Page 103
microphone.
2.2Broad-band signals
2.2.1Pink Noise signal
The pink noise signal is available for Multicurve 2 cm3 coupler gain and output frequency response measurements and for Insertion gain and Speechmap real-ear measurements. It is a pseudo-random signal composed of 1024 simultaneous tones summed to provide a crest factor of 12 dB. The spectrum of the pink noise signal is controlled by the reference microphone in conjunction with a digital feedback loop with a frequency resolution of 1/12 octave and a response time of about 1/3 of a second. Pink noise was selected as a test signal because it has equal energy per octave, producing a flat spectrum when analyzed in 1/12 or 1/3 octave bands. Figure 1 shows the 1/3 octave spectrum of the pink noise signal and the noise signal specified in ANSI S3.42-1997. Note that the ANSI S3.42-1997 spectrum represents speech peaks not the long-term average.
1/3 OCTAVE SPECTRA for the PINK NOISE SIGNAL
and ANSI S3.42 NOISE re overall SPL
-8.00
Band Level re overal SPL - Pink Noise
| -9.00 | | | | | | | | | | | Band Level re overall SPL - ANSI S3.42 |
| | | | | | | | | | | | | | | | |
| -10.00 | | | | | | | | | | | | | | | | |
SPL) | -11.00 | | | | | | | | | | | | | | | | |
| | | | | | | | | | | | | | | | |
re overall | -12.00 | | | | | | | | | | | | | | | | |
-13.00 | | | | | | | | | | | | | | | | |
SPL (dB | | | | | | | | | | | | | | | | |
-14.00 | | | | | | | | | | | | | | | | |
BAND | -15.00 | | | | | | | | | | | | | | | | |
| | | | | | | | | | | | | | | | |
| -16.00 | | | | | | | | | | | | | | | | |
| -17.00 | | | | | | | | | | | | | | | | |
| -18.00 | | | | | | | | | | | | | | | | |
| 200 | 250 | 315 | 400 | 500 | 630 | 800 | 1000 | 1250 | 1600 | 2000 | 2500 | 3150 | 4000 | 5000 | 6300 | 8000 |
BAND CENTER FREQUENCY (Hz)
Figure 1: 1/3 octave spectra for the pink noise signal and the noise signal specified in ANSI S3.42- 1997.
2.2.2Dual-direction pink noise signal (Verifit only)
This unique test signal is available in the dual-source Verifit test chamber and for real-ear measurements by adding an additional sound-field speaker. It is similar to the pink noise signal described previously except that the component tones are presented simultaneously from two speakers, half from each. The level of each tone is independently controlled at the reference microphone. Two frequency response curves are generated simultaneously -one for each speaker. This provides a real-time measure of the functioning of the directional features of hearing aids that is independent of compression or noise reduction algorithms. Measurement methods which sequentially measure response from different directions work only with these features disabled.
2.2.3Real-speech signals
Real-speech signals are provided in Speechmap for both REM and S-REM measurement modes. Four different
07/05/06 | © Etymonic Design Incorporated, 41 Byron Ave., Dorchester, ON, Canada N0L 1G0 | Page 2 |
| USA 800-265-2093 519-268-3313 FAX 519-268-3256 www.audioscan.com | |
Contents
USER’S Guide
General Operation
Table of Contents
General Setup
Getting Started
Ansi S3.22 Hearing Aid Tests
Hearing Instrument Tests Setup
Other Hearing Instrument Tests
REM Manual Control
Troubleshooting
Speechmap Fitting Procedures
Speechmap Technical Details
Section overview
About RM500SL
Help is available on the following topics by pressing Help
FastFacts 1.3 Electronic users guide
FastFacts 1.4 Warranty, trademarks, acknowledgements
Warranty, Trademarks, Acknowledgements
Cispr
Electromagnetic Compatability EMC
Following topics are covered in this section
FastFacts 2.1 Getting started Unpacking and connecting
FastFacts 2.2 Unpacking and connecting
Getting Started
FastFacts 2.3 Microphone connection
Microphone connection
Probe microphone
FastFacts 2.4 Microphone care Battery pill use and care
FastFacts 2.5 Battery pill use and care
Microphone care Coupler microphone
FastFacts 2.6 Mouse, keyboard, barcode scanner
Mouse, keyboard, barcode scanner
Section overview This section covers the following topics
General Operation
External printer, auxiliary audio outputs
FastFacts 2.7 External printer, auxiliary audio outputs
Function Keypad
Input device operation
FastFacts 3.2 Input device operation
FastFacts 3.3 Barcode data input
Barcode data input
FastFacts 3.4 Keypad keys
Keypad keys Function Button
FastFacts 3.5 Menus, lists and buttons
Menus, lists and buttons Setup Box List Box
Screen messages
FastFacts 3.7 Program memory removal and replacement
Software updating
General Setup
FastFacts 4.1 General setup Date and time setup
FastFacts 4.2 Date and time setup
FastFacts 4.3 Display settings Saving test setup
Display settings
FastFacts 4.4 Saving test setup
FastFacts 5.1 Networking Networking requirements
Networking
FastFacts 5.2 Networking requirements
FastFacts 5.3 Networking setup
Networking setup
On the RM500SL, see
Single computer connection
FastFacts 5.4 Single computer connection
FastFacts 6.1 Printing and storing results
Printing and Storing Results
To load a new roll of paper
Internal printer paper loading
FastFacts 6.2 Internal printer paper loading
Barcodes, headers and comments on printouts
FastFacts 6.4 Local printer setup
FastFacts 6.5 Printing to a USB memory stick
Printing to a USB memory stick
FastFacts 6.6 Network printer setup
Network printer setup
FastFacts 6.8 Data storage to a network file
Printing to a network file
FastFacts 7.4 Output or gain scale
Hearing Instrument Tests Setup
FastFacts 7.2 HIT a or B display Graph or table format
FastFacts 7.3 Graph or table format Output or gain scale
To change the averaging frequencies
Hide or Show HIT curves
FastFacts 7.7 Ansi test frequencies
FastFacts 7.8 HIT calibration facts
HIT calibration facts
FastFacts 7.9 Calibration of HIT reference microphone
FastFacts 7.10 Calibration check for coupler microphone
Calibration check for coupler microphone
Coupling the hearing instrument
FastFacts 7.12 Positioning the hearing instrument
Positioning the hearing instrument
FastFacts 8.2 Ansi S3.22-1996 facts
Ansi S3.22 Hearing Aid Tests
Turn the instrument on
Ansi S3.22-2003 facts
FastFacts 8.4 Ansi 1996 Linear and AGC tests
FastFacts 8.5 Ansi 2003 Linear and AGC tests
Ansi 2003 Linear and AGC tests
FastFacts 8.6 Ansi test results
Ansi test results
FastFacts 8.7 Ansi input-output curves
Ansi input-output curves
FastFacts 8.8 Ansi telecoil terminology
Ansi telecoil terminology
FastFacts 8.9 Ansi telephone simulator Tmfs test
Ansi telephone simulator Tmfs test
FastFacts 8.10 Telecoil test results
Telecoil test results
Turn the hearing instrument on
Other Hearing Instrument Tests
FastFacts 9.2 Harmonic distortion
FastFacts 9.3 Multicurve procedure Multicurve results
Multicurve procedure
FastFacts 9.4 Multicurve results
Spectral analysis in Multicurve
FastFacts 9.6 Battery drain test
FastFacts 9.8 Sound level meter using manual control
Manual control procedure
FastFacts 10.2 REM calibration facts
Real-Ear Measurement Setup
FastFacts 10.3 Calibration of REM probe microphone
Calibration of REM probe microphone
Calibration check for probe module
FastFacts 10.5 Max TM SPL setup 10.6 ABR nHL to eHL setup
Max TM SPL setup
FastFacts 10.6 ABR nHL to eHL setup
FastFacts 10.8 Positioning the probe tube
Positioning the client
FastFacts 11.3 Graph, table or 2cc target format
REM Screen Setup
FastFacts 11.5 Hide or show REM curves
REM Manual Control
SPL or HL scale
FastFacts 11.4 SPL or HL scale Hide or show REM curves
Manual control measurements
FastFacts 12.3 Sound level meter using REM probe
Sensory loss simulator description
Sensory Loss Simulator
FastFacts 13.2 Sensory loss simulator operation
FastFacts 14.1 Occlusion effect measurement
Occlusion Effect Test
Occlussion effect measurement
Turn the hearing aid OFF
FastFacts 15.2 Recd facts
Recd real-ear to coupler difference
FastFacts 15.3 Recd coupler response
Recd coupler response
Allow the foam tip to fully expand in the ear
Recd real-ear response
FastFacts 15.4 Recd real-ear response
FastFacts 15.5 Recd results
Recd results
FastFacts 16.1 Insertion gain Insertion gain in SPL
Insertion Gain
FastFacts 16.2 Insertion gain in SPL
FastFacts 16.3 Insertion gain in HL
Insertion gain in HL
FastFacts 16.4 Audiometric data entry
Audiometric data entry
To enter data from a barcode see 3.3 Barcode data input
To enter assessment data and parameters manually
FastFacts 16.5 Reur measurement procedure
Reur measurement procedure
FastFacts 16.6 Rear measurement procedure
Rear measurement procedure
FastFacts 16.7 SII calculation in Insertion gain
SII calculation in Insertion gain
FastFacts 17.1 Speechmap 17.2 Speechmap facts
Speechmap
FastFacts 17.2 Speechmap facts
FastFacts 17.3 DSL 5.0 in Speechmap
DSL 5.0 in Speechmap
FastFacts 17.4 DSL 5.0 changes NAL-NL1 in Speechmap
DSL 5.0 changes
FastFacts 17.5 NAL-NL1 in Speechmap
FastFacts 17.6 Using Speechmap
Using Speechmap
To enter audiometric data, see 18.2 Assessment data entry
0611 RM500SL User’s Guide Version
FastFacts 17.9 REM or S-REM mode
REM or S-REM mode
FastFacts 17.10 SII calculation in Speechmap
17.10SII calculation in Speechmap
To enter assessment data manually
Speechmap Fitting Procedures
FastFacts 18.2 Assessment data entry
FastFacts 18.3 Assessment data choices
Assessment data choices
FastFacts 18.4 Assessment data choices
From 18.3 Assessment data choices
FastFacts 18.5 Fitting to targets for soft speech
Fitting to targets for soft speech
FastFacts 18.6 Fitting to targets for loud sounds
Fitting to targets for loud sounds
FastFacts 18.8 Open fittings in Speechmap
Fitting to targets for mid-level speech
FastFacts 18.9 FM fitting and verification
FM fitting and verification
FastFacts 19.2 Speechmap stimuli
Speechmap Technical Details
FastFacts 19.3 Stimulus levels
Stimulus levels
FastFacts 19.4 Microphone location effects
Microphone location effects Microphone Location Effects
FastFacts 19.5 Speech signal analysis
Speech signal analysis
FastFacts 20.1 Troubleshooting Self test failures
Troubleshooting
FastFacts 20.2 Self test failures
FastFacts 20.4 HIT curves inconsistent
FastFacts 20.7 No HIT reference mic. detected
No HIT reference mic. detected
FastFacts 20.10 Invalid REM calibration
Invalid HIT calibration Invalid HIT Calibration
FastFacts 20.12 Barcode scanner malfunction
20.11REM speaker overdriven
HIT Chamber
Technical Specifications
Storage & Transportation
General
Glossary
Page
Page
Page
References
Page
Page
0611 RM500SL User’s Guide Version
Test signals
Verifit and RM500SL Test Signals and Analysis
Introduction
Broad-band signals
Loud Vocal Effort Effect re nominal band SPL
Pink noise signal analysis
Analysis of broad-band signals
Verifit / RM500SL Speechmap
07/05/06
Example of speech envelope and Ltass
07/05/06
Ltass for Speechmap speech signals at average vocal effort
Ltass
Dynamic