7.5Hide or Show HIT curves

On screens capable of showing more than one curve, you may show or hide test curves. Hiding a curve does not erase it.

To hide or show test curve(s):

1)Highlight, then <PICK> Curve [Hide/Show].

2)Highlight, then <PICK> the desired test curve to toggle between Hide or Show. The curve will toggle on or off when selected.

3)Press <Continue> to return to the test screen.

FastFacts 7.5: Hide or Show HIT curves

7.61996 or 2003 ANSI standard

Automated tests of linear and AGC hearing aids may be conducted according to either the ANSI S-3.22-1996 or ANSI S3.22-2003 test standards. For differences, see 8.2: ANSI S3.22-1996 facts and 8.3: ANSI S3.22-2003 facts.

To select the standard to be used:

1)Press <Setup>, then highlight and <PICK> [ANSI]

2)Highlight and <PICK> the version of the ANSI standard you wish to use.

3)Press <Continue> to exit

FastFacts 7.6: 1996 or 2003 ANSI standard

7.7ANSI test frequencies

The Frequencies feature allows hearing instruments to be evaluated at the Special Purpose Average (SPA) frequencies defined by ANSI S3.22, rather than the default High Frequency Average (HFA) frequencies. The choice of SPA (or HFA) frequencies influences the average OSPL90, average full-on gain, reference test gain, bandwidth and equivalent input noise. The frequencies at which distortion is measured are one-half the selected SPA frequencies. You can also select the frequencies at which attack and release times are measured.

To change the averaging frequencies:

1)Press <Setup>, then highlight and <PICK> [Frequencies].

2)Highlight and <PICK> the HFA/SPA frequencies box, to generate a drop-down list box of alternate test frequencies.

3)Highlight the desired HFA/SPA test frequencies and press <PICK> to select.

To change the frequencies for attack and release time measurements:

In the list of available frequencies, highlight and <PICK> the frequencies to be used to measure attack and release time.

Press <Continue> to exit.

FastFacts 7.7: ANSI test frequencies

0611

RM500SL User’s Guide Version 2.8

Page 32

Image 32
Contents USER’S Guide Table of Contents General SetupGetting Started General OperationOther Hearing Instrument Tests Hearing Instrument Tests SetupAnsi S3.22 Hearing Aid Tests REM Manual Control Speechmap Technical Details Speechmap Fitting ProceduresTroubleshooting Help is available on the following topics by pressing Help About RM500SLSection overview FastFacts 1.3 Electronic users guide Warranty, Trademarks, Acknowledgements FastFacts 1.4 Warranty, trademarks, acknowledgementsElectromagnetic Compatability EMC CisprFastFacts 2.1 Getting started Unpacking and connecting FastFacts 2.2 Unpacking and connectingGetting Started Following topics are covered in this sectionMicrophone connection FastFacts 2.3 Microphone connectionFastFacts 2.4 Microphone care Battery pill use and care FastFacts 2.5 Battery pill use and careMicrophone care Coupler microphone Probe microphoneMouse, keyboard, barcode scanner FastFacts 2.6 Mouse, keyboard, barcode scannerGeneral Operation External printer, auxiliary audio outputsFastFacts 2.7 External printer, auxiliary audio outputs Section overview This section covers the following topicsFastFacts 3.2 Input device operation Input device operationFunction Keypad Barcode data input FastFacts 3.3 Barcode data inputKeypad keys Function Button FastFacts 3.4 Keypad keysMenus, lists and buttons Setup Box List Box FastFacts 3.5 Menus, lists and buttonsFastFacts 3.7 Program memory removal and replacement Screen messagesGeneral Setup FastFacts 4.1 General setup Date and time setupFastFacts 4.2 Date and time setup Software updatingFastFacts 4.4 Saving test setup Display settingsFastFacts 4.3 Display settings Saving test setup FastFacts 5.2 Networking requirements NetworkingFastFacts 5.1 Networking Networking requirements Networking setup FastFacts 5.3 Networking setupFastFacts 5.4 Single computer connection Single computer connectionOn the RM500SL, see Printing and Storing Results FastFacts 6.1 Printing and storing resultsFastFacts 6.2 Internal printer paper loading Internal printer paper loadingTo load a new roll of paper FastFacts 6.4 Local printer setup Barcodes, headers and comments on printoutsPrinting to a USB memory stick FastFacts 6.5 Printing to a USB memory stickNetwork printer setup FastFacts 6.6 Network printer setupPrinting to a network file FastFacts 6.8 Data storage to a network fileHearing Instrument Tests Setup FastFacts 7.2 HIT a or B display Graph or table formatFastFacts 7.3 Graph or table format Output or gain scale FastFacts 7.4 Output or gain scaleFastFacts 7.7 Ansi test frequencies Hide or Show HIT curvesTo change the averaging frequencies HIT calibration facts FastFacts 7.8 HIT calibration factsFastFacts 7.9 Calibration of HIT reference microphone Calibration check for coupler microphone FastFacts 7.10 Calibration check for coupler microphoneCoupling the hearing instrument Positioning the hearing instrument FastFacts 7.12 Positioning the hearing instrumentAnsi S3.22 Hearing Aid Tests FastFacts 8.2 Ansi S3.22-1996 factsFastFacts 8.4 Ansi 1996 Linear and AGC tests Ansi S3.22-2003 factsTurn the instrument on Ansi 2003 Linear and AGC tests FastFacts 8.5 Ansi 2003 Linear and AGC testsAnsi test results FastFacts 8.6 Ansi test resultsAnsi input-output curves FastFacts 8.7 Ansi input-output curvesAnsi telecoil terminology FastFacts 8.8 Ansi telecoil terminologyAnsi telephone simulator Tmfs test FastFacts 8.9 Ansi telephone simulator Tmfs testTelecoil test results FastFacts 8.10 Telecoil test resultsFastFacts 9.2 Harmonic distortion Other Hearing Instrument TestsTurn the hearing instrument on FastFacts 9.4 Multicurve results Multicurve procedureFastFacts 9.3 Multicurve procedure Multicurve results FastFacts 9.6 Battery drain test Spectral analysis in MulticurveManual control procedure FastFacts 9.8 Sound level meter using manual controlReal-Ear Measurement Setup FastFacts 10.2 REM calibration factsCalibration of REM probe microphone FastFacts 10.3 Calibration of REM probe microphoneCalibration check for probe module FastFacts 10.6 ABR nHL to eHL setup Max TM SPL setupFastFacts 10.5 Max TM SPL setup 10.6 ABR nHL to eHL setup Positioning the client FastFacts 10.8 Positioning the probe tubeREM Screen Setup FastFacts 11.3 Graph, table or 2cc target formatREM Manual Control SPL or HL scaleFastFacts 11.4 SPL or HL scale Hide or show REM curves FastFacts 11.5 Hide or show REM curvesFastFacts 12.3 Sound level meter using REM probe Manual control measurementsFastFacts 13.2 Sensory loss simulator operation Sensory Loss SimulatorSensory loss simulator description Occlusion Effect Test Occlussion effect measurementTurn the hearing aid OFF FastFacts 14.1 Occlusion effect measurementRecd real-ear to coupler difference FastFacts 15.2 Recd factsRecd coupler response FastFacts 15.3 Recd coupler responseFastFacts 15.4 Recd real-ear response Recd real-ear responseAllow the foam tip to fully expand in the ear Recd results FastFacts 15.5 Recd resultsFastFacts 16.2 Insertion gain in SPL Insertion GainFastFacts 16.1 Insertion gain Insertion gain in SPL Insertion gain in HL FastFacts 16.3 Insertion gain in HLAudiometric data entry To enter data from a barcode see 3.3 Barcode data inputTo enter assessment data and parameters manually FastFacts 16.4 Audiometric data entryReur measurement procedure FastFacts 16.5 Reur measurement procedureRear measurement procedure FastFacts 16.6 Rear measurement procedureSII calculation in Insertion gain FastFacts 16.7 SII calculation in Insertion gainFastFacts 17.2 Speechmap facts SpeechmapFastFacts 17.1 Speechmap 17.2 Speechmap facts DSL 5.0 in Speechmap FastFacts 17.3 DSL 5.0 in SpeechmapFastFacts 17.5 NAL-NL1 in Speechmap DSL 5.0 changesFastFacts 17.4 DSL 5.0 changes NAL-NL1 in Speechmap To enter audiometric data, see 18.2 Assessment data entry Using SpeechmapFastFacts 17.6 Using Speechmap 0611 RM500SL User’s Guide Version REM or S-REM mode FastFacts 17.9 REM or S-REM mode17.10SII calculation in Speechmap FastFacts 17.10 SII calculation in SpeechmapFastFacts 18.2 Assessment data entry Speechmap Fitting ProceduresTo enter assessment data manually Assessment data choices FastFacts 18.3 Assessment data choicesFrom 18.3 Assessment data choices FastFacts 18.4 Assessment data choicesFitting to targets for soft speech FastFacts 18.5 Fitting to targets for soft speechFitting to targets for loud sounds FastFacts 18.6 Fitting to targets for loud soundsFitting to targets for mid-level speech FastFacts 18.8 Open fittings in SpeechmapFM fitting and verification FastFacts 18.9 FM fitting and verificationSpeechmap Technical Details FastFacts 19.2 Speechmap stimuliStimulus levels FastFacts 19.3 Stimulus levelsMicrophone location effects Microphone Location Effects FastFacts 19.4 Microphone location effectsSpeech signal analysis FastFacts 19.5 Speech signal analysisFastFacts 20.2 Self test failures TroubleshootingFastFacts 20.1 Troubleshooting Self test failures FastFacts 20.4 HIT curves inconsistent No HIT reference mic. detected FastFacts 20.7 No HIT reference mic. detectedInvalid HIT calibration Invalid HIT Calibration FastFacts 20.10 Invalid REM calibration20.11REM speaker overdriven FastFacts 20.12 Barcode scanner malfunctionTechnical Specifications Storage & TransportationGeneral HIT ChamberGlossary Page Page Page References Page Page 0611 RM500SL User’s Guide Version Introduction Verifit and RM500SL Test Signals and AnalysisTest signals Broad-band signals Loud Vocal Effort Effect re nominal band SPL Analysis of broad-band signals Pink noise signal analysisVerifit / RM500SL Speechmap 07/05/06 Example of speech envelope and Ltass 07/05/06 Ltass for Speechmap speech signals at average vocal effort Ltass Dynamic