center toll-free number 1 800 654 7768. The center, which is available 24 hours a day, will assist you with any needed information or materials.

The center will ask for product model and serial numbers, and will provide a Return Material Authorization (RMA) number. The center will also ask for the process material to which the product was last exposed.

Individuals who handle products exposed to a hazardous substance can avoid injury if they are informed of, and understand, the hazard. If the product being returned was exposed to a hazardous substance as defined by OSHA, a copy of the required Material Safety Data Sheet (MSDS) for each hazardous substance identified must be included with the returned goods.

SHIPPING CONSIDERATIONS FOR WIRELESS PRODUCTS (LITHIUM BATTERIES):

The unit was shipped to you without the power module installed. Please remove the power module prior to shipping the unit.

Primary lithium batteries (charged or discharged) are regulated during transportation by the U.S. Department of Transportation. They are also covered by IATA (International Air Transport Association), ICAO (International Civil Aviation Organization), and ARD (European Ground Transportation of Dangerous Goods). It is the responsibility of the shipper to ensure compliance with these or any other local requirements. Consult current regulations and requirements before shipping.

1-5

Image 11
Contents Reference Manual Rosemount 708 Wireless Acoustic TransmitterPage International Rosemount 708 Wireless Acoustic TransmitterCustomer Central National Response CenterExplosions could result in death or serious injury Shipping considerations for wireless productsTable of Contents Appendix C Appendix aAppendix B Section Overview Safety MessagesGeneral OverviewTransmitter ConsiderationsRecommended Practices Wireless ConsiderationsPower Up Sequence Antenna PositionEnvironmental MechanicalElectrical Service Support RECYCLING/DISPOSAL ProductSafety Messages Device Configuration Section ConfigurationField Communicator Device Network ConfigurationAMS Wireless Configurator AMSFast Key Sequence Remove Power ModuleField Communicator Menu Tree Hart Menu TreeMounting SectionTransmitter Alignment Piping, Insulation Side View Mounting ConsiderationsRosemount Verify Operation CommissioningAMS Suite Wireless Configurator Smart Wireless Gateway ExplorerSmart Wireless Gateway Network Settings TroubleshootingKey Function Sequence Menu Items Field Communicator UseSafety Messages Operation Operation and MaintenanceAMS Device Manager Overview Screen Normal Conditions AlertsAlert Levels Device Alert ConfigurationAlert Setup Alert Setup Install and LearnAlert Setup Manual Alert Setup Guided User Alert SetupEnvironmental Considerations Power Module ReplacementHandling Considerations Shipping Considerations Overview Safety Messages TroubleshootingWireless Troubleshooting Symptom Device Troubleshooting Corrective ActionsConfiguration Troubleshooting Functional Specifications Physical Specifications Appendix a Specifications and Reference DataSpecifications Performance Specifications Wireless Output Specifications Rosemount 708 Direct Mount Dimensional DrawingsOrdering Information Appendix B Product Certifications Standards Hazardous Locations CertificatesNorth American Certifications Special Conditions for Safe UseSafe Use Conditions IECEx System CertificationsBrazilian Certifications CMaster-PRINTEDCOPIESAREUNCONTROLLED- RosemountProprietary Effective Range Appendix C Recommended PracticesRosemount Page Asia Pacific Private Limited 00809-0100-4708, Rev AC September