20.5 HIT curves differ from specifications

The OSPL90 and response curves don’t look like the manufacturer’s.

vCheck that all controls or program settings are the same as those used by the manufacturer.

vCheck that vents on earmolds or ITE instruments are plugged at the faceplate.

vCheck that ITE instruments are well sealed to the coupler.

vWhen puttying a hearing instrument into the ITE coupler, be sure to keep the canal portion of the instrument flush with the inside of the 2cc volume.

vEnsure that the HIT chamber is sealed tightly.

vEnsure that the HIT microphone calibration was performed correctly.

vCheck that the coupler is screwed tightly onto the coupler microphone.

FastFacts 20.5: HIT curves differ from specifications 20.6 HIT speaker overdriven

HIT speaker overdriven!

This message indicates that the test chamber speaker is unable to produce the SPL required for a test.

vCheck that the test chamber lid is closed and sealed.

vCheck that the HIT calibration looks normal.

vCheck that the hearing instrument under test is not feeding back.

vTry running a REM test.

FastFacts 20.6: HIT speaker overdriven 20.7 No HIT reference mic. detected

No HIT reference mic. detected!

This message may be displayed when starting a hearing instrument test. It indicates that the RM500SL is not measuring any sound at the HIT reference microphone. The HIT reference microphone is the white microphone located in the test chamber.

vEnsure that the white HIT reference microphone is secure. Unplug it and plug it in again, ensuring that it ‘clicks’ into place. Try the HIT calibration procedure again.

vUnplug the HIT reference microphone and plug the REM probe microphone in its place. Try HIT calibration using the reference microphone of the REM probe microphone in place of the HIT reference microphone. If this is successful, the HIT reference microphone is defective and must be replaced.

FastFacts 20.7: No HIT reference mic. detected

0611

RM500SL User’s Guide Version 2.8

Page 90

Image 90
Contents USER’S Guide Getting Started Table of ContentsGeneral Setup General OperationHearing Instrument Tests Setup Ansi S3.22 Hearing Aid TestsOther Hearing Instrument Tests REM Manual Control Speechmap Fitting Procedures TroubleshootingSpeechmap Technical Details About RM500SL Section overviewHelp is available on the following topics by pressing Help FastFacts 1.3 Electronic users guide Warranty, Trademarks, Acknowledgements FastFacts 1.4 Warranty, trademarks, acknowledgementsElectromagnetic Compatability EMC CisprGetting Started FastFacts 2.1 Getting started Unpacking and connectingFastFacts 2.2 Unpacking and connecting Following topics are covered in this sectionMicrophone connection FastFacts 2.3 Microphone connectionMicrophone care Coupler microphone FastFacts 2.4 Microphone care Battery pill use and careFastFacts 2.5 Battery pill use and care Probe microphoneMouse, keyboard, barcode scanner FastFacts 2.6 Mouse, keyboard, barcode scannerFastFacts 2.7 External printer, auxiliary audio outputs General OperationExternal printer, auxiliary audio outputs Section overview This section covers the following topicsInput device operation Function KeypadFastFacts 3.2 Input device operation Barcode data input FastFacts 3.3 Barcode data inputKeypad keys Function Button FastFacts 3.4 Keypad keysMenus, lists and buttons Setup Box List Box FastFacts 3.5 Menus, lists and buttonsFastFacts 3.7 Program memory removal and replacement Screen messagesFastFacts 4.2 Date and time setup General SetupFastFacts 4.1 General setup Date and time setup Software updatingDisplay settings FastFacts 4.3 Display settings Saving test setupFastFacts 4.4 Saving test setup Networking FastFacts 5.1 Networking Networking requirementsFastFacts 5.2 Networking requirements Networking setup FastFacts 5.3 Networking setupSingle computer connection On the RM500SL, seeFastFacts 5.4 Single computer connection Printing and Storing Results FastFacts 6.1 Printing and storing resultsInternal printer paper loading To load a new roll of paperFastFacts 6.2 Internal printer paper loading FastFacts 6.4 Local printer setup Barcodes, headers and comments on printoutsPrinting to a USB memory stick FastFacts 6.5 Printing to a USB memory stickNetwork printer setup FastFacts 6.6 Network printer setupPrinting to a network file FastFacts 6.8 Data storage to a network fileFastFacts 7.3 Graph or table format Output or gain scale Hearing Instrument Tests SetupFastFacts 7.2 HIT a or B display Graph or table format FastFacts 7.4 Output or gain scaleHide or Show HIT curves To change the averaging frequenciesFastFacts 7.7 Ansi test frequencies HIT calibration facts FastFacts 7.8 HIT calibration factsFastFacts 7.9 Calibration of HIT reference microphone Calibration check for coupler microphone FastFacts 7.10 Calibration check for coupler microphoneCoupling the hearing instrument Positioning the hearing instrument FastFacts 7.12 Positioning the hearing instrumentAnsi S3.22 Hearing Aid Tests FastFacts 8.2 Ansi S3.22-1996 factsAnsi S3.22-2003 facts Turn the instrument onFastFacts 8.4 Ansi 1996 Linear and AGC tests Ansi 2003 Linear and AGC tests FastFacts 8.5 Ansi 2003 Linear and AGC testsAnsi test results FastFacts 8.6 Ansi test resultsAnsi input-output curves FastFacts 8.7 Ansi input-output curvesAnsi telecoil terminology FastFacts 8.8 Ansi telecoil terminologyAnsi telephone simulator Tmfs test FastFacts 8.9 Ansi telephone simulator Tmfs testTelecoil test results FastFacts 8.10 Telecoil test resultsOther Hearing Instrument Tests Turn the hearing instrument onFastFacts 9.2 Harmonic distortion Multicurve procedure FastFacts 9.3 Multicurve procedure Multicurve resultsFastFacts 9.4 Multicurve results FastFacts 9.6 Battery drain test Spectral analysis in MulticurveManual control procedure FastFacts 9.8 Sound level meter using manual controlReal-Ear Measurement Setup FastFacts 10.2 REM calibration factsCalibration of REM probe microphone FastFacts 10.3 Calibration of REM probe microphoneCalibration check for probe module Max TM SPL setup FastFacts 10.5 Max TM SPL setup 10.6 ABR nHL to eHL setupFastFacts 10.6 ABR nHL to eHL setup Positioning the client FastFacts 10.8 Positioning the probe tubeREM Screen Setup FastFacts 11.3 Graph, table or 2cc target formatFastFacts 11.4 SPL or HL scale Hide or show REM curves REM Manual ControlSPL or HL scale FastFacts 11.5 Hide or show REM curvesFastFacts 12.3 Sound level meter using REM probe Manual control measurementsSensory Loss Simulator Sensory loss simulator descriptionFastFacts 13.2 Sensory loss simulator operation Turn the hearing aid OFF Occlusion Effect TestOcclussion effect measurement FastFacts 14.1 Occlusion effect measurementRecd real-ear to coupler difference FastFacts 15.2 Recd factsRecd coupler response FastFacts 15.3 Recd coupler responseRecd real-ear response Allow the foam tip to fully expand in the earFastFacts 15.4 Recd real-ear response Recd results FastFacts 15.5 Recd resultsInsertion Gain FastFacts 16.1 Insertion gain Insertion gain in SPLFastFacts 16.2 Insertion gain in SPL Insertion gain in HL FastFacts 16.3 Insertion gain in HLTo enter assessment data and parameters manually Audiometric data entryTo enter data from a barcode see 3.3 Barcode data input FastFacts 16.4 Audiometric data entryReur measurement procedure FastFacts 16.5 Reur measurement procedureRear measurement procedure FastFacts 16.6 Rear measurement procedureSII calculation in Insertion gain FastFacts 16.7 SII calculation in Insertion gainSpeechmap FastFacts 17.1 Speechmap 17.2 Speechmap factsFastFacts 17.2 Speechmap facts DSL 5.0 in Speechmap FastFacts 17.3 DSL 5.0 in SpeechmapDSL 5.0 changes FastFacts 17.4 DSL 5.0 changes NAL-NL1 in SpeechmapFastFacts 17.5 NAL-NL1 in Speechmap Using Speechmap FastFacts 17.6 Using SpeechmapTo enter audiometric data, see 18.2 Assessment data entry 0611 RM500SL User’s Guide Version REM or S-REM mode FastFacts 17.9 REM or S-REM mode17.10SII calculation in Speechmap FastFacts 17.10 SII calculation in SpeechmapSpeechmap Fitting Procedures To enter assessment data manuallyFastFacts 18.2 Assessment data entry Assessment data choices FastFacts 18.3 Assessment data choicesFrom 18.3 Assessment data choices FastFacts 18.4 Assessment data choicesFitting to targets for soft speech FastFacts 18.5 Fitting to targets for soft speechFitting to targets for loud sounds FastFacts 18.6 Fitting to targets for loud soundsFitting to targets for mid-level speech FastFacts 18.8 Open fittings in SpeechmapFM fitting and verification FastFacts 18.9 FM fitting and verificationSpeechmap Technical Details FastFacts 19.2 Speechmap stimuliStimulus levels FastFacts 19.3 Stimulus levelsMicrophone location effects Microphone Location Effects FastFacts 19.4 Microphone location effectsSpeech signal analysis FastFacts 19.5 Speech signal analysisTroubleshooting FastFacts 20.1 Troubleshooting Self test failuresFastFacts 20.2 Self test failures FastFacts 20.4 HIT curves inconsistent No HIT reference mic. detected FastFacts 20.7 No HIT reference mic. detectedInvalid HIT calibration Invalid HIT Calibration FastFacts 20.10 Invalid REM calibration20.11REM speaker overdriven FastFacts 20.12 Barcode scanner malfunctionGeneral Technical SpecificationsStorage & Transportation HIT ChamberGlossary Page Page Page References Page Page 0611 RM500SL User’s Guide Version Verifit and RM500SL Test Signals and Analysis Test signalsIntroduction Broad-band signals Loud Vocal Effort Effect re nominal band SPL Analysis of broad-band signals Pink noise signal analysisVerifit / RM500SL Speechmap 07/05/06 Example of speech envelope and Ltass 07/05/06 Ltass for Speechmap speech signals at average vocal effort Ltass Dynamic