HP manual Verifit and RM500SL Test Signals and Analysis, Introduction, Test signals

Page 102

VERIFIT and RM500SL Test Signals and Analysis

Bill Cole BASc, PEng

 

 

 

1 Introduction

 

Some useful facts about broad-band signals

The Verifit and RM500SL are full-duplex

Ø

Overall SPL is the SPL in a band containing all significant frequency

dual-channel audio measurement systems

 

components of the signal.

 

designed for the testing and fitting of all

Ø

Spectrum level is the SPL in a band 1 Hz wide.

types of hearing instruments and many

Ø

Band SPL is the SPL in a restricted frequency range. If the signal is

assistive listening devices. They provide

Ø

uniform in the band, band SPL = Spectrum level + 10*log(bandwidth).

an acoustic test chamber for the testing of

A spectrum is the band SPL, or spectrum level, in a series of

 

adjacent bands.

 

devices coupled to standard 2 cm3

 

 

couplers as well as sound-field speaker(s)

Ø For a broad-band signal, the overall SPL is greater than the band SPL

 

and the band SPL is greater than the spectrum level. For a pure tone,

for real-ear measurement (REM) of hearing

 

the overall SPL, the band SPL and the spectrum level are the same.

aid performance. The Verifit test chamber

Ø

A white noise signal has a spectrum level that is independent of

contains two loudspeakers for the

 

frequency (i.e. constant SPL per Hertz).

 

functional testing of directional hearing

Ø

A pink noise signal has a spectrum level that is inversely proportional

instruments. Test signals are delivered to

 

to frequency (i.e. constant SPL per octave).

 

the device under test via the test chamber

Ø

A fractional octave band (e.g. 1/3 octave) has a bandwidth that is

loudspeaker(s), the test chamber telephone

 

proportional to frequency.

 

magnetic-field simulator (TMFS), the test

Ø A pink noise signal has a flat spectrum when analyzed in fractional-

 

octave bands.

 

chamber magnetic field test loop (Verifit

Ø

A critical band is a band within which the loudness of a continuously-

only) or the sound-field loudspeaker(s).

 

distributed signal of constant SPL is independent of bandwidth.

Device output signals are measured in the

 

 

 

2 cm3 coupler via the coupler microphone or in a real ear via the probe microphone. Data from the 2 cm3 coupler

may be viewed as coupler SPL, coupler gain or as estimated SPL in the ear canal (simulated real-ear measurements, S-REM). Data from the probe microphone may be viewed as ear canal SPL, as ear canal SPL re normal hearing threshold (i.e. dB HL), as insertion gain or as ear canal SPL in the context of an auditory area (Speechmap®).

2 The test signals

Four types of test signals are currently available. These are tonal signals, pseudo-random noise, digitized real speech signals and the ICRA distorted speech signal. Test signals are generated in real time by the digital signal processor (DSP) or played from 16 bit binary audio files stored in the on-board flash memory. In the Verifit, these audio files are up-loaded from the internal compact disc (CD) drive at power up. In order to provide a repeatable acoustic signal to the device under test, a reference microphone is used in conjunction with a digital control loop to maintain the desired band level at each frequency. Live audio (speech, music etc.) may also be used as a test signal but it is not controlled for spectrum or level.

2.1Narrow-band signals

2.1.1Pure tones

Pure-tones are used in the ANSI hearing aid tests, in Multicurve 2 cm3 frequency response tests (gain or output), in 2 cm3 and real-ear single frequency (manual) tests, in real-ear insertion gain tests and in tone-burst maximum output (MPO) tests. Frequencies used are 1/12 octave except for the MPO test which uses 1/3 octave frequencies. Levels from 40 to 90 dB SPL are available at the reference microphone. Control of the level at each frequency is maintained by measuring the frequency response of the signal path to the reference microphone 256 ms before each test using a 256 ms, 50 – 55 dB SPL, multi-tone complex or a 256 ms chirp, depending on the test to follow. Drive levels at each frequency are then set to achieve the desired band levels at the reference

07/05/06

© Etymonic Design Incorporated, 41 Byron Ave., Dorchester, ON, Canada N0L 1G0

Page 1

 

USA 800-265-2093 519-268-3313 FAX 519-268-3256 www.audioscan.com

 

Image 102
Contents USER’S Guide Getting Started Table of ContentsGeneral Setup General OperationHearing Instrument Tests Setup Ansi S3.22 Hearing Aid TestsOther Hearing Instrument Tests REM Manual Control Speechmap Fitting Procedures TroubleshootingSpeechmap Technical Details About RM500SL Section overviewHelp is available on the following topics by pressing Help FastFacts 1.3 Electronic users guide Warranty, Trademarks, Acknowledgements FastFacts 1.4 Warranty, trademarks, acknowledgementsElectromagnetic Compatability EMC CisprGetting Started FastFacts 2.1 Getting started Unpacking and connectingFastFacts 2.2 Unpacking and connecting Following topics are covered in this sectionMicrophone connection FastFacts 2.3 Microphone connectionMicrophone care Coupler microphone FastFacts 2.4 Microphone care Battery pill use and careFastFacts 2.5 Battery pill use and care Probe microphoneMouse, keyboard, barcode scanner FastFacts 2.6 Mouse, keyboard, barcode scannerFastFacts 2.7 External printer, auxiliary audio outputs General OperationExternal printer, auxiliary audio outputs Section overview This section covers the following topicsInput device operation Function KeypadFastFacts 3.2 Input device operation Barcode data input FastFacts 3.3 Barcode data inputKeypad keys Function Button FastFacts 3.4 Keypad keysMenus, lists and buttons Setup Box List Box FastFacts 3.5 Menus, lists and buttonsFastFacts 3.7 Program memory removal and replacement Screen messagesFastFacts 4.2 Date and time setup General SetupFastFacts 4.1 General setup Date and time setup Software updatingDisplay settings FastFacts 4.3 Display settings Saving test setupFastFacts 4.4 Saving test setup Networking FastFacts 5.1 Networking Networking requirementsFastFacts 5.2 Networking requirements Networking setup FastFacts 5.3 Networking setupSingle computer connection On the RM500SL, seeFastFacts 5.4 Single computer connection Printing and Storing Results FastFacts 6.1 Printing and storing resultsInternal printer paper loading To load a new roll of paperFastFacts 6.2 Internal printer paper loading FastFacts 6.4 Local printer setup Barcodes, headers and comments on printoutsPrinting to a USB memory stick FastFacts 6.5 Printing to a USB memory stickNetwork printer setup FastFacts 6.6 Network printer setupPrinting to a network file FastFacts 6.8 Data storage to a network fileFastFacts 7.3 Graph or table format Output or gain scale Hearing Instrument Tests SetupFastFacts 7.2 HIT a or B display Graph or table format FastFacts 7.4 Output or gain scaleHide or Show HIT curves To change the averaging frequenciesFastFacts 7.7 Ansi test frequencies HIT calibration facts FastFacts 7.8 HIT calibration factsFastFacts 7.9 Calibration of HIT reference microphone Calibration check for coupler microphone FastFacts 7.10 Calibration check for coupler microphoneCoupling the hearing instrument Positioning the hearing instrument FastFacts 7.12 Positioning the hearing instrumentAnsi S3.22 Hearing Aid Tests FastFacts 8.2 Ansi S3.22-1996 factsAnsi S3.22-2003 facts Turn the instrument onFastFacts 8.4 Ansi 1996 Linear and AGC tests Ansi 2003 Linear and AGC tests FastFacts 8.5 Ansi 2003 Linear and AGC testsAnsi test results FastFacts 8.6 Ansi test resultsAnsi input-output curves FastFacts 8.7 Ansi input-output curvesAnsi telecoil terminology FastFacts 8.8 Ansi telecoil terminologyAnsi telephone simulator Tmfs test FastFacts 8.9 Ansi telephone simulator Tmfs testTelecoil test results FastFacts 8.10 Telecoil test resultsOther Hearing Instrument Tests Turn the hearing instrument onFastFacts 9.2 Harmonic distortion Multicurve procedure FastFacts 9.3 Multicurve procedure Multicurve resultsFastFacts 9.4 Multicurve results FastFacts 9.6 Battery drain test Spectral analysis in MulticurveManual control procedure FastFacts 9.8 Sound level meter using manual controlReal-Ear Measurement Setup FastFacts 10.2 REM calibration factsCalibration of REM probe microphone FastFacts 10.3 Calibration of REM probe microphoneCalibration check for probe module Max TM SPL setup FastFacts 10.5 Max TM SPL setup 10.6 ABR nHL to eHL setupFastFacts 10.6 ABR nHL to eHL setup Positioning the client FastFacts 10.8 Positioning the probe tubeREM Screen Setup FastFacts 11.3 Graph, table or 2cc target formatFastFacts 11.4 SPL or HL scale Hide or show REM curves REM Manual ControlSPL or HL scale FastFacts 11.5 Hide or show REM curvesFastFacts 12.3 Sound level meter using REM probe Manual control measurementsSensory Loss Simulator Sensory loss simulator descriptionFastFacts 13.2 Sensory loss simulator operation Turn the hearing aid OFF Occlusion Effect TestOcclussion effect measurement FastFacts 14.1 Occlusion effect measurementRecd real-ear to coupler difference FastFacts 15.2 Recd factsRecd coupler response FastFacts 15.3 Recd coupler responseRecd real-ear response Allow the foam tip to fully expand in the earFastFacts 15.4 Recd real-ear response Recd results FastFacts 15.5 Recd resultsInsertion Gain FastFacts 16.1 Insertion gain Insertion gain in SPLFastFacts 16.2 Insertion gain in SPL Insertion gain in HL FastFacts 16.3 Insertion gain in HLTo enter assessment data and parameters manually Audiometric data entryTo enter data from a barcode see 3.3 Barcode data input FastFacts 16.4 Audiometric data entryReur measurement procedure FastFacts 16.5 Reur measurement procedureRear measurement procedure FastFacts 16.6 Rear measurement procedureSII calculation in Insertion gain FastFacts 16.7 SII calculation in Insertion gainSpeechmap FastFacts 17.1 Speechmap 17.2 Speechmap factsFastFacts 17.2 Speechmap facts DSL 5.0 in Speechmap FastFacts 17.3 DSL 5.0 in SpeechmapDSL 5.0 changes FastFacts 17.4 DSL 5.0 changes NAL-NL1 in SpeechmapFastFacts 17.5 NAL-NL1 in Speechmap Using Speechmap FastFacts 17.6 Using SpeechmapTo enter audiometric data, see 18.2 Assessment data entry 0611 RM500SL User’s Guide Version REM or S-REM mode FastFacts 17.9 REM or S-REM mode17.10SII calculation in Speechmap FastFacts 17.10 SII calculation in SpeechmapSpeechmap Fitting Procedures To enter assessment data manuallyFastFacts 18.2 Assessment data entry Assessment data choices FastFacts 18.3 Assessment data choicesFrom 18.3 Assessment data choices FastFacts 18.4 Assessment data choicesFitting to targets for soft speech FastFacts 18.5 Fitting to targets for soft speechFitting to targets for loud sounds FastFacts 18.6 Fitting to targets for loud soundsFitting to targets for mid-level speech FastFacts 18.8 Open fittings in SpeechmapFM fitting and verification FastFacts 18.9 FM fitting and verificationSpeechmap Technical Details FastFacts 19.2 Speechmap stimuliStimulus levels FastFacts 19.3 Stimulus levelsMicrophone location effects Microphone Location Effects FastFacts 19.4 Microphone location effectsSpeech signal analysis FastFacts 19.5 Speech signal analysisTroubleshooting FastFacts 20.1 Troubleshooting Self test failuresFastFacts 20.2 Self test failures FastFacts 20.4 HIT curves inconsistent No HIT reference mic. detected FastFacts 20.7 No HIT reference mic. detectedInvalid HIT calibration Invalid HIT Calibration FastFacts 20.10 Invalid REM calibration20.11REM speaker overdriven FastFacts 20.12 Barcode scanner malfunctionGeneral Technical SpecificationsStorage & Transportation HIT ChamberGlossary Page Page Page References Page Page 0611 RM500SL User’s Guide Version Verifit and RM500SL Test Signals and Analysis Test signalsIntroduction Broad-band signals Loud Vocal Effort Effect re nominal band SPL Analysis of broad-band signals Pink noise signal analysisVerifit / RM500SL Speechmap 07/05/06 Example of speech envelope and Ltass 07/05/06 Ltass for Speechmap speech signals at average vocal effort Ltass Dynamic