Reference Manual
Rosemount 708
Device Alert Configuration
Alert Levels
The Rosemount 708 provides 4 user configurable alerts.
Acoustic High - Triggers when acoustic level exceeds a user configured threshold.
Acoustic Low - Triggers when acoustic level falls below a user configured threshold.
Temperature High - Triggers when temperature level exceeds a user configured threshold.
Temperature Low - Triggers when temperature level falls below a user configured threshold.
Alert Setup
There are three methods that can be used to set up these alerts using AMS Wireless Configurator or a field communicator.
Figure 5-3. Alert Setup screen
Alert Setup - Install and Learn
To use the install and learn function on the device to setup the alerts the device must be installed and the equipment being monitored must be in its normal state. After installation and operating conditions have been confirmed, go to the configuration menu under guided setup and click on “alert setup.” This will launch a dialog menu (Figure
If the input conditions are too close to the high or low measurement limits of the device, the install and learn function will not be a good method to set the alert levels. If this is the case, it is recommended that installation conditions be checked and that the status is verified before moving ahead. If current input conditions do not allow for the install and learn function, use the manual or guided alert setup to set alerts.