HP RM500SL manual FastFacts 19.3 Stimulus levels

Page 85

19.3 Stimulus levels

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1/3 octave band SPL

 

 

Add to band SPL for

 

 

 

 

Speech

 

ICRA

Pink

Loud

Own

FM

FM

 

 

Hz

Standard

Female

Child

Noise

Noise

Voice

Chest

Boom

 

 

200

56.8

54.9

32.3

52.9

53

-2.2

.5

9.0

19.2

 

 

250

56.0

53.1

51.5

55.9

53

3.3

10.7

12.0

22.2

 

 

315

53.0

48.4

57.3

56.9

53

9.6

11.0

16.6

26.0

 

 

400

57.0

55.8

54.1

57.5

53

6.1

10.0

13.3

22.8

 

 

500

58.0

57.3

53.2

57.8

53

7.7

10.1

13.2

25.3

 

 

630

55.0

58.1

61.2

54.3

53

12.4

10.5

21.0

29.1

 

 

800

52.5

54.6

56.5

52.8

53

14.4

10.8

26.4

32.1

 

 

1000

51.0

52.0

52.7

50.4

53

15.0

10.5

26.4

32.7

 

 

1250

50.5

51.0

45.4

51.3

53

14.8

5.5

22.0

33.6

 

 

1600

48.0

51.1

41.1

48.5

53

15.5

4.0

22.1

33.9

 

 

2000

45.0

51.2

39.7

47.0

53

16.2

6.0

21.0

33.0

 

 

2500

44.0

49.3

35.5

44.7

53

14.8

6.0

18.2

31.9

 

 

3150

42.5

46.0

28.8

43.3

53

14.5

2.0

14.4

30.4

 

 

4000

42.0

40.0

31.2

41.1

53

13.2

-0.5

11.8

29.0

 

 

5000

40.0

36.1

30.6

39.6

53

9.2

-1.0

12.4

27.8

 

 

6300

41.5

43.5

32.6

33.5

53

4.2

0.0

11.4

26.2

 

 

8000

41.5

39.9

35.2

28.2

53

2.7

-4.0

11.0

25.8

 

 

Overall

65.0

65.0

65.0

65.0

65

75.0

75.0

84.0

93.0

 

 

LTASS

for standard

speech

is from

Cox &

Moore

(1988).

Loud spectrum is from

 

ANSI S3.5-1997. Own voice spectrum is from Cornelisse et al (1991) adjusted to a

 

1 m reference.

 

 

 

 

 

 

 

 

 

FastFacts 19.3: Stimulus levels

 

 

 

 

 

 

 

0611

RM500SL User’s Guide Version 2.8

Page 85

Image 85
Contents USER’S Guide General Setup Table of ContentsGetting Started General OperationAnsi S3.22 Hearing Aid Tests Hearing Instrument Tests SetupOther Hearing Instrument Tests REM Manual Control Troubleshooting Speechmap Fitting ProceduresSpeechmap Technical Details Section overview About RM500SLHelp is available on the following topics by pressing Help FastFacts 1.3 Electronic users guide FastFacts 1.4 Warranty, trademarks, acknowledgements Warranty, Trademarks, AcknowledgementsCispr Electromagnetic Compatability EMCFastFacts 2.2 Unpacking and connecting FastFacts 2.1 Getting started Unpacking and connectingGetting Started Following topics are covered in this sectionFastFacts 2.3 Microphone connection Microphone connectionFastFacts 2.5 Battery pill use and care FastFacts 2.4 Microphone care Battery pill use and careMicrophone care Coupler microphone Probe microphoneFastFacts 2.6 Mouse, keyboard, barcode scanner Mouse, keyboard, barcode scannerExternal printer, auxiliary audio outputs General OperationFastFacts 2.7 External printer, auxiliary audio outputs Section overview This section covers the following topicsFunction Keypad Input device operationFastFacts 3.2 Input device operation FastFacts 3.3 Barcode data input Barcode data inputFastFacts 3.4 Keypad keys Keypad keys Function ButtonFastFacts 3.5 Menus, lists and buttons Menus, lists and buttons Setup Box List BoxScreen messages FastFacts 3.7 Program memory removal and replacementFastFacts 4.1 General setup Date and time setup General SetupFastFacts 4.2 Date and time setup Software updatingFastFacts 4.3 Display settings Saving test setup Display settingsFastFacts 4.4 Saving test setup FastFacts 5.1 Networking Networking requirements NetworkingFastFacts 5.2 Networking requirements FastFacts 5.3 Networking setup Networking setupOn the RM500SL, see Single computer connectionFastFacts 5.4 Single computer connection FastFacts 6.1 Printing and storing results Printing and Storing ResultsTo load a new roll of paper Internal printer paper loadingFastFacts 6.2 Internal printer paper loading Barcodes, headers and comments on printouts FastFacts 6.4 Local printer setupFastFacts 6.5 Printing to a USB memory stick Printing to a USB memory stickFastFacts 6.6 Network printer setup Network printer setupFastFacts 6.8 Data storage to a network file Printing to a network fileFastFacts 7.2 HIT a or B display Graph or table format Hearing Instrument Tests SetupFastFacts 7.3 Graph or table format Output or gain scale FastFacts 7.4 Output or gain scaleTo change the averaging frequencies Hide or Show HIT curvesFastFacts 7.7 Ansi test frequencies FastFacts 7.8 HIT calibration facts HIT calibration factsFastFacts 7.9 Calibration of HIT reference microphone FastFacts 7.10 Calibration check for coupler microphone Calibration check for coupler microphoneCoupling the hearing instrument FastFacts 7.12 Positioning the hearing instrument Positioning the hearing instrumentFastFacts 8.2 Ansi S3.22-1996 facts Ansi S3.22 Hearing Aid TestsTurn the instrument on Ansi S3.22-2003 factsFastFacts 8.4 Ansi 1996 Linear and AGC tests FastFacts 8.5 Ansi 2003 Linear and AGC tests Ansi 2003 Linear and AGC testsFastFacts 8.6 Ansi test results Ansi test resultsFastFacts 8.7 Ansi input-output curves Ansi input-output curvesFastFacts 8.8 Ansi telecoil terminology Ansi telecoil terminologyFastFacts 8.9 Ansi telephone simulator Tmfs test Ansi telephone simulator Tmfs testFastFacts 8.10 Telecoil test results Telecoil test resultsTurn the hearing instrument on Other Hearing Instrument TestsFastFacts 9.2 Harmonic distortion FastFacts 9.3 Multicurve procedure Multicurve results Multicurve procedureFastFacts 9.4 Multicurve results Spectral analysis in Multicurve FastFacts 9.6 Battery drain testFastFacts 9.8 Sound level meter using manual control Manual control procedureFastFacts 10.2 REM calibration facts Real-Ear Measurement SetupFastFacts 10.3 Calibration of REM probe microphone Calibration of REM probe microphoneCalibration check for probe module FastFacts 10.5 Max TM SPL setup 10.6 ABR nHL to eHL setup Max TM SPL setupFastFacts 10.6 ABR nHL to eHL setup FastFacts 10.8 Positioning the probe tube Positioning the clientFastFacts 11.3 Graph, table or 2cc target format REM Screen SetupSPL or HL scale REM Manual ControlFastFacts 11.4 SPL or HL scale Hide or show REM curves FastFacts 11.5 Hide or show REM curvesManual control measurements FastFacts 12.3 Sound level meter using REM probeSensory loss simulator description Sensory Loss SimulatorFastFacts 13.2 Sensory loss simulator operation Occlussion effect measurement Occlusion Effect TestTurn the hearing aid OFF FastFacts 14.1 Occlusion effect measurementFastFacts 15.2 Recd facts Recd real-ear to coupler differenceFastFacts 15.3 Recd coupler response Recd coupler responseAllow the foam tip to fully expand in the ear Recd real-ear responseFastFacts 15.4 Recd real-ear response FastFacts 15.5 Recd results Recd resultsFastFacts 16.1 Insertion gain Insertion gain in SPL Insertion GainFastFacts 16.2 Insertion gain in SPL FastFacts 16.3 Insertion gain in HL Insertion gain in HLTo enter data from a barcode see 3.3 Barcode data input Audiometric data entryTo enter assessment data and parameters manually FastFacts 16.4 Audiometric data entryFastFacts 16.5 Reur measurement procedure Reur measurement procedureFastFacts 16.6 Rear measurement procedure Rear measurement procedureFastFacts 16.7 SII calculation in Insertion gain SII calculation in Insertion gainFastFacts 17.1 Speechmap 17.2 Speechmap facts SpeechmapFastFacts 17.2 Speechmap facts FastFacts 17.3 DSL 5.0 in Speechmap DSL 5.0 in SpeechmapFastFacts 17.4 DSL 5.0 changes NAL-NL1 in Speechmap DSL 5.0 changesFastFacts 17.5 NAL-NL1 in Speechmap FastFacts 17.6 Using Speechmap Using SpeechmapTo enter audiometric data, see 18.2 Assessment data entry 0611 RM500SL User’s Guide Version FastFacts 17.9 REM or S-REM mode REM or S-REM modeFastFacts 17.10 SII calculation in Speechmap 17.10SII calculation in SpeechmapTo enter assessment data manually Speechmap Fitting ProceduresFastFacts 18.2 Assessment data entry FastFacts 18.3 Assessment data choices Assessment data choicesFastFacts 18.4 Assessment data choices From 18.3 Assessment data choicesFastFacts 18.5 Fitting to targets for soft speech Fitting to targets for soft speechFastFacts 18.6 Fitting to targets for loud sounds Fitting to targets for loud soundsFastFacts 18.8 Open fittings in Speechmap Fitting to targets for mid-level speechFastFacts 18.9 FM fitting and verification FM fitting and verificationFastFacts 19.2 Speechmap stimuli Speechmap Technical DetailsFastFacts 19.3 Stimulus levels Stimulus levelsFastFacts 19.4 Microphone location effects Microphone location effects Microphone Location EffectsFastFacts 19.5 Speech signal analysis Speech signal analysisFastFacts 20.1 Troubleshooting Self test failures TroubleshootingFastFacts 20.2 Self test failures FastFacts 20.4 HIT curves inconsistent FastFacts 20.7 No HIT reference mic. detected No HIT reference mic. detectedFastFacts 20.10 Invalid REM calibration Invalid HIT calibration Invalid HIT CalibrationFastFacts 20.12 Barcode scanner malfunction 20.11REM speaker overdrivenStorage & Transportation Technical SpecificationsGeneral HIT ChamberGlossary Page Page Page References Page Page 0611 RM500SL User’s Guide Version Test signals Verifit and RM500SL Test Signals and AnalysisIntroduction Broad-band signals Loud Vocal Effort Effect re nominal band SPL Pink noise signal analysis Analysis of broad-band signalsVerifit / RM500SL Speechmap 07/05/06 Example of speech envelope and Ltass 07/05/06 Ltass for Speechmap speech signals at average vocal effort Ltass Dynamic